Hitachi AJ-CE series Operation Manual page 42

High-precision tuning fork electronic balance
Table of Contents

Advertisement

(2)
Instruction for tare subtraction (zero adjustment)
C1
T
C2
(SP)
Code
54H
20H
Description
Instruction for tare
subtraction (zero
adjustment)
Value
None
Response
A00: Normal end
E01: Tare subtraction cannot be
executed due to an error in
the weight value.
(3)
Setup of output control
C1
o
o
o
o
o
o
o
o
o
o
C2
0
1
2
3
4
5
6
7
8
9
Code
4FH
4FH
4FH
4FH
4FH
4FH
4FH
4FH
4FH
4FH
30H
31H
32H
33H
34H
35H
36H
37H
38H
39H
Description
Stop output
Output continuous at all times
Output continuous if stable (stop output if unstable)
Outputs once by pressing | Memory | key (irrespective of
whether stable).
Outputs once if stable.
Outputs if the balance is stable when a
sample is loaded after the preceding sample has been removed
and the balance indicated zero, or less.
Outputs once if stable, and stops output when unstable.
Even
if the sample is not replaced, the balance is output once when it
stabilizes next time (including the zero indication).
Outputs once if stable, and outputs continuously when
unstable.
Even if the sample is not replaced, output of the
balance stops when it stabilizes after being output once.
Pressing | Memory | key causes the balance to output once
when stable.
Output once immediately.
Output once after stabilization.
output controls executed with commands [OO] - [07] work the same as the output controls
executed through function setup on the main unit of the balance.
The commands [08] and [09] are data request commands issued to the balance.
i&Once any command from [OO] to [09] is executed, the balance runs that function until another
command is entered.
However, if the balance is switched off and on again, the output control is
reset to the initial function (function set value).
10.5.3 Response Output
(1)
Response output format
Composed of five characters, including the terminators (CR = ODH; LF = OAH)
1
2
3
4
5
A1
A2
A3
CR
LF
(2)
Types
A1
i
A
i
E
j
of response outputs
A2
0
0
A3
0
1
Code
41H
45H
30H
30H
30H
31H
(Abnormal
Description
Normal end
Command error
command received, other errors)
38