Page 1
Compressor Nebulizer System Model IRC 1705 Dealer: This manual MUST be given to the end user. User: BEFORE using this product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com...
Page 2
WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment ‐ otherwise, injury or damