Performance Assurance Checks
Procedure
Overview
See the following table for test values.
System Self-test Values
Module
Test Numeric
ECG and ECG/
RESP
ECG
100 bpm in ADULT mode
125bpm in NEO/PED mode
RESP
15 rpm in ADULT mode
30 nrpm in PEDI mode
55 rpm in NEO mode
Pressure
Systolic
120 mmHg ADULT
60 mmHg PEDI/NEO
Diastolic
0 mmHg ADULT
0 mmHg PEDI/NEO
SpO
/Pleth
2
SpO
100%
2
Pleth
Wave on display
13-14
Maintenance
1. Choose a patient category (Adult/Pedi/Neo) by pressing:
Monitor Setup
2. Display 30 seconds of the waveforms and numerics of all
connected modules by pressing:
Monitor Setup
3. Repeat Step 2 as necessary to check the individual test waveforms
and numerics for the category.
4. Use Step 1 to choose a new patient category and repeat Steps 2
and 3.
Config Sets
Change Pat Categ
Pts
Refernce
Test Signals
Pts
Refernce
Signals
Test Waveform
Simulated Square Wave
Simulated Wave
Simulated square wave
No Wave
Simulated wave
Pts
Refernce