Omron HEM-6111 Instruction Manual

Omron HEM-6111 Instruction Manual

Automatic wrist blood pressure monitor

Advertisement

Quick Links

2. Preparation
Instruction Manual
2.1 Installing/Replacing the Batteries
1.
Remove the battery compartment cover
Automatic Wrist
by pulling it off in the direction of
Blood Pressure Monitor
the arrow.
1) Grasp both ends of the battery cover.
HEM-6111
Model
2) Pull the cover off the main unit.
2.
Install or replace two "AAA" size
batteries so that the + (positive)
and - (negative) polarities match
1667074-4D
the polarities indicated on the battery
Introduction
compartment.
Thank you for purchasing the OMRON HEM-6111 Automatic Wrist Blood
3.
Put the battery cover back in place.
Pressure Monitor.
Note: The measurement values continue to be stored in memory even
after the batteries are replaced.
The OMRON HEM-6111 is compact and easy-to-use at home, at work, and
during travel. Perfect for people who frequently monitor their own blood pressure.
2.2 Battery Life & Replacement
The unit also stores up to 30 measurement results in the memory.
If the battery low symbol (
replace both batteries at the same time.
Please read this instruction manual thoroughly before using the unit. For specific
information about your own blood pressure, CONSULT YOUR DOCTOR.
- When the battery low symbol (
the unit for a short while. You should replace the batteries with new ones
Important Safety Information
ahead of time.
- When the symbol (
) remains lit, the batteries are exhausted. You should
Consult your doctor during pregnancy, arrhythmia and arteriosclerosis.
replace the batteries with new ones at once. Turn the unit off before replacing
You should never change the dosage of medication prescribed by your doctor.
the batteries.
Use the HEM-6111 only for yourself.
• Remove the batteries if the unit will not be used for three months or more.
• Dispose of batteries according to applicable local regulations.
Caution:
• Do not leave the unit unattended with infants or persons who cannot express
Two new "AAA" alkaline batteries will last for approximately 300 measurements,
their consent.
when used to take two measurements a day.
• People with severe blood flow problems, or blood disorders, should consult
Since the supplied batteries are for monitoring use only, they may have
a doctor before using the unit. Cuff inflation may cause internal bleeding.
a shorter life and not last for 300 measurements.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not use a cellular phone near the unit. This may result in incorrect
operation of the unit.
3. Using the Unit
• Do not operate the unit in a moving vehicle (car, airplane).
• Do not apply strong shocks and vibrations to or drop the unit.
3.1 Applying the Wrist Cuff
• Please observe the following points.
You can take a measurement on either your left or right wrist.
- Do not insert the batteries with their polarities incorrectly aligned.
- Replace old batteries with new ones immediately.
Notes:
- When you are not going to use the unit for a long period of time
• To ensure correct measurement, apply the wrist cuff so that it fits comfortably
(approximately three months or more), take out the batteries.
around your wrist.
- Do not use new and used batteries together.
• Roll up your sleeve so that the unit covers bare skin.
- Do not use batteries of a different type together.
• Do not apply over clothing.
- Only use two "AAA" alkaline batteries with this unit. Do not use other types of
batteries.
1.
Place the wrist cuff over your left wrist with your left thumb
- When using batteries, the battery liquid may leak and damage the main unit.
facing upward.
• Do not inflate the wrist cuff when it is not wrapped around your wrist.
1. Overview
Main unit:
A
C
Note: Make sure that the wrist cuff does not cover the protruding part of
D
E
the wrist bone (ulna) on the outside of the wrist.
B
2.
Hold the bottom part of the wrist
cuff and wrap it around the wrist
while pulling so that it fits
comfortably.
A. Display
C. Wrist cuff
B. Battery compartment
D. START/STOP button
E. MEM (Memory) button
Note: Unless the wrist cuff is wrapped securely around the wrist, it may
Package contents:
not be possible to take correct measurements.
F . Two "AAA" alkaline (LR03)
Smooth the cuff with your fingers
batteries
to ensure that it fits
G. Storage case
snugly around the wrist.
• Instruction manual
F
G
• Blood pressure monitor unit
Notes:
Display:
• If the fastener is not wrapped securely
H
around your wrist, the cuff may slip
during measurement.
I
• If you have a slender wrist, the cuff may
L
feel slightly loose.
J
M
However, this will not affect measurement
N
K
results and you should not force the cuff to fit.
• The remaining part of the wrist cuff can be
H. Systolic blood pressure
L. Heartbeat symbol
conveniently folded back out of the way.
I. Diastolic blood pressure
1. Flashes during measurement
J. Memory symbol
2. If flashing after measurement
(Displayed when viewing
complete indicates blood pressure
values stored in memory)
out of recommended range
K. Battery low symbol
M. Pulse display
• If the remaining part of the cuff is difficult
N. Deflation symbol
to fold back, measurement results will not
be effected.
Note: If your systolic or diastolic
pressure is outside the
standard range
(above 135/85 mmHg),
the Heartbeat symbol (
)
Taking a reading on the right wrist
will blink. Refer to
"3.3 Taking a Reading".
Measurements can also be made on
