Software
Revision
3. 1 0
August 19, 2004
September 20, 2004 •
3.20
3.30
October 28, 2004
3.40
Not released
3.50
Not released
3.60
Not released
3.70
February 22, 2006
Page 106
SITRANS LR 400 (Solids) – INSTRUCTION MANUAL
Date
•
DDs: Initial quality value for PROFIBUS
blocks set to 0x4C "uncertain: initial
value"
•
DDs: Corrected several default values
that did not match the device and were
different on a master reset.
•
HCF 275 Hand Held DD changes to
include the new parameters in release
•
All parameters available locally at the
LR 400 device are now documented in
the manual
Extended range calibration limits
•
Specific dead bands per horn type have
been removed. All horn types now have
a dead band of 0.35 meters.
•
Several new values have been added to
antenna and flange parameters to
support the new antenna and flange
options.
•
Reset failures caused by certain filling
speed settings have been corrected.
•
Fault Simulation Feature added, for
service purposes only.
•
MP&F maintenance support added.
New parameters and support added for
Device lifetime, Sensor lifetime,
maintenance required and calibration
required.
•
Corrected "inconsistent parameter
values" warning when tank height
causes a maximum volume less than
the max. volume parameter.
•
SILO2 SEARCH_BIG_X_MAX now
defaults to OFF (not user-editable).
•
Limited density value to reasonable
value. Large invalid values were
causing the LR 400 to reset.
•
Fixed issue regarding Volume/Mass
table entry deletion
•
Fixed issue using g/cm
Changes
3
density units
7ML19985JC02