HP 5890 Series II Plus Reference Manual page 229

Hewlett-packard reference manual
Table of Contents

Advertisement

Test Sample Chromatograms
Test sample chromatograms
Figure 10-5.
Detector Type
Temp 250 DEGREES C
Inlet Type Ded On-Col Cap
Oven Track
Temp N/A
Operating Mode
Purge Time On
Purge Time Off
Oven Temp Temp Programmed (1
ramp)
Init Temp
Init Time
Ramp
Rate
Fin Temp
Fin Time
Flow Param (EPP)
Constant Flow
Range
4
COLUMN:
Part No.
19095(#100)
Dimensions
Sta Phase
Methyl Silicone
FID-On-Column Capillary Inlet
HP 5890 Test Sample Operating
Conditions
FIDw/MUG
On
C
N/A
N/A
min
N/A
min
60 DEGREES C
0.5
min
20
185
1
Off
¿
530
MID; 5 M
START
FLOW RATES
Carrier (He)
15 +/• 1
Hydrogen
30 +/• 1
Air
400 +/• 20
Makeup (N2)
20 +/• 1
Split Vent
Septum Purge
5 +/• 1
SAMPLE:
Type
FID Sample
¿
Inj Volume
1
Part No. 18710•60170
Composition 0.03%(V/V)each:
C14,C15, and C16 normal paraffin
hydrocarbons in hexane.
ml/min
ml/min
ml/min
ml/min
N/A
ml/min
ml/min
l
.112
.331
.267
4.743
5.398
3
6.
22
C14
C15
C16
229