Checking the Battery
Status (cont.)
Advanced Programming
When the battery voltage reads:
12.0 Volts - the battery is fully charged and will supply
approximately 480 minutes (8 hours) of power.
11.4 Volts - the MMS-3 displays the low battery message (refer to
Recharging the Battery Pack in Chapter 2 of the Service Manual).
10.9 Volts - the MMS-3 will shut down (refer to Recharging the
Battery Pack in Chapter 2 of the Service Manual).
Note: Continuous use of the backlight and alarms shortens the
battery life by 1-2 hours from the estimated battery life
displayed on the screen.
[YES]
Press
to proceed.
You may now do one of the following:
Recharge the battery as described in Recharging the Battery Pack
in Chapter 2 of the Service Manual.
Refer to another section and perform a different procedure. Refer
to Figure 3-2 on page 3-48 and Figure 3-3 on page 3-49 to navigate
through the user program.
[RUN]
Exit by pressing the
key.