In case of alterations of the device or the additional equipment; the latest version of the manual or the corresponding additional information always should be considered. The latest manual version is always available in PDF version on the h/p/cosmos website: http://www.h-p-cosmos.com/en/company/downloads.htm...
Page 3
All kinds of installation and repair work and most maintenance work are to be performed only by trained and authorized technicians who have been certified by h/p/cosmos. The following symbols will indicate which work can be done by the customer and which work has to be done only by authorized technicians: The customer/user should perform this maintenance and monitoring work.
Operation ..................................24 General ................................... 24 Switch on procedure ............................... 26 5.2.1 Switch on procedure model range 150/50 up to 190/65 ..................26 5.2.2 Switch on procedure model range 200/75 up to 450/300 (venus/saturn) ............27 Switch off procedure ............................... 28 5.3.1 Switch off procedure model range 150/50 up to 190/65 ..................
Page 8
12.5 Handrails, adjustable: Parallel bars with gas struts and locking lever ................162 12.6 Therapist’s seat and foot support, adjustable ....................... 164 12.7 Unweighting system h/p/cosmos airwalk se 135 ......................165 12.8 Unweighting systems airwalk 70/160 ........................... 166 12.9 Vests for subjects for airwalk systems.......................... 166 12.10 h/p/cosmos robowalk...
A running machine (also called “treadmill ergometer”) is an exercise device for running or walking while staying in one place. The h/p/cosmos running machines are stationary treadmills which are used for training and testing, simulating various adjustable speeds and various adjustable elevations (also called “slope” or “inclination”). The medical versions can also be used for diagnostics and therapy applications.
Unweighting systems as an accessory (for example the h/p/cosmos airwalk) can be used for body weight support of the patient, in case the patient should not bear his full body weight on his legs.
Page 13
It can, for example, be duplicated on the opposite handrail. h/p/cosmos conforms with the norm EN 957-6: “The push-button type or pull cord safety stop actuator shall be in a reachable position in front of the user and in an area of ± 180 mm parallel to the centre line of the running surface. If the push-button type safety stop actuator is outside this area, it shall be duplicated on the opposite side except when a pull- cord actuator is provided.”...
The medical treadmills may be operated only in medical facilities by medical staff. The safety requirements in particular requirements for medical electrical equipment and medical electrical systems for medical care in a home environment (IEC 60601-1-11) - are not considered in the medical device, h/p/cosmos treadmill ergometer.
Page 17
Recommendation and application of weight support system and/or fall stop device, such as the h/p/cosmos safety arch with harness and automatic shut-off of the running belt. Comprehensive instruction of the patient before the start of treadmill therapy with an explanation that the drive of the running belt is very strong and does not stop automatically when the patient stops.
A safety arch with chest belt and fall stop is also obligatory in this case. h/p/cosmos devices must not be used in environmental conditions other than those specified in the chapter „Technical specifications“ and “Environmental conditions” (e.g. in wet and humid areas, swimming pools, sauna, environmental chambers, high-pressure-, low-pressure, altitude- and oxygen-chambers, etc.).
Any other use than that explicitly listed in the field of application is prohibited. Use of the h/p/cosmos medical treadmills in a home environment and/or in public and/or operation by persons who are not medically trained staff is not allowed.
Additionally h/p/cosmos recommends the use of adjustable handrails. When using the treadmill for load tests, a medical doctor and a defibrillator are to be within reach at any time.
Page 22
If the treadmill is being controlled via ECG, pay special attention to the ECG manufacturer’s manual! The summary of clinical data and the risk management of h/p/cosmos does NOT cover any measured parameters and/or vital functions which are measured and/or detected by host devices (e.g. ECG, ergospirometry devices, EMG, etc.) which are connected to the treadmill.
Setting OP 41 to value 1 affects the safety of the subject and third parties. A down movement of the elevation after pressing STOP can result in confusion leading to serious injuries. h/p/cosmos strongly recommends keeping the standard setting of this option. Stop means stop for all movements and should not activate the running machine in any way.
5.2 Switch on procedure 5.2.1 Switch on procedure model range 150/50 up to 190/65 If the running machine is at an angle of elevation > 0 when switched on, it will automatically move to the position zero (display elevation „INIT“). Please ensure that while lowering the elevation no harm is done to persons or objects under the running machine.
