Technical Alarms (INOPs)
Alarm Text
TELE CONFIG
UNSUPPORTED
Source - Patient Monitor
TELE DISCONNECTED
Sourece - Telemetry
TELE INCOMPATIBLE
Source - Patient Monitor
TELEMETRY STANDBY
5-18
Alarms
Draft - 1 Aug 08
Technical Alarms (Inoperative Conditions)
Priority
Condition
Hard
Transceiver not supported
(companion mode).
Red,
The interface cable is
Yellow or
not connected.
Hard
Short-range radio
Technical
connection between
Alarm
the transceiver and
the patient monitor
has been lost.
Hard
Transceivers equipped with
SRRA are not supported
with this revision of the
Information Center.
Soft
Information Center standby
mode timer is active, or
patient was not returned to
telemetry coverage area.
There is no data from bed.
What to do
Make sure the transceiver is
the correct revision.
Contact Service.
If the disconnection is not
intentional:
Re-connect the
interface cable.
Identify and remove
interference sources
such as bluetooth
devices, DECT
phones, microwaves,
etc.
Reduce the number of
devices equipped with
short-range radio
capability.
Check equipment
configuration/compatibility.
Contact Service.
Cancelled when patient is
removed from Standby.