Continuous Wave Transducers; Protective Transducer Holder - Siemens Acuson S1000 Instructions For Use Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

Continuous Wave Transducers

Continuous wave transducer port.
Caution: You must freeze the system before connecting or disconnecting a CW transducer.
To connect a continuous wave transducer:
1.
Align the connector key until it fits smoothly into the receptacle.
2.
Insert the connector into the system port until it locks in position.
To disconnect a continuous wave transducer:
Caution: To avoid damaging the transducer cable, do not pull on the cable to disconnect the
transducer. Use the following instructions.
1.
Pull on the connector housing ring to disengage the locking mechanism. Continue pulling
on the ring to remove the connector from the system port.
2.
Store each transducer in its protective carrying case.
Example of CW transducer connector.

Protective Transducer Holder

Caution: Transducer holders have variable sizes, both in depth and diameter. To avoid
transducer damage, you must use the holder or insert provided for transducers that have small
or large diameter handles or for specialty transducers, such as endocavity transducers.
After connecting a transducer to the system, place the transducer in the protective holder
attached to the control panel platform. There is also a holder for the coupling agent (gel).
Transducer Holders
The transducer holders on the sides of the control panel are interchangeable and replaceable.
I n s t r u c t i o n s f o r U s e
1
Connector
2
Retractable connector housing
3
Transducer cable
3 S y s t e m S e t u p