Nidek Medical Mark 5 Nuvo M5C5 User Manual
Nidek Medical Mark 5 Nuvo M5C5 User Manual

Nidek Medical Mark 5 Nuvo M5C5 User Manual

Oxygen concentrator

Advertisement

Quick Links

User's Guide
Mark 5 Nuvo
M5C5
Oxygen Concentrator
Manufactured by:
Nidek Medical Products, Inc.
BIRMINGHAM, ALABAMA, USA
Federal Law (US) restricts this device to sale by, or on the
order of, a licensed physician. Read and understand this
entire manual before using this device.
IMPORTANT SAFETY GUIDELINES
DESCRIPTION / PURPOSE OF USE
START UP / INSTALLATION
CLEANING / MAINTENANCE
• If you are experiencing any signs of discomfort consult your Equipment Provider or your physician immediately.
• The unit should only be used for oxygen therapy and by medical prescription. The prescribed daily duration and flow rate must be
followed, otherwise it may present a risk to the health of the patient.
• Do not leave unit on or unattended if not in use by the patient.
• Do not use extension cords or adapters.
• Do not smoke, or allow others to smoke near the unit while in use.
• Do not use in a magnetic environment such as MRI, X-ray, etc.
• Do not remove the covers of the unit.
• Do not place unit where the intake filter ventilation could be obstructed or blocked.
• Oxygen is not a flammable gas, but accelerates the combustion of materials. To avoid all risks of fire, the unit must be kept away
from all flames, incandescent sources and sources of heat, as well as any combustible products such as oil, grease, solvents,
aerosols, etc.
• This device complies with the requirements of the FDA Quality System Regulation but its operation may be affected by other
devices being used near by.
!
ATTENTION
DO
NOT
EXPOSE
FLAMES
DO NOT SMOKE
The Mark5 Nuvo M5C5 is intended solely for medical use in oxygen therapy programs under the supervision of a physician. This
device is available by prescription only and is not intended to support or sustain life. It produces oxygen enriched product by
concentrating the oxygen contained in room air. The prescribed daily duration and flow rate must be followed, otherwise it may
present a risk to the patient's health.
1
TABLE OF CONTENTS
1
1
2
2
IMPORTANT SAFETY GUIDELINES
GLOSSARY OF SYMBOLS
DO NOT GREASE
TO
OPEN
TYPE B DEVICE
FRAGILE – HANDLE WITH CARE
DESCRIPTION / PURPOSE OF USE
ACCESSORIES AND SPARE PARTS
ALARMS AND SAFETY DEVICES
SPECIFICATIONS
TROUBLESHOOTING AND SYMBOLS
09/03
2
3
3
3
CLASS II DEVICE
TECHNICAL INFORMATION
KEEP
IN
THE
VERTIFICAL
POSITION
2010-2224

Advertisement

loading
Need help?

Need help?

Do you have a question about the Mark 5 Nuvo M5C5 and is the answer not in the manual?

Questions and answers

Summary of Contents for Nidek Medical Mark 5 Nuvo M5C5

  • Page 1 DESCRIPTION / PURPOSE OF USE The Mark5 Nuvo M5C5 is intended solely for medical use in oxygen therapy programs under the supervision of a physician. This device is available by prescription only and is not intended to support or sustain life. It produces oxygen enriched product by concentrating the oxygen contained in room air.
  • Page 2 START UP / INSTALLATION 6 & 7 1. Ensure that the Power Switch (1) is in the (0) Off position. 2. Position unit where the External Cabinet Filter (8) is not obstructed. 3. Plug the Power Cord (10) into a power outlet of the correct voltage and frequency as defined on the Technical Label (9) located on the rear panel of the machine.
  • Page 3: Specifications

    The oxygen flow is interrupted or the flow Tubing is disconnected or not Check tubing connections at the cannula feels irregular tight Straighten tubing Tubing is restricted Contact your Equipment Provider NIDEK MEDICAL PRODUCTS, INC. 3949 VALLEY EAST INDUSTRIAL DRIVE BIRMINGHAM, ALABAMA 35217 USA 09/03 2010-2224...