the right wrist.
Fit the unit on the right wrist as shown.
3.2 Correct Posture
Correct posture during measurement helps you get accurate results.
You should also try to measure your blood pressure at the same time each day.
Notes:
• Do not take measurement after bathing, drinking alcohol, or exercising.
2
• Do not move or talk during measurement.
Sit comfortably with your
Position your arm so that
1
back straight, hold your arm
the unit is at the same
across your chest and relax.
level as your heart
(height of a nipple).
Do not use your other
hand to support the wrist cuff.
) appears on the display,
This can result in inaccurate
measurement results.
Relax your wrist and hand.
) starts to blink, you will still be able to use
Do not bend
Do not clench
your wrist back.
your fist.
3.3 Taking a Reading
1.
Press the START/STOP button to start measurement.
The wrist cuff will start to inflate automatically after the
power is turned on.
Note: To cancel a measurement, press the START/STOP button to turn off
the unit and to release the air in the wrist cuff.
2.
After the unit has detected your blood pressure and
pulse rate, the cuff automatically deflates and your blood
pressure and pulse rate are displayed.
3.
Check the measurement results.
The unit automatically stores blood
Systolic blood
ulna
pressure
pressure and pulse rate into its memory.
Diastolic blood
Refer to "3.4 Using the Memory
pressure
Function".
Pulse display
Notes:
• Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
• Wait 2-3 minutes before taking another blood pressure reading.
Waiting between readings allows the arteries to return to the condition prior
to taking the blood pressure measurement.
Important:
• If your systolic or diastolic pressure is outside
the standard range, the heartbeat symbol will
blink when the measurement result is displayed.
Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure
Above 85 mmHg
This criteria is for home blood pressure measurement.
4.
Undo the fastener and remove the unit.
5.
Press the START/STOP button to turn the unit off.
Note: If you forget to turn the unit off, it will automatically shut itself off
after two minutes.
3.4 Using the Memory Function
This unit has a memory capable of storing 30 sets of readings. Every
time you complete the measurement, the unit automatically stores the
blood pressure and pulse rate.
Note: When 30 sets of readings are stored in memory, the oldest set will be
deleted to store a new set.
1.
Press the MEM button to view readings stored in
memory from the most recent to the oldest.
The Memory number appears for a second before the
pulse rate is displayed.
The newest set is numbered "1".
Important:
If your systolic or diastolic pressure is outside
the standard range, the heartbeat symbol will
blink when the measurement result is displayed.
Refer to "3.3 Taking a Reading".
Note: If there are no measurements results
stored in memory, the screen to the right
is displayed.
2.
Press the MEM button repeatedly
to cycle through the previous
measurement results. The Memory
number appears for a second before
the pulse rate is displayed.
The newest set is numbered "1".
3.
Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after two minutes.
To Delete All the Values Stored in Memory
You cannot delete individual stored readings, all the
readings in the unit will be deleted.
1.
To delete stored readings, first press the MEM button.
Then while holding it down, press the START/STOP
button simultaneously for about 2-3 seconds. All readings
will then be deleted.
2
1
2.
Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after two minutes.
Do not bend
your wrist forward.
4. Handling Errors and Problems
4.1 Error Messages
Error Display
Cause
Remedy
Repeat measurement.
Cuff over inflated.
Remain still and do not move
during measurement.
(Refer to "3.3 Taking a
Reading".)
Movement during
measurement.
The wrist cuff
Carefully read and repeat
is not fastened
the steps listed under
securely.
"3.1 Applying the Wrist Cuff".
An Er mark with
a code/number
Consult your OMRON retail
indicates the
outlet or distributor.
device has a
hardware failure.
Replace the batteries with two
The battery
new "AAA" alkaline (LR03)
power is low.
batteries.
Blinks or appears
continuously
4.2 Troubleshooting
Problem
Cause
Remedy
Measure while in
Are you holding
the correct posture.
the wrist cuff at
(Refer to "3.2 Correct
heart level?
Posture".)
Is the cuff
Wrap the cuff correctly.
wrapped snugly
(Refer to "3.1 Applying
The reading is
around the wrist?
the Wrist Cuff".)
extremely low (or
Relax and try taking
high).
Are your arms and
the measurement again.
shoulders tense?
(Refer to "3.3 Taking
a Reading".)
Remain still and do not
Movement or
talk during measurement.
talking during
(Refer to "3.3 Taking
measurement.
a Reading".)
Wrist cuff pressure
Air is leaking from
Consult your OMRON
does not rise.
the wrist cuff.
retail outlet or distributor.
Apply the cuff correctly
so that it is securely
Wrist cuff deflates
The wrist cuff is loose.
wrapped around the wrist.
too soon.
(Refer to "3.1 Applying
the Wrist Cuff".)
Blood pressure readings
constantly vary with
time of measurement
The blood pressure is different each time.
and nervous condition.
The reading is extremely low (or high).
Take deep breaths to
relax before taking a
measurement.
The unit loses power
The batteries are
Replace the batteries
during measurement.
empty.
with new ones.
The batteries are
Replace the batteries
