Bayer HealthCare Contour Next USB User Manual

Bayer HealthCare Contour Next USB User Manual

Blood glucose monitoring system
Hide thumbs Also See for Contour Next USB:

Advertisement

Blood Glucose
Monitoring System
Uses only Bayer's C
n
blood glucose test strips
®
ontour
ext
USER GUIDE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Contour Next USB and is the answer not in the manual?

Questions and answers

Summary of Contents for Bayer HealthCare Contour Next USB

  • Page 1 Blood Glucose Monitoring System Uses only Bayer’s C blood glucose test strips ® ontour USER GUIDE...
  • Page 2: Your Meter And Lancing Device

    INTENDED USE • If your blood sugar reading is under 50 mg/dL, follow medical The C USB blood glucose monitoring system is an over ® ontour the counter (OTC) device utilized by persons with diabetes in home advice immediately. settings for the quantitative measurement of glucose in whole blood, • If your blood sugar reading is over 250 mg/dL, wash and dry your and is for single-patient use only and should not be shared.
  • Page 3 Medical Advice Immediately” screen will display and the meter will beep twice. Contact your health care professional. • If your control solution test result is out of range, call Bayer Diabetes Care Customer Service at 1-800-348-8100 (available If your blood sugar reading is over 600 mg/dL the next о...
  • Page 4: Table Of Contents

    Table of Contents rechargeable Battery Bayer's C USB meter has a rechargeable ® ontour battery. You must charge your battery before you can use Your Meter and Lancing Device ..........2 your meter and you cannot do a blood sugar test while the Meter Overview ..............4...
  • Page 5: Glucose Monitoring System

    Gray Square End (palm) Insert this end into the test strip port with the gray end face up. Sample Tip *Available on request from Bayer Diabetes Care Customer Service, Blood is drawn in here. Call 1-800-348-8100. CAUTION: Bayer’s C USB meter...
  • Page 6: Meter Overview

    meter overview Using the Selection/Scrolling Buttons Turning the meter on or off • Press and hold the top Menu button to turn the meter ▲ Button on or off. OK Button • Insert a test strip to turn the meter on and remove the test ▼...
  • Page 7: Initial Setup

    Initial Setup CuRRenT DATe & TIMe 3. The date and time will appear and you can Accept 10/20/11 Accept or Change. Before use, charge your Change 5:57 PM meter. See page 68. When you accept the date and time as they appear, you are also accepting the date format and time format.
  • Page 8 HINT: Alerts can only be set outside your hIGh AnD LoW ALeRTs selected Target Range(s). For more information, call Bayer Customer Service at 1-800-348-8100 (available 24 High: mg/dL hours a day, 7 days a week), or visit our website at...
  • Page 9: Getting Ready To Test

    Examine product for missing, damaged, or broken parts. If the test strip bottle is open inside a new box of strips, WARNING do not use those strips. Contact Bayer Diabetes Care Discuss your Target ranges with your health care professional. Customer Service at 1-800-348-8100 (available 24 hours a day, 7 days a week) for replacement parts.
  • Page 10 autoLog Your meter comes with an AutoLog feature that lets you 2. Hold the test strip with the gray end facing toward the meter. mark your test result as Fasting, Before Meal, After Meal and No Mark. 3. Insert the gray square end firmly into the test strip port Fasting –...
  • Page 11: Preparing The Lancing Device

    preparing the Lancing Device Read the M 2 lancing device insert for complete ® iCrolet 3. Insert the lancet firmly instructions. For Alternative Site Testing instructions, see into the lancing device page 20. until it comes to a full stop. This will re-set the WARNING: Potential Biohazard device.
  • Page 12: Getting The Blood Drop For Fingertip Testing

    Getting the Blood Drop for fingertip Testing WARNING 1. Press the endcap firmly • The meter and lancing device are for single patient against the puncture use. Do not share them with anyone including other site and press the blue family members! Do not use on multiple patients! release button.
  • Page 13 Test results with autoLog on Test results with autoLog off Before your test result is displayed, the AutoLog screen appears. Accepted Select Fasting, Before AuToLoG Meal, After Meal, or Fasting No Mark by pressing Before Meal the OK button when the After applying blood to the test strip when AutoLog is turned After Meal selection is highlighted.
  • Page 14: For Alternative Site Testing (Ast) - Palm

    If you do not have a clear end cap to perform AST, call Testing is right for you. Bayer Diabetes Care Customer Service at 1-800-348-8100 • Do not calibrate your continuous glucose (available 24 hours a day, 7 days a week).
  • Page 15 apply more Blood if Necessary If the first blood drop is not enough, the meter may beep twice and display “STRIP UNDERFILLED” and “APPLY MORE BLOOD NOW”. You have about 30 seconds to apply more blood onto 5. Maintain steady pressure the same strip.
  • Page 16: Test Results

