10. Additional Notes
IEC 60601-1 classification
Type of Protection Against Electric
Shock
Degree of Protection Against
Electric Shock
Degree of Protection against
Ingress of Water
Mode of Operation
Electrical
AC Power Consumption
Fuses
USB Charging Port Output
Electrical for the Battery Pack
Run Time
Output voltage range
Battery technology
Capacity
Input voltage range
Output power (max continuous)
Minimum life cycle
Recharge time
46
Class II Equipment/Internally Powered
Type BF Applied Part
Device: Drip Proof, IP22
First characteristic numeral - 2 - Protection against
ingress of solid foreign objects ≥ 12.5 mm diameter.
Explanation: Protected against access to hazardous
parts with a finger and protected against solid
foreign objects of 12.5 mm diameter and greater.
Second characteristic numeral - 2- Protection
against ingress of water with harmful effects
dripping (15° tilted).
Explanation: Protected against vertically falling
water drops when enclosure tilted up to 15°.
Continuous
100–240 VAC, 50/60 Hz, 2.0-1.0 A
There are no user-replaceable fuses.
5 VDC, 7.5 W (1.5 A)
> 8 hours
18-24.6 VDC
Lithium Ion
62 Wh
100-240 VAC, 50/60 Hz, 2.0-1.0 A
50W
≥ 70% of rated capacity after 500 cycles
< 5 hours