Command
:PULS:DEL[1]:UNIT
Long
[:SOURce]:PULSe:DELay[1]:UNIT
Form
Set & Query
Parameter
S | SEC | PCT | DEG | RAD
*RST value
S
Description
Use this command to set/read the default units for the pulse delay
parameter. The default unit of a parameter is the unit used when the
parameter is programmed to a value without a unit suffix.
Example
To set the pulse delay to 50% of period:
:PULS:DEL:UNIT PCT
:PULS:DEL 50
Command
:PULS:DOUB[1]
Long
[:SOURce]:PULSe:DOUBle[1][:STATe]
Form
Set & Query
Parameter
OFF | ON
*RST value
OFF
Description
Use this command to switch double-pulse mode on or off. In double-
pulse mode two pulses are generated per pulse period. The delay
between the leading edges of the first and second pulse can be adjusted.
Programming Reference
SCPI Instrument Command List
Set OUTPUT delay unit to %
Set OUTPUT delay to 50% of period
59