Mr. Anderson
March 9, 2025
The TensCare TouchTENS XL-Y1 device may not turn on even with new batteries properly installed due to the following reasons:
1. Battery Polarity: The batteries may not be inserted in the correct orientation. Ensure they are placed as shown in the battery compartment.
2. Battery Leakage: If there was previous battery leakage, it could have damaged the contacts. Check for any corrosion or residue in the battery compartment.
3. Battery Type: The device requires alkaline batteries. Using rechargeable or incompatible batteries may prevent it from working.
4. Low Battery Level: Even if the batteries are new, they may not be fully charged or may not provide enough power.
5. Battery Cover Not Locked: If the battery cover is not properly locked, the device may not make proper contact with the batteries.
If the issue persists after checking these factors, the device may have a malfunction.
Touch TENS Instructions for Use Section Page No Introduction How TENS works Side Effects Contents Installation of Batteries Belt clip The Programmes Programme Settings Controls and Display Controls Connecting Lead wires Positioning of Electrode Pads General Pad Advice Cleaning Troubleshooting Caution and Warnings Consumables and servicing WEEE...
Sections 9 & 10. reach a maximum level which can last for hours after the machine is switched off. The Tenscare Touch TENS is a dual channel TENS. It has: By using TENS you can expect to achieve a significant reduction in pain if not complete pain relief.
Warning 5. INSTALLATION OF BATTERIES • Keep batteries out of the reach of small children. Remove battery tray from side of unit • If battery leakage occurs and comes in contact with the skin or eyes, wash thoroughly with lots of water. •...
The following grid summarises the programmes available and USAGE how they vary. TENS can be used for as long as is necessary. Continuous treatment is fine, but electrode pads should be repositioned Output Sensation Specific Uses Frequ- Pulse ency Hz Width regularly (at least every 12 hours) to allow the skin to be exposed µS...
9. CONTROLS AND DISPLAY 10. CONTROLS ON/OFF To turn the unit ON, press the ON/OFF button on Programme Channel 1 the side of the unit and hold for 2 secs until the selected Strength display shows. To turn the unit OFF, press the ON/OFF button on Channel 2 Keypad Lock the side of the unit and hold for 2 seconds until the...
Page 7
The unit has 60 levels of strength. If you hold down the button for You can set session times of 20,40, or 60 minutes by pressing 2 seconds, the strength will start scrolling. the green button repeatedly. You may feel nothing over the first few presses. Continue pressing The LCD shows the session duration next to the clock symbol.
12. POSITIONING OF ELECTRODE PADS 3) Elapsed Time. Press Programme and Right blue ▲ together and hold down for 3 seconds. The display will show U (for Usage) The following body maps in this instruction booklet will show and a number from 1 to 9999 hours, indicating the total number you where to place the electrode pads for a range of common of hours the unit has been above zero strength.
ELECTRODE PAD POSITIONS 13. GENERAL PAD ADVICE • The electrode pads supplied are reusable but are for single Low Back Pain Neck and Sciatica patient use. Shoulder Tension • In order to obtain the best conductivity through the pads always ensure that they are in good condition and tacky.
(electrosurgery/microwave report unexpected operation or events, call TensCare or your cookers/mobile phones) may affect the correct operation of this local dealer (address on back cover) for advice. unit - See section 21. If it appears to behave unusually, move it away from these devices.
Page 11
Effect on patients who have had strokes or seizures is not known. May affect sense of balance contact details), by mail order from TensCare, by telephone using The effects of stimulation of the brain are unknown a credit or debit card, or through our website.
Your TensCare device is guaranteed for two years from the date applied part. of purchase. If a fault develops return the unit to TensCare at the This symbol on the unit means “ Refer to address below, together with a copy of your invoice and details Instructions for use”...
21. EMC PRECAUTIONS Use special precautions regarding EMC according to the information provided below. Other portable and mobile RF communications equipment can The Touch TENS is intended for use in the electromagnetic affect performance. Do not use when adjacent to or stacked with environment specified below.
Page 14
EC Declaration of Conformity communications equipment. Rated maximum Separation distance according to frequency of transmitter M TensCare Ltd hereby declare that an examination output power of of the production quality assurance system has transmitter W been...
Need help?
Do you have a question about the TouchTENS XL-Y1 and is the answer not in the manual?
Questions and answers
שלום. אני מעונין לקנות את המכשיר
המכשיר לא נדלק למרות שהוכנסו סוללות חדשות, והן מונחות במקום הנכון.
The TensCare TouchTENS XL-Y1 device may not turn on even with new batteries properly installed due to the following reasons:
1. Battery Polarity: The batteries may not be inserted in the correct orientation. Ensure they are placed as shown in the battery compartment.
2. Battery Leakage: If there was previous battery leakage, it could have damaged the contacts. Check for any corrosion or residue in the battery compartment.
3. Battery Type: The device requires alkaline batteries. Using rechargeable or incompatible batteries may prevent it from working.
4. Low Battery Level: Even if the batteries are new, they may not be fully charged or may not provide enough power.
5. Battery Cover Not Locked: If the battery cover is not properly locked, the device may not make proper contact with the batteries.
If the issue persists after checking these factors, the device may have a malfunction.
This answer is automatically generated