For models built before the year 2012 the green lamp extinguishes after the switch on delay of 1 minute. 5.3 Switch off procedure 5.3.1 Switch off procedure model range 150/50 up to 190/65 Illustration Description [01] Switch off the running machine by pressing the black or red “OFF”...
The emergency stop button is to be pressed immediately! h/p/cosmos is in compliance with the norm EN 957-6 and offers more safety features than required, as shown in the following tables. In all below mentioned cases, all displayed values will be reset to “0”.
More recent computers will only have a USB interface instead of the RS232 interface. In that case, a „USB to RS232 interface adapter“ is available at h/p/cosmos under order number [cos12769-01]. For control via USB interface the processor must be a Pentium 1.8 GHz or higher. To use all functions described below and for maintenance and service, we recommend the PC software h/p/cosmos para control (Freeware).
After booting up, h/p/cosmos para control will start automatically and show these safety instructions. Before the running machine can be used, the safety instructions have to be read and confirmed with a click on the respective confirm button: “I fully understand and accept”.
Page 37
To make the software as user-friendly as possible it has been designed in exactly the same style as the physical UserTerminal of the h/p/cosmos running machine. The great advantage is that everyone who is familiar with the running machine is also able to control the machine with h/p/cosmos para control.
® ® Kistler force measurement plates. It is the only h/p/cosmos treadmill with these features. The system measures the vertical ground reaction forces and thereby collects a large amount of force and time-based gait parameters. The total force path can be displayed with the Kistler Gaitway software.
Page 39
Pathologic gait pattern, arthritis in ankles Femoropatellar syndrome Knee joint problems Anterior cruciate ligament reconstruction The following figures show four examples of the clinical benefit of h/p/cosmos gaitway with KISTLER force plates: ® Arthritis in the upper ankle 4 weeks later...
The acceleration levels are accessible by pressing the "speed-keys" + and minus several times and then holding the key. The acceleration levels are also accessible via interface and optional PC software h/p/cosmos para control and h/p/cosmos para graphics ®...
5.9.2 Manual mode The following table shows the operating steps for running the treadmill via firmware or software h/p/cosmos para control. Operation of profile mode Initiation: Running belt is not moving. One of the modes LED is flashing: (manual, profile, cardio, test) Step no.
Page 47
The following table shows the operating steps for running the treadmill via firmware or software h/p/cosmos para control ® Operation of profile mode Initiation: Running belt is not moving. One of the modes LED is flashing: (manual, profile, cardio, test) Step no.
Page 50
15.1 ... 30 30.1 ... 50 > 50 The following table show the operating steps to run the treadmill via firmware or software h/p/cosmos para control ® Operation of cardio mode Initiation: Running-Belt is not moving. One of the modes LED is flashing: (manual, profile, cardio, test) Step no.
Proper warm-up and cool-down periods have to be considered. Use a safety arch with chest belt system when making max. load and max. heart rate tests. The following table shows the operating steps to run the treadmill via firmware or software h/p/cosmos para control Operation of test mode Initiation: running machine is in standby mode.
Index”. For correct calculation of the UKK fitness index, the following parameters have to be entered: gender, age, weight and height. The following table shows the operating steps to run the treadmill via firmware or software h/p/cosmos para control ®...
Pr. 21. Up to 40 program steps can be stored per program. If more program steps are required, ask for the optional PC Software h/p/cosmos para graphics , which is able to control all functions of the ®...
Optional settings serve for example as confirmation of error notes on the display, or for the standard setting of the device: selection of RS232 interface protocol, etc. For optional settings on devices without UserTerminal, an external UserTerminal or a connected PC with the software h/p/cosmos para control is required.
Page 72
OP01 Reset (deleting) of error messages The required maintenance work has to be performed before deleting error messages. The h/p/cosmos service department has to be contacted before one of the service displays or error codes will be deleted! Note: This option resets exclusively the causing variable.
Page 74
(standard setting COM 1 and COM 2) h/p/cosmos coscom protocol see JAEGER OXYCON (up to 1995); if available use: www.coscom.org h/p/cosmos coscom (= 1). Not to be used after 21.03.2010! recommend using printer protocol (serial printer or converter required) h/p/cosmos coscom v3 protocol...
