175, 177, & 179
Calibration Information
Test
(Switch
Position)
L
3 V
DC Volts
300 V
1000 V
−1000 V
* Hz
1 V 45 Hz
Frequency
1 V 2 kHz
1.5 V 20 kHz
5 V 50 kHz
0.03 V 0 Hz
E
-0.3 V
Millivolts DC
0.6 V
1.5 V 20 kHz
5 V 50 kHz
25 Ω
R
Continuity
250 Ω
19 Ω
e
2
Ohms
19000000 Ω
G
Apply 2.0 V
1
Diode
E
Apply 0.90 uF
1
Capacitance
?
4.0 A 45 Hz
Amps AC
9.0 A 1 kHz
F
0.003 A 45 Hz
AC Milliamps
0.05 A 1 kHz
0.4 A 1 kHz
I
0.003 A, 0 Hz
m
1
DC Milliamp
0.05 A
0.4 A
A
4.0 A
1
DC Amps
−9.0 A
1.
Press the YELLOW button to access this function.
2.
Does not include test lead resistance.
12
Table 2. Performance Tests 175 (continued)
5520 Output
Meter Response
Lower Limit
2.993 V DC
299.3 V DC
996 V DC
−1004 V DC
44.94 Hz
1.997 kHz
19.97 kHz
49.94 kHz
29.8 mV DC
−300.7 mV DC
598.9 mV DC
19.97 kHz
49.94 kHz
Beeper On
Beeper Off
18.6 Ω
18.68 MΩ
1.978 V DC
887 nF
3.937 A AC
8.83 A AC
2.92 mA AC
49.22 mA AC
393.7 mA AC
2.94 mA DC
49.47 mA DC
−404.3 mA DC
3.957 A DC
−9.12 A DC
Upper Limit
3.007 V DC
300.7 V DC
1004 V DC
−996 V DC
45.06 Hz
2.003 kHz
20.03 kHz
50.06 kHz
30.2 mV DC
−299.3 mV DC
601.1 mV DC
20.03 kHz
50.06 kHz
19.4 Ω
19.32 MΩ
2.022 V DC
913 nF
4.063 A AC
9.17 A AC
3.08 mA AC
50.78 mA AC
406.3 mA AC
3.06 mA DC
50.53 mA DC
−395.7 mA DC
4.043 A DC
−8.88 A DC