Imaging Functions that Change Acoustic Output

WARNING: Observe the real-time display of mechanical and thermal indices (MI/TI) at all times.
In addition to the adjustment of the transmit power, adjustment of the following imaging
functions and/or controls may affect the acoustic output:
Automatic Time-out
Color Ensemble Size
Color ROI Position
Color ROI Size
Doppler Gate Position
Doppler Color PRF
Doppler Gate Size
Exam Type
Field of View (Scan Angle)
Focus
Frame Rate
Freeze
Image Depth
2 - 1 8
Imaging Mode
Line Density/Resolution
M-mode ROI Position
Multi-Frequency
Power On/Off
Presets
Reset
Transducer
Transmit Power
Update
Gel Pad
Zoom