Page 75
28. Note: If the peripheral equipment sends an acceleration and deceleration command via the h/p/cosmos coscom protocol, the selected level in option number 29 is not valid for these speed commands. OP40...
All kind of installation and repair work as well as most maintenance work is to be carried out only by trained and authorized technicians who have been certified by h/p/cosmos. The following symbols will indicate which work can be done by the customer and which work has to be done by authorized technicians: The customer / user should perform this maintenance and monitoring work.
Maintenance and repair of the devices (also opening of the device) have to be performed by service engineers authorized and certified by h/p/cosmos, preferably within the scope of a maintenance contract. The installation by unqualified personnel may cause electric shock or fire or other damages or electric shock or injury due to a malfunction.
For this reason the worst case malfunction scenario has to be taken into account. h/p/cosmos classifies the internal firmware and software components as class B due to a number of risk reducing design features and measurements.
Within Germany, most devices are delivered and assembled by h/p/cosmos directly or by an authorised forwarder. If delivered by h/p/cosmos, the packaging will be removed and recycled.
Transport to upper or lower floors and through narrow doorways: With the standard models (sizes with running surface 150/50 cm, 170/65 cm, 190/65 cm), the left handrail is easily removed (unless it is an adjustable handrail), as there are no cables inside. The handrail at the right can be unscrewed and turned 90°...
The bearing capacity of floor and ceiling in the building must be higher than the weight of the machine. An authorized body of the operator must certify the bearing capacity for the h/p/cosmos device. For example, the h/p/cosmos saturn 300/100r: dimensions ground frame running machine: L: 3400 mm x B: 1490 mm = 5.066 m² platform. Net static weight of running machine: 1300 kg, static body weight of subject: 200 kg, dynamical weight of subject: 1200 kg (up to 6 times the body weight), total weight load of system at platform: 2500 kg = required dynamic bearing capacity (493.48 kg...
The transportation locks have to be removed after setup on the h/p/cosmos gaitway model with integrated force plate. Drive elevation up to max of 25 %. Loosen the four M8 screws, marked with the silver arrows on the underside of the aluminium force plates, and remove them.
To start again, switch on the running machine at the main ”ON“ switch and press START at the keyboard. h/p/cosmos running machines are to be installed at a voltage power connection with ground wire (earthing,) including leakage-current protection switch (interrupter) and according to VDE 0100 or/and the currently valid regulations and directives.
Page 91
Please ask your electrical engineer or h/p/cosmos if you have further questions. Before installing the running machine, please compare the specifications on the name plate concerning the mains voltage and the mains frequency with your local characteristics.
All other interface protocols (older versions coscom v2, coscom v1, coscomekg.dll, other treadmill manufacturers, etc.) are not validated by h/p/cosmos according to the norms EN 62304 and EN 14971 and must not be used for medical applications (e.g. ergometry, stress-tests, cardiorespiratory diagnostics, performance diagnostics, etc.) since March 21,...
In case the manufacturer declares multiple sockets as valid, these must not be located on the floor in order to prevent soaking by liquids or mechanical damage. h/p/cosmos herewith informs about all above and below mentioned risks, warnings and obligations. Responsibility for system integration: The operator of ME equipment and ME systems (ME=medical-electrical) is to appoint a so-called system integrator, responsible for taking care of the tasks that result from the norm.
Then, the h/p/cosmos employee / h/p/cosmos partner instructs the user according to the instruction protocol in appendix 2. It is important to include all people in the instruction and commissioning who are going to work with the running machine. After the instruction is completed, the instruction protocol has to be signed by the h/p/cosmos technician and all trained persons and returned, together with the signed delivery note and the registration form, to h/p/cosmos.
8.1 Preventive maintenance The h/p/cosmos authorised service engineers are happy to help you in case problems occur. Preventive maintenance can prevent future problems and is indispensable for the safety of technical devices. Therefore, ask our service department for an annual preventive maintenance contract, which is highly recommended by the manufacturer.
The main inspection sticker on the device (e.g. running machine) also certifies the inspection of the optional equipment and the accessories. However, inspection intervals for optional equipment and accessories (e.g. rope of h/p/cosmos airwalk unweighting system, chest belt for safety harness, compressors, etc.) may deviate significantly from inspection intervals of the main device.
An h/p/cosmos authorized technical service engineer will then advise what corrective action needs to be taken. Use original h/p/cosmos special oil only. The use of other oils or the failure to lubricate regularly may damage the running belt, running deck or the lubrication system and reduce warranty and liability. h/p/cosmos recommends a maintenance contract with an authorized h/p/cosmos service partner.
Lubricant is available on request at h/p/cosmos. The running belt should be lubricated after 1000 km at the latest and/or if dry grinding noises are heard during operation.
Based on previous experience, h/p/cosmos is able to evaluate the quantity of oil absorbed by measuring the difference in weight. However, the user should realize that this test can only show approximate values and trends and is not meant to replace the direct inspection by an authorized h/p/cosmos technician at the prescribed time intervals.
Please contact h/p/cosmos to order the special oil. The running belt should be lubricated after 1000 km at the latest or/and if dry grinding noises are detected during operation. Special devices may have different intervals.
For those models a lubrication test (tissue-test) according the instructions (available from h/p/cosmos) should be performed during the obligatory yearly maintenance. The test should be sent back to h/p/cosmos for evaluation.
Open the lid of the tank and refill the oil-tank with approximately 0.5 litres of the provided original oil, which can be ordered at h/p/cosmos. For devices constructed before 2002 and without an external control unit, add approximately 1 to 1.5 litres) Close the lid of the oil-tank, plug in the power plug and switch the device on.
8.4.6 Resetting of oil message An external UserTerminal or the PC-software h/p/cosmos para control is required for devices without ® UserTerminal. Initiation: running belt is not moving. One of the mode LEDs is flashing: manual, profile, cardio, test...
The allen key must be pulled out of the screw immediately after usage due to risk of injury! The oversize machine h/p/cosmos saturn 450/300rs has an automatic belt tightening system with motors and sensors and does not need to be tightened manually.
The allen key must be pulled out of the screw immediately after use due to risk of injury! The oversize machine h/p/cosmos saturn 450/300rs has an automatic belt centring system with motors and sensors and does not need to be centred manually; however, the user should observe the central belt position.
The surfaces of the treadmill can be cleaned with a light moistened cloth. For disinfection h/p/cosmos recommends Bacillol plus which can be ordered from h/p/cosmos, order number [cos12179]. Before use of cleaning and disinfection substances, always test the compatibility of the substance at poorly visible places.
In case of replacement see picture for right positioning of the battery. The MCU control unit also has a number of status LEDs which can assist in diagnosis. When communicating with the h/p/cosmos headquarters during error search and trouble shooting, the authorized technician should observe the indications on these LEDs and report to the h/p/cosmos headquarters.
Page 131
Running machine is still working. At MCU5 control electronic (as of manufacturing date 12/2007), this error shows low or defective a defect of the option memory (FRAM). In this case, please contact h/p/cosmos battery and request a service technician. Optional setup restores to default factory setting and device type.
Page 134
Reset of error messages on devices without UserTerminal The OIL message and SERVICE message are quit with OP01. Other error messages have to be quit by OP01 at the administrator level. In cases, an external UserTerminal or the software h/p/cosmos para control are needed. For further ®...
10 Optional settings: administrator options The administrator options are only accessible through h/p/cosmos staff and authorised service engineers. Access is only possible with a special code which is not given here. Changing an administrator option might cause serious problems if you are not familiar with all details. Therefore the h/p/cosmos service has to be contacted before modifying administrator options.
Page 136
The standard maximum speed of the selected device type (in OPTION 03) will be indicated in the SPEED display. Example: h/p/cosmos mercury 4.0 = 22.0 km/h; h-p-cosmos discovery 4.0 = 40 m/min For safety purposes (e.g. for beginners), the max. speed of every treadmill model can be limited.
Page 137
(OP11=1). If the running machine is not equipped with a speed measuring system, this can be retrofitted easily. To activate the motor brake, some parameters of the frequency inverter must be changed. Ask the service department of h/p/cosmos sports & medical gmbh for detailed instructions. Attention!
Page 138
km is flashing, adjustment range: OFF, 100 ... 5.000 km Do not change the interval without prior communication with h/p/cosmos. For the models h/p/cosmos venus and saturn, JAEGER CareFusion LE 580 CE and larger as well as COSMED T200 and larger, OP35 has the standard setting OFF because the machines are equipped with an automatic oil pump and tank.
Page 139
A downward movement of the elevation after pressing STOP can result in a jumpy reaction and in severe injuries and contusions. h/p/cosmos strongly recommends adhering to the standard setting of this option. Stop means stop for all movements and must not result in activity or movement on the part of the running machine.
Page 140
It should be 10 V. Make a note of the voltage in the service report (not possible with digital control via RS485)! Read the speed and make a note of the value in the h/p/cosmos service report: OP 48: _____ km/h Stop the running belt with the STOP key.
Page 141
240 mm (h/p/cosmos mercury since 1997) 5 = elevation control: linear motor for 0 - 28% elevation, increment sensor, lift range 300 mm (h/p/cosmos quasar 4.0 since 2002) 6 = elevation control: linear motor for 0 - 25% elevation, increment sensor, lift range 300 mm (h/p/cosmos pulsar 4.0 since 2002)
10.1 List of running machines types If you have questions regarding the correct device type, contact the h/p/cosmos service for authorization in writing before making any changes. OP03 OP04 OP05 OP07 OP11 OP34 OP48 OP89 Ø...
6 LCD displays / 6 keys / display of mode and unit with LED. UserTerminal E: no Terminal “lt-models“ and “h/p/cosmos locomotion 150/50 E med“ have no UserTerminal (no display, no (display & keyboard) DE: TouchScreen keyboard). Control via interface.
Page 147
** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25 %.
Page 149
** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25 %.
Page 151
** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25 %.
Page 153
** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx.
The clinical evaluation refers only to the h/p/cosmos treadmills and the “intersection risks“ (e.g. risk during stress test until fatigue in treadmill ergometry with cardiorespiratory diagnostic) and does not refer to the optional components themselves.
All other systems and devices which are not in the above stated list must not be connected to h/p/cosmos running machines. Please contact service@h-p-cosmos.com for the latest updated list of compatible devices and systems in writing. Only connect and/or use accessories, software and host equipment if it is approved as compatible by all manufacturers.
12.1 Arm support The h/p/cosmos arm supports (patent no.: DE 199 16 508 A1) are applicable especially in the field of rehabilitation and walking therapy, as well as for the safety of seniors and insecure persons. Therapies can start earlier and the therapist is supported efficiently in his work.
6 keys incl. 2 m spiral cord cos10106 mount for additional keyboard on arm support cos10111 h/p/cosmos additional stop button (red button on top of handgrip) for arm support right cos10107 h/p/cosmos additional stop button (red button on top of handgrip) for arm support left cos10108...
12.3 Special handrails h/p/cosmos offers a number of optional handrails and custom-made devices on request. Details and order numbers are available at the manufacturer h/p/cosmos or at your local authorized distributor. Handrail crossbar adjustable Handrail crossbar bended...
Stabilising elements (at the back) (at the front) h/p/cosmos adjustable handrails [cos10030] only available for the running machines with running surface 150/50cm (e.g. h/p/cosmos mercury med) ® Reproducibility of adjustments To obtain reproducibility, the individually optimal adjustments have to be determined for each subject (see scaling at the locking elements at the back) and noted accordingly (e.g.
The training phases can be extended, which results in a more effective therapy for the patient. The “h/p/cosmos airwalk se 135” can be used for a very wide field of applications. Therapeutic applications such as functional motion therapy, rehabilitation or neurology as well as competitive sports (e.g. coordination training for athletes, etc.).
35/70/160 cos13752 It is possible to use the h/p/cosmos airwalk with the running machine series with track size 150/50 cm or 170/65 cm or 190/65 cm, and as well with most running machines produced by other manufacturers.
Wheelchair stabilizer for deck length 200cm: [cos10227] 250cm: [cos10226] 300cm: [cos00096110031] Available for the h/p/cosmos running machine series with track size 200/75 cm (and bigger) with the mark “r” Adjust the range limiters properly to limit the operation range on the deck and to ensure the wheelchair does not fall off from the deck to the front or to the back.
Page 168
The length of the rope should be adjusted so that the subject cannot use the rear 30% of the running deck. Put on the chest belt so that the h/p/cosmos logo is at the front. In order to attach the carabiner, put the vertical belt (1) below the shoulder belts (2), facing the body, thereby unloading the seam at the joint (3).
