2. On the Graphics Setup screen, select the type of chromatogram to save: Calibration (Cal),
Benchmark (Bench), most recent (Last), Alarm, or typical (Typ).
3. Next select the amount of screen offset, which determines where on the screen the
chromatogram will display. Start with "0" and adjust if necessary.
4. Then select the amount of screen attenuation, to ensure the entire chromatogram will appear on
the screen.
5. Determine whether the chromatogram needs to be compressed to fit by selecting No or Yes.
6. Set the chromatogram's start time ("0" to start at the beginning of the analysis).
7. Set the end time according to how much of the analysis you want to see.
8. Verify the selections and then press the F2 (Execute The Command) soft key. The
chromatogram will start to scroll on the display screen.
9. Press the F1 (Escape) soft key to return to the Display Saved Data screen.
Printing a Raw Data Report
1. On the Display Saved Data screen, cursor to REPORT TO PRINTER.
2. On the Send Report To Printer screen, select the type of data to be printed.
3. Next select whether you want calibration and benchmark data printed as raw data.
4. Verify the selections and press the F2 (Send Report To Printer) soft key. "The Command is being
updated" will appear on the message line and the raw data report will be printed. The report will
not appear on the display screen.
5. Press the F1 (Escape) soft key to return to the Display Saved Data screen.