MagMaster
TM
Electromagnetic
Flowmeters
Keypad Version
IM/MM-QRG Iss 1 5/96
ABB Kent-Taylor
ASEA BROWN BOVERI
CONTROLS AND DISPLAY
Kent-Taylor
MagMaster
Upper
32.8
Display
Ltr/s
>42315
Lower
Display
Membrane
Resets totaliser, if parameter
Switches
'Tot Clr En' is enabled
Controls and Display
Upper display gives continual update of flow rate in
selected units.
By pressing the
key, the lower display steps through
the following sequence:
>
Forward flow total value.
<
Reverse flow total value
*
Net flow total value
Alm
Active alarms
Any alarms are displayed
sequentially if more than one alarm is present.
'Alm Clr' is displayed when no alarms are present.
Vel
Flow Velocity
%
% of Flow Range.
Pressing the
key resets the flow total displayed on the
upper display, if parameter 'Tot Clr En' is enabled.
Pressing the
key accesses the Login Parameter
where it is necessary to enter a security code before any
other parameters can be accessed – see SECURITY
ACCESS.
Oldends Lane, Stonehouse, Gloucestershire, England GL10 3TA
Telephone: 01453 826661, Facsimile: 01453 826358
MENU LAYOUT
Press
moves
Anlg Fsd
Pls Fact
Alm No1 Idle
Flow Rng
Tot Unit
Flow Unit
Anlg Zero
Pls Cutoff
Tot Mult
Alm No1 En
Press
Flow Mult
Anlg No2
Pls Max
Tot ClrEn
Alm No1 Fault
Flow Time
Anlg mA
Pls Hz
Alm No1 Fwd
Flow Rspns
Anlg Dir Fwd
Pls Idle
Alm No1 Rev
moves
Flow %
Anlg Dir Rev
Pls Size
Alm No1 Cutoff
Flow Probe Ins
Alm No1 Mtsnsr
Flow Probe Prf
Alm No1 Hi
Flow Cutoff
Alm No1 Lo
Alm No1 Anlg
Alm No1 Pls
CONTROLS AND DISPLAY
A – Advancing to Next Page
Advance to
next page
Page 1
Page 2
Parameter 1
Parameter 1
Parameter 2
Parameter 2
or
Parameter 3
Parameter 4
Parameter 5
Parameter 6
B – Moving Between Parameters
Parameter 1
Advance to
next parameter
Parameter 2
Parameter 3
Parameter 4
Inpt
Mtsnsr Trip
Alm No2 Idle
Alm Trip Hi
Inpt Idle
Mtsnsr mv
Alm No2 En
Alm Trip Lo
Alm Trip Hyst
Alm No2 Fault
Alm Trip Disp
Alm No2 Fwd
Alm No2 Rev
Alm No2 Cutoff
Alm No2 Mtsnsr
Alm No2 Hi
Key
Alm No2 Lo
Security Level 1
Alm No2 Anlg
Security Level 2
Alm No2 Pls
C – Adjusting and Storing a Parameter Value
Parameter Value
Adjust
or unit
Advance to next digit
New value is
automatically stored
D – Selecting and Storing a Parameter Choice
Parameter X
Y
Select
Z
New value is
automatically stored
Depressing this switch for 5 seconds and then releasing
it will exit the menu system and return to normal operating
mode.
...MENU LAYOUT
Return to 'Flow Rng' Page
Snsr No
Disp Res
Login Key 1
Test Mode
Login Key 2
Snsr Tag
Test Flow
Disp Mode
Snsr Size
Test %
Snsr Vel
Test Hz
Snsr Fact 1
Test mA
Snsr Fact 2
Test Vel
Snsr Fact 3
Test Alm
Snsr Fact 4
Test Txv
SECURITY ACCESS
Two security code levels, 1 and 2, are available, and are
each accessed with a five digit number.
User Code Level 1 default number is 10760.
Engineer Code Level 2 default number is 56360.
Parameters accessible by the two levels are shown
above.
At the flashing cursor on the first digit of the Login code
number, press either
or
membrane switches to
reach the required digit.
To set this digit and pass to the next digit, depress the
switch. Continue until all digits have been set, and
depress the
switch to enter the complete code.
If an incorrect value is entered, access to subsequent
programming pages is prevented and the display reverts
to the Operating Page.