    2 package insert ® iCrolet a flat surface with the for instructions on how to properly remove and Bayer logo facing down. dispose of the lancet. 3. With the used lancet still in the lancing device, Test results push the lancet needle...
  • Page 17: Cleaning And Disinfection

    1. Before disinfecting, clean the meter as described above in of cleaning and disinfection without damage. You should Cleaning Your Meter. call Bayer Diabetes Care Customer Service for assistance if your device malfunctions for any cause or if you notice For proper disinfection, you must keep all meter surfaces wet for 60 seconds.
  • Page 18 Cleaning Your Lancing Device. without damage. However if your device malfunctions for any cause, contact Bayer Diabetes Care Customer Service at 1-800-348-8100 (available 24 hours a day, 7 days a week) to troubleshoot the issue and provide a solution.
  • Page 19 High and Low Blood Sugar alert Screens Your meter has been preset with a low blood sugar WARNING (hypoglycemia) value of 70 mg/dL and a high blood sugar (hyperglycemia) value of 250 mg/dL. These are the preset • If your blood sugar reading is under 50 mg/dL, values, but can be customized by you and/or your health follow medical advice immediately.
  • Page 20: Control Solution Testing

    Bayer Diabetes Customer Service at 1-800-348-8100 (available 24 hours a day, 7 days a week). 1. Remove the test strip from the bottle and firmly snap the 10. Compare your control test result with the lid closed.
  • Page 21: Setting A Test Reminder After Testing

    Setting a Test reminder after Testing A Reminder for your next test can be set after testing or ReMInDeR 3. Press the ▲ or ▼ from the Setup menu. A Reminder can be set in 15 minute CounTDoWn button to select the increments from 15 minutes to 23 hours, 45 minutes.
  • Page 22: Getting A Reminder

    Turning off a Test reminder after Testing Getting a reminder A Reminder can be turned off or changed after testing or When the Reminder time is reached, 20 beeps will sound. from the Setup menu. The meter will turn on and a Reminder screen will appear. HINT: If you do a blood sugar test within 15 minutes of Time for your next when a Reminder is set, the Reminder will turn off.
  • Page 23: Logbook - Adding Carbs, Insulin And Notes Information

    Logbook – adding Carbs, Insulin and 3. You can enter carbs you are eating right now, or anytime Notes Information today or yesterday. CARBs Log Carbs Information If you have just eaten and When did you eat want to use the current You can log carbohydrates (carbs) information after testing carbs? time and date of your...
  • Page 24 Log Insulin Information after Testing 5. The meter will then ask if you want to log 10:30AM | 10/13 1. Press the Options insulin information. button from your test Reminder results screen and ConfIRM Select the Yes or No Options press Log Insulin.
  • Page 25: Adding Notes

    adding Notes You may add Notes to your test result that may help explain 3. Select your insulin type. results. Your notes will be saved in the Logbook. InsuLIn Fast-Acting 1. From the test result screen, press the Options button. Select insulin Long-Acting 10:30AM | 10/13...
  • Page 26: Log Carbs Information From The Logbook Menu

    Log Carbs Information from the Logbook menu reviewing the Logbook 1. Press and hold the Menu button until the meter turns on. The Logbook menu contains Log Carbs, Log Insulin and View Logbook. 2. Go to the Main Menu by pressing the Menu button on the To review entries in the Logbook: top of the meter.
  • Page 27: Trends (Averages And Summaries)

    Trends The Trends feature displays your averages and your results 2. Press the ▼ button to display the 14 Day Fasting Average. as they compare to your targets, over a period of time that If you have marked any test results as Fasting, you will you select.
  • Page 28: Setup

    Setup Viewing Trends with autoLog off 1. Press and hold the Menu button to turn the meter on. You can change options on your meter and personalise it from the Setup menu. The Setup menu displays the current Menu Logbook settings in the meter.
  • Page 29 Turning off a reminder 4. Once you have selected hours and minutes, press OK. To turn off a reminder, you can use the Setup menu, or do a When you have selected the time for your Reminder, blood sugar test within 15 minutes of the Reminder time. To use the Setup menu: select OK.
  • Page 30 Setting the Date Setting the Time 1. Press the Menu button to go to the Main Menu. 1. Press and hold the Menu button to turn the meter on. 2. Press the Setup button. 2. Press the Setup button. 3. Press the ▲ or ▼ button to select Date. 3.
  • Page 31 Turning the Sound on/off Turning autoLog on/off Sound is turned ON when you receive your meter. Some Your meter comes with AutoLog turned on. We error messages and the Reminder bell will override the recommend that you keep this feature turned on. Sound Off setting.
  • Page 32 1. Use the ▲ or ▼ buttons to highlight your choice. selected Alert level(s). For more information, call 2. Select OK. Bayer Customer Service at 1-800-348-8100 (available 24 hours a day, 7 days a week), or visit our website at www.bayercontour.com.
  • Page 33 Changing Blood Sugar Target range with autoLog off 4. To make changes to your Fasting target range, press the Change button, otherwise Accept. 1. Press the Menu button to go to the Main Menu. 2. Select Setup. TARGeT seTuP Fasting: Accept Carbs 70 - 130...
  • Page 34 Target Range(s). For more information, call 1. Press and hold the Menu button to turn the meter on. Bayer Customer Service at 1-800-348-8100 (available 2. Press Setup. 24 hours a day, 7 days a week), or visit our website at www.bayercontour.com.
  • Page 35: Customer Service

