SCPI Instrument Command List
Range coupling
Amplitude
Description
Use this command to program the offset current of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of current, you first have to
execute the [:SOURce]:HOLD CURRent command to enable the
:SOURce]:CURRent subsystem.
The available current range is limited by the combination of:
• Specified Voltage limits
• Actual OUTPUT Impedance setting :OUTPut:IMPedance
• Actual Expected Load impedance setting
Example
To program the offset current of the output signal:
:HOLD CURR
:CURR:OFF 50MA
Command
:CURR[1]:HIGH
Long
[:SOURce]:CURRent[1][:LEVel][:IMMediate]:HIGH
Form
Set & Query
Parameter
Numeric
Parameter suffix
A with engineering prefixes.
Value coupling
Range coupling
Low level
+10 mA (50 Ω into 50 Ω)
*RST value
Specified Limits
10 V Output (from high Z into short): –396 mA to 400 mA typical
3.8 V (from 50 Ω into short): –82 mA to 152 mA typical
Description
Use this command to program the high level current of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of current, you first
50
Amplitude = High – Low
High – Low
Offset =
Enable CURRENT subsystem
Set OUTPUT offset to 50 mA
2