Wear appropriate clothing. 12.16 Bottle holder, drinking bottle The bottle holder is available for all h/p/cosmos running machines and can be easily fixed with a clamp on a handrail with 60 mm diameter. The position of the bottle holder should not be at the rear end of the handrail, because this end must be reachable for the subject.
(without holder for handrail of the treadmill) [cos10002-02] UserTerminal Display Keyboard MCU5 for h/p/cosmos "lt-models", manufactured as of mid-January 2009 with MCU5 and pre-installed UserTerminal housing and blind cover. Instead of the existing blind cover on the pre-installed UserTerminal, a circuit board with 6 LCD displays (4 digits, 7-segments), 24 LEDs and a keyboard with 6 keys are installed on the UserTerminal.
12.19 Safety multiple socket The stable safety multiple socket can be used for the safe connection of a medical h/p/cosmos treadmill with peripheral devices such as PCs, printer or monitors. It meets all requirements of EN60601-1 for multiple sockets for creating connections to a medical electrical system within the patients environment, i.e.
“Technical data“. More recent computers will only have a USB interface instead of the RS232 interface. In that case a “USB to RS232 interface adapter cable“ is available at h/p/cosmos under order number [cos12769]. To control via USB interface, the processor must be a Pentium 1.8 GHz or higher.
Note: Even when using the coded senders, the transmission of the heart rate to the treadmill is still un- coded. The models h/p/cosmos locomotion, h/p/cosmos pulsar, h/p/cosmos venus and h/p/cosmos saturn are equipped with the POLAR W.I.N.D. transmitter with WearLink technology and coded transmission. The transmitter is only activated if placed correctly on the body.
® well-known version 3.0. In addition to the pure remote control function of h/p/cosmos treadmill ergometers via the PC, the software now provides additional functions such as an option setting configuration tool with integrated storage function, firmware update over flash eprom, quick-stop key, cool-down key, pause key, target speed settings, target elevation settings, stop countdown, display reset, etc.
The treadmill can be controlled automatically. The software is not designed for medical diagnostics or evaluation. The treadmill data can support the doctor’s decision. Data can be converted by the software for data transfer to EXCEL and further evaluation e.g. by EXCEL or h/p/cosmos para analysis software.
The evaluation of heart rate and lactate curves after graded tests is possible on the basis of individually selected threshold value models in graphics and tables. Test data can be entered into h/p/cosmos para analysis manually or by importing from h/p/cosmos para graphics : The entry and ®...
12.26 PC Software h/p/cosmos para motion ® The h/p/cosmos para motion is a software for recording and evaluating video analysis with individual report for printout, incl. control functions of h/p/cosmos running machines. The software is not designed for medical diagnostics or evaluation.
13 Disposal Upon request and at the expense of the client, h/p/cosmos might take over the disposal of old or defective devices. Please contact service@h-p-cosmos.com for a detailed offer. Note the information for possible disposal of the running machine parts or components through the client or a subcontractor.
A serial printer or a PCL printer with parallel/centronics plug und converter [cos10056] can be connected directly to the h/p/cosmos running machine. Following data can be printed after the UKK Walk Test even without using any PC: UKK Fitness Index, date, time, test duration, distance, elevation, heart rate, age, gender, body mass index (BMI).
After the instruction is completed, the instruction protocol is to be signed by the h/p/cosmos technician and all trained persons. Signed instruction protocol, signed delivery note and the registration form are to be sent to h/p/cosmos.
Page 187
Only connect to coscom v3 - compatible and authorized host equipment. Do not connect to equipment or software which is not explicitly declared as compatible by h/p/cosmos and by the other manufacturer. Inform about the h/p/cosmos software programs h/p/cosmos para control...
Page 188
Lubrication of running belt when "OIL"-message is indicated (quantity, intervals according to this manual, distribution of oil by "duck-walking"). Use only original h/p/cosmos lubricant authorized for this model of running machine. Observe additional lubrication requirements. After running belt lubrication, the position of the running belt must be checked once more and be adjusted if necessary.
Page 189
Advice and support documents Hand over and explain delivery papers (folder). Fill out registration form and send it to h/p/cosmos sales. Let the customer sign the delivery note and instruction protocol and send it back to h/p/cosmos sales immediately.
This will furthermore result in loss of liability and warranty. Please fill out the instruction protocol and send it back to h/p/cosmos via Fax to +49 86 69 86 42 49, or via email to sales@h-p-cosmos.com or via post...