    It is not for use 1. Plug your meter into a high-power USB port on your any other time. Bayer Diabetes Care Customer Service is computer. A window will open. In the window, select available 24 hours a day, 7 days a week at 1-800-348-8100.
  • Page 36 ® diabetes luCofaCts eluxe management software. Bayer is not responsible for any erroneous results from the use of other software. For more information, call Bayer Diabetes Care Customer Service at 1-800-348-8100 (available 24 hours a day, 7 days a week), or visit our website at www.bayercontour.com.
  • Page 37: Error Detection Displays

    OK shown, press the button next to it to continue. turns yellow when the battery is low and then red when your battery is almost out of charge. • If you do not resolve the problem, contact Bayer Diabetes Care Customer Service at 1-800-348-8100 Menu Logbook (available 24 hours a day, 7 days a week).
  • Page 38 rapid Charge If the battery is low when you plug in your meter, it will Rapid If you do not charge the battery, a display will alert you: “Shutting down, Battery is dead.” You must charge Charge for about 1 minute. You can run a blood sugar test as immediately.
  • Page 39: Symptoms Of High And Low Blood Sugar

    End of meter Life/Battery removal Symptoms of High and Low Blood Sugar You can better understand your blood sugar results by being HINT: Carry out this process only when you no longer aware of the symptoms of high or low blood sugar. Some of intend to use the meter.
  • Page 40: Technical Information

    Technical Information measuring range Your meter’s measuring range is 20 - 600 mg/dL. principles of the procedure • If your blood sugar reading is under 20 mg/dL, the The C USB blood glucose test is based on ® ontour “Follow Medical Advice Immediately” screen will measurement of electrical current caused by the reaction display and the meter will beep twice.
  • Page 41 accuracy precision The C USB blood glucose monitoring system A repeatability study was conducted with the ® ontour was tested on 204 lay users using capillary blood samples USB blood glucose monitoring system using ® ontour and three C test strip lots. Results were 100 venous whole blood specimens with glucose levels from ®...
  • Page 42: Battery Status

    Check List This check list may be helpful when speaking with Bayer Diabetes Care Customer Service: 1. Locate the model number (A) and serial number (B) on the back of the meter.
  • Page 43: System Specifications

    System Specifications The C ® USB meter has been tested for radio ontour frequency interference at the frequency range and test levels Test Sample: Capillary and venous whole blood specified by ISO 15197: 2003. To avoid radio frequency interference, do not use the C USB meter near Test Result: Referenced to plasma/serum glucose ontour...
  • Page 44: Caring For Your Meter

    *Part numbers are subject to change without notice. cap, call 1-800-348-8100 (available 24 hours a day, To order parts, please call Bayer Diabetes Care Customer 7 days a week), for a replacement. Service at 1-800-348-8100, 24 hours a day, 7 days a week.
  • Page 45: Warranty

    REPLACEMENT, DESCRIBED ABOVE, IS BAYER DIABETES modification, misuse, neglect, maintenance by someone CARE’S ONLY OBLIGATION UNDER THIS WARRANTY. IN other than Bayer Diabetes Care, or failure to operate the NO EVENT SHALL BAYER DIABETES CARE BE LIABLE instrument in accordance with instructions. Further, Bayer...
  • Page 46: Index

    Index Service Information ................76 Alert (High and Low Alerts) ..............8 Setup ....................49 Alternative Site Testing (AST) ............20 AutoLog On or Off ..........55 AutoLog (Meal Markers) ..............13 Carbs Settings ..........56 Averages (see Trends) ..............46 Date ..............52 Battery (To Charge) ................
  • Page 47 Bayer, the Bayer Cross, C ontour , and the luCofaCts iCrolet No Coding logo are registered trademarks of Bayer. No Coding is a trademark of Bayer. All other trademarks are the property of their respective owners. ©2012 Bayer. All rights reserved. 84651797...