In Case Of Ix81 - Olympus IX2-DSU Instruction Manual

Disk scanning biological microscope
Table of Contents

Advertisement

ASSEMBLY
/ Z board setting and installation
(2)
(3)

8-23-3 In case of IX81

II . DSU
II . 8 - 2 6
Page
(1)
Z board installation
Note
1.
Loosen fixing knobs (6 pieces) for two slot covers at back side of control
box (1) and remove covers and knobs.
2.
Aligning direction of connector of control box and connector of Z board
(2), insert board along with board rails.
Push in until connectors are firmly connected.
3.
Using knobs you removed, fix Z board (2). Likewise, attach the other
cover (3).
Note
Note
For Z board, use type U-ZPCB(T2) for sure.
In case that U-ZPCB(T2) or subsequent is not used, the following
troubles may occur even if correct setting was done.
● Rotation of focusing handle and actual objective lens movement would
be reversed.
● Focusing control may not be able to initialize, depending upon the
position. (Microscope does not boot up).
TIP
All DIP switches on Z board, S1, S2 and S3 are turned to OFF at ex-factory.
You must install Z board after you turn power of control box
to OFF.
Keep one piece of slot cover that remained with care as it
may be required for Z board when it is removed.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IX2-DSU and is the answer not in the manual?

Questions and answers

Fawn Sanchez
March 21, 2025

I have an Olympus IX70 and I my "burner on" bottom is blinking. Can you please help me troubleshoot the issue?

This manual is also suitable for:

Bx-dsu

Table of Contents