Siemens sitrans LR 400 Instruction Manual page 109

Hide thumbs Also See for sitrans LR 400:
Table of Contents

Advertisement

Software
Revision
3. 1 0
August 19, 2004
September 20, 2004 •
3.20
3.30
October 28, 2004
3.40
Not released
3.50
Not released
3.60
Not released
3.70
February 22, 2006
Page 106
SITRANS LR 400 (Solids) – INSTRUCTION MANUAL
Date
DDs: Initial quality value for PROFIBUS
blocks set to 0x4C "uncertain: initial
value"
DDs: Corrected several default values
that did not match the device and were
different on a master reset.
HCF 275 Hand Held DD changes to
include the new parameters in release
All parameters available locally at the
LR 400 device are now documented in
the manual
Extended range calibration limits
Specific dead bands per horn type have
been removed. All horn types now have
a dead band of 0.35 meters.
Several new values have been added to
antenna and flange parameters to
support the new antenna and flange
options.
Reset failures caused by certain filling
speed settings have been corrected.
Fault Simulation Feature added, for
service purposes only.
MP&F maintenance support added.
New parameters and support added for
Device lifetime, Sensor lifetime,
maintenance required and calibration
required.
Corrected "inconsistent parameter
values" warning when tank height
causes a maximum volume less than
the max. volume parameter.
SILO2 SEARCH_BIG_X_MAX now
defaults to OFF (not user-editable).
Limited density value to reasonable
value. Large invalid values were
causing the LR 400 to reset.
Fixed issue regarding Volume/Mass
table entry deletion
Fixed issue using g/cm
Changes
3
density units
7ML19985JC02