Performing A Control Solution Test - Abbott FreeStyle Lite Owner's Booklet

Blood glucose monitoring system
Hide thumbs Also See for FreeStyle Lite:
Table of Contents

Advertisement

Available languages

Available languages

Performing a Control Solution Test

1
System Check
2
3
If the sound prompt is on, your meter will beep when the test strip
is full. A short line then moves clockwise around the display screen
while the sample is processing.
Insert the top of the test strip into your meter
until it stops. This will power on the meter.
System Check Screen
When the meter powers on, this screen
will appear so that you can make sure the
display is working properly.
Wait for the meter to display the Blood
Drop and Test Strip symbols.
Gently touch only the Sample Area of the test
strip to the control solution.
Important: Do not apply control solution to both
Test Strip Sample Areas.