Nothing happens
empty.
with new ones.
when you press
The batteries have
Insert the batteries
the buttons.
been inserted
with the correct (+/-)
incorrectly.
polarity.
• Press the START/STOP button and repeat
measurement.
• If the problem continues, try replacing
Other problems.
the batteries with new ones.
If this still does not solve the problem,
contact your OMRON retail outlet or distributor.
5. Maintenance and Storage
Important information regarding Electro Magnetic Compatibility (EMC)
Maintenance
With the increased number of electronic devices such as PC's and mobile
(cellular) telephones, medical devices in use may be susceptible to
To protect your unit from damage, please observe the following:
electromagnetic interference from other devices.
Electromagnetic interference may result in incorrect operation of the medical
• Do not subject the main unit and cuff to extreme
device and create a potentially unsafe situation.
temperatures, humidity, moisture or direct sunlight.
Medical devices should also not interfere with other devices.
• Do not inflate the wrist cuff over 299 mmHg.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
• Do not disassemble the unit.
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
• Do not subject the unit to strong shocks or vibrations
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
(for example, dropping the unit on the floor).
emissions for medical devices.
• Do not use volatile liquids to clean the main unit.
The unit should be cleaned with a soft, dry cloth.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
• Do not use petrol, thinners or similar solvents to clean
Nevertheless, special precautions need to be observed:
the wrist cuff.
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct
operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
• Do not carry out repairs of any kind by yourself.
of its working life. To prevent possible harm to the environment or
If a defect occurs, consult your OMRON distributor.
human health from uncontrolled waste disposal, please separate
this from other types of wastes and recycle it responsibly to
Calibration and Service
promote the sustainable reuse of material resources.
• The accuracy of this blood pressure monitor has been
Household users should contact either the retailer where they purchased this
carefully tested and is designed for a long service life.
product, or their local government office, for details of where and how they can
• It is generally recommended to have the unit inspected
take this item for environmentally safe recycling.
every two years to ensure correct functioning and
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
accuracy. Please consult your authorised OMRON
wastes for disposal.
dealer.
This product does not contain any hazardous substances.
• Do not carry out any repairs yourself. If a defect occurs
or you have doubts about the correct functioning of the
device, consult your authorised OMRON dealer.
Storage
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity,
direct sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will
be at an angle.
Asia Pacific HQ
• Locations exposed to chemicals or corrosive vapours.
Production Facility
6. Technical Data
Product Description
Automatic Wrist Blood Pressure Monitor
Model
HEM-6111
Display
LCD Digital Display
Measurement Method
Oscillometric method
Measurement Range
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180 beats/min.
Memory
Memory 30 Measurements
Accuracy
Accuracy Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation
Automatic by electric pump
Deflation
Automatic pressure release valve
Power Source
2 "AAA" batteries 1.5V
Battery life
Capacity of new alkaline batteries is approximately
300 measurements
Applied Part
= Type B
Protection Against
Internally powered ME equipment
Electric Shock
Operating
10°C to 40°C/ 30 to 85% RH
temperature/ Humidity
Storage temperature/
-20°C to 60°C/ 10 to 95% RH/ 700-1060 hPa
Humidity/ Air Pressure
Console Weight
Approximately 117 g without batteries
Outer Dimensions
Approximately 71 (w) mm × 41 (h) mm × 70 (l) mm
Measurable
Approximately 13.5 to 21.5 cm
circumference of wrist
Package Content
Main unit, Storage case, Instruction manual, Battery
set
Notes:
• Subject to technical modification without prior notice.
• Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.
Please read this instruction manual carefully before using the device.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
HEM-6111-C1_C_M02_100309.pdf
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 Japan
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08, Alexandra Technopark,
Singapore 119967
OMRON(DALIAN) CO., LTD.
Dalian, China

Advertisement

loading
Need help?

Need help?

Do you have a question about the HEM-6111 and is the answer not in the manual?

Questions and answers

Summary of Contents for Omron HEM-6111

  • Page 1 Recommendation is to keep a minimum distance of 7 m. Verify correct after the batteries are replaced. The OMRON HEM-6111 is compact and easy-to-use at home, at work, and button simultaneously for about 2-3 seconds. All readings operation of the device in case the distance is shorter.
  • Page 2 OMRON HEALTHCARE EUROPE B.V. • Kruisweg 577, 2132 NA Hoofddorp THE NETHERLANDS “1” • • • 299 mmHg START/STOP OMRON HEALTHCARE SINGAPORE PTE LTD. HEM-6111 • Asia Pacific HQ 438A Alexandra Road, #05-05/08, Alexandra Technopark, • Singapore 119967 “AAA” •...