OHAUS Explorer Balances Service Manual page 25

Balances
Hide thumbs Also See for Explorer Balances:
Table of Contents

Advertisement

IDNR error
Type data error
--operation not
There is no sample on the pan when
allowed--
doing Tare
Shortcut buttons
are gray and no
The current application is in procedure.
action when
pressing them.
Does not retain
Battery backup for the real time clock
the correct time
has lost its charge.
and date.
Note: There is 40-second timeout control for print under stable requirement. If the unstable
condition  continues  over  40  seconds,  balance  will  send  "ES"  to  the  printer.
®
Explorer
Series Service Manual
CHAPTER 2 - DIAGNOSTIC GUIDE
Contact the authorized dealer
Place sample on the pan and redo Tare
Cancel the procedure first.
Replace the real time clock battery on the
Base Main PCBA, see 3.4.2