Page 2
The information in this publication has been carefully checked and is believed to be entirely accurate at the time of publication. HUVITZ assumes no responsibility, however, for possible errors o r omissions, or for any consequences resulting from the use o f the ormatio n contained herein.
Introduction 1.1. Indication for use (intended use) The SLIT LAMP, HS-5500 is intended for use in eye examination of the ant segments, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment.
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 Safety Information 2.1. Introduction Safety is everyone’s responsibility. The safe use of this equipment is largely dependent upon the installer, user, operator, and maintainer. It is imperative that personnel study and become familiar with this entire manual before attempting to install use, clean, service or adjust this equipment and any associated accessories.
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- 2.2. Safety Symbols The International Electro technical Commission (IEC) has established a set of symbols for medical electronic equipment which classify a connection or warn of any potential hazards. The classifications and symbols are shown below. This applies only to the instrument that has the certification symbol printed explicitly on the product label or sticker.
Page 9
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 M EDIC AL EUIP M ENT W ITH RESP EC T TO ELEC TRIC SHOCK UL6 06 01 FIRE M E CH ANIC AL H AZ ARDS CAN/CS A C22 .2 ONLY IN ACCO RD ANCE W ITH U L 6 0601- NO.601 .1...
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- 2.3. Environment factors Avoid the following environments for operation or storage: W here the equipment is expo sed to water vapor. ’ t operate equipment with a wet hand. W here the equipment is expo sed to direct sunlight. W here the temperature changes extremely °...
Page 11
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 W here equipment is exposed to chemical material or explosive gas. Be careful not to be inserted dust, especially, metal. ’ disassemble the pro duct or o pen. We aren ’ respo nsible for it fo r nothing.
The equipment must be operated only by, or under direct supervision of a properly trained and qualified person. Modifications of this equipment may only be carried out by Huvitz’s service technicians or other authorized persons. Customer maintenance of this equipment may only be performed as stated in the User’s Manual and Service Manual.
Page 13
The manufacturer is not liable for damage caused by unauthorized tampering with the device(s). Such tampering will forfeit any rights to claim under warranty. This equipment may only be used together with accessories supplied by Huvitz’s. If the customer makes use of other accessories, use them only if their safe usability under technical safety aspects has been proved and confirmed by Huvitz or the manufacturer of the accessory.
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- Features & Operating Condition 3.1. Brief Description The device works like all slit lamps: Stereomicroscopic examination of the eye under the slit light. Microscopic examination of the ocular fundus and posterior vitreou body (using hruby’s lens). These devices are designed for use by ophthalmologists and optometrists (within the realms of their respective professions) for specific diagnostic procedures (bio- microscopic examination of the eye)
Page 15
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 CAUTION Because prolo nged intense light exposure can damage the retina, the use of the device for ocular examinatio n sho uld no t be unnecessarily prolonged, and the brightness setting sho uld not exceed what is needed to pro vide cl visualizatio n o f the target structures.
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- Standard Components The instrument is delivered in package. Taking it out the package, make sure that none of the following parts is missing. A table top(N.B. the table top isn’t supplied with the slit lamp for twin table) on which are placed: A SMPS box(15) including the main switch &...
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 Accessories The following standard accessories are provided with the instrument; protection cover(46) a setting rod(43) spare bulbs 12V 30W Halogen bulb (53) chin-rest paper(47) spare fuses 250Vac T2.0AL a breath screen(5) power cable [ F i g u r e 1 . Ac c e sso r i e s ]...
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- Composition Fixation point Control lever for changing enlargement Removable eye-pieces Knob for clamping the breath screen Breath screen Knobs to change the slit length Connecting plug of the projector lamp Knurled ring for filters insertion Joystick for lateral, longitudinal and vertical movements(x y z) 10) Base with orthogonal movements 11) Shaped tabletop...
Page 19
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 26) Locking pins for chin rest paper 27) Chin rest 28) Knobs to change the slit width 29) Pointer for the eye positioning 30) Forehead Rest 31) Microscope Lamp socket 33) Knob to lock the base of the instrument...
Page 20
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- 54) Screw hole cover 55) Prism 56) Capture connector ( For the image system) 57) Capture switch 58) Luminosity selector connector ( Connect the Luminosity selector to the SMPS) 59) Head rest module 60) Diffuser...
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 Installation NOTE The appliances are always supplied with excellent packaging to be able to stand up to all kinds of transport and standard sto rage co nditions. P lease co ntact the suppliers if any irregularities are fo und in the appliance due to transport after unpacking 7.1.
Page 24
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- Place the head rest module(59) and make sure that its holes fit together with Those of the table top. Tighten the screw with the screw driver [ Figure 5. Assembling the instrument I] Place the base with orthogonal movements on the guides situated on the upper part of the table top, checking that the wheels are in line.
