Hitachi AJ-CE series Operation Manual page 32

High-precision tuning fork electronic balance
Table of Contents

Advertisement

Calibrating the Balance (cont.)
4
Calibrating the capacity point
\
/
o n
\
/-.
/
yCfi
L
Load the calibration weight in the center
of the balance.
The display flashes, indicating that
capacity point calibration is in progress.
5
Terminating the span adjustment
on F.S
I
N
' I
N
>
GAL
When the calibration of the capacity
point is finished, the original
measurement mode is restored.
0 - | - | - I F
EOO.no Gig
it Key Points of the Procedure it
1.
Pressing the | Function | key in Step 2 interrupts the span adjustment and returns you to the
original measurement mode.
2.
The calibration weight used for span adjustment should be heavier than half the capacity of the
balance.
To implement a calibration as precisely as possible, use a weight close to the capacity of the
balance.
•& Calibration weights can be ordered from Shinko.
For ordering information, please contact
Shinko.
3.
If problems arise during span adjustments, one or more of the following error messages will
appear:
(1)
[o - E r r ]: The calibration weight exceeds the capacity of the balance.
(2)
[ i - E r r ]: The calibration weight is less than half the capacity of the balance.
(3)
[£' - E r r ]: The difference between before and after calibration values is too large (1.0%
or more).
2<;lf error messages are displayed, calibration cannot take place.
Check the weight and re-calibrate.
If the same error continues after repeated calibrations using
the correct weight, please contact our Marketing Division or Technical Service Division.
28

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ajh-ce series