2 Safety instructions

2.3 Emergency procedure

2.3

Emergency procedure

This section describes how to do an emergency shutdown of the Typhoon FLA 7000 in-
strument. The section also describes the results of a power failure.
Emergency shutdown
Step
1
2
3
Power failure
The results of a power failure depends on the unit or units affected.
Unit affected by power fail-
ure
Typhoon FLA 7000 instrument
Computer running the control
software
30
Action
Click the Stop button in the Typhoon FLA 7000.
Turn off the Typhoon FLA 7000 by pressing the power switch on the left side
of the instrument body.
Disconnect the power cord from the power outlet.
Results
The reading is interrupted immediately. The in-
strument is in an undefined state.
The data collected up to the time of the power
failure is available in the file created when
starting the scan.
The computer shuts down immediately.
The run continues, but no data is saved.
Getting Started with Typhoon FLA 7000 28-9607-64 AD