Page 25
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 Assemble an optical group of the slit projector (49) on the Base with orthogonal movements (10). Then tighten the Guide screw with the screw driver(-) Put the microscope(31) in the right position making sure that it is placed well in the stop(37);...
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 Trial Operations Assemble the chin-rest paper (47) on the chin-rest (27). Let the patient sit down comfortable with his chin on the chin-rest and his forehead against the forehead-rest. Lift or lower the chin-rest by moving the handle (25) so that the patient’s eyes are in line with the notches on the chin-rest (29).
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- Main Features Principal operative characteristics of the instrument: Observation without settling or tiring and rapid change of enlargements. Base with cross-wise movement (which can be stopped) commanded by a single joystick for all adjustments(x y z). Stereomicroscopical observation of the eye with enlargements 6x-10x-16x-25x- 40x(when use 12.5x eyepieces ) ¢ª...
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 10. Ordinary Maintenance WARNING The repairs illustrated below should all be carried out with the power cable disconnected from the main supply plug. P lease contact the suppliers for any other pro blems that cannot be resolved by the operatio ns des ibed below.
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- Fasten the bulb with the Bulb fastening knob Make sure that the stems of the bulb are completely inserted; insert the plug in the main supply socket [ F ig ure 7. Re p la c e Ha lo g e n b ulb ] WARNING Do n ’...
Page 31
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 The protection fuses are placed on the back of the SMPS, and inserted power socket. Before doing anything else, disconnect the plu from the main supply socket Remove the burned-out fuses Substitute the fuses and check Replace the plug in the main supply socket.
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- 10.3. Cleaning & Disinfection To keep the device clean, apply clinical alcohol and cidex or cidex plus on a soft cloth and scrub the body and the probe once a month. Do not lacquer, thinner, ethylene, or the oxidizing substance.
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 11. EMC Information 11.1. Guidance and manufacturer’s declaration – electromagnetic emissions The EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT should assure that it is used in such an environment.
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- 11.2. Guidance and manufacturer’s declaration – electromagnetic immunity The EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT should assure that it is used in such an environment.
Page 35
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 <5% U? U? ) EUT image intensifier be (<95% dip in for 5 s Powered from U? ) uninterruptible power for 5 s supply or a battery. Power frequency 3 A/m 3 A/m Power frequency (50/60Hz)
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- 11.3. Guidance and manufacturer’s declaration – electromagnetic immunity The EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT should assure that it is used in such an environment.
Page 37
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as deter-mined by an electromagnetic site survey, Should be less than the compliance level in each frequency range. Interference may occur in the vicinity of...
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- 11.4. Recommended separation distances between portable and mobile RF communications equipment and the EUT This is intended for use in an electromagnetic environment which radiated RF disturbances are controlled. The customer or the user the EUT can help Prevent electromagnetic interference by maintaining a minimum nce between portable and mobile RF communications equipment (transmitters)
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- 11.6. Guidance and manufacturer’s declaration – electromagnetic immunity The EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT should assure that it is used in such an electromagnetic environment.
Page 41
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 NOTE 1) These guidelines may not apply in all situatio Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. NOTE 2) It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to assure that they meet the minimum specification.
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 13. Technical Specifications M IC R O SC O P E - H S-55005x T ype G alilean conv erging bin ocular M agnification 5 p osition s rotatingdrum E y epiece 12.5x T otal m ag nification...
Page 44
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- BASE Vertical Movement Min. 28mm Longitudinal Movement Min. 78mm. Lateral Movement Min. 98mm Fine base Movement Min. 10mm Power supply 100~ 24 0 V AC 50/ 60HZ 2.0A Power consumption 70VA Instrument Voltage 12V DC Halogen Projection bulb 12V 30W Fixation point LED...
Seri al No .: ’ If you can t contact with your local distributo r, you can directly get in touch with the service department of the HUVITZ using the pho ne number and the address written in the below table.
Page 46
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- How to Co ntact HUVITZ Co ., Ltd Wri te us at: : +82-31 442-8868 HUVITZ Co ., Ltd. F ax : +82 31-442-8619 Huvitz B/D, 68 9 3 Geumjeong dong URL http://www.huvi tz.com Gunpo...
------------------------------------------------------ -----------------------------SLIT LAMP HS-5500 15. Particular Information 15.1. Relative spectral output graph (1) Halogen Lamp (2) Fixat ion Lamp...
Page 48
SLIT LAMP HS- 5500------------------- ---------------------------------------------------------------- Note: this graph be shown for the beam after it exists the instrument, when the instrument is operating at maximum light intensity and maximum aperture. CAUTION “CAUTION – The light emitted from this instrument is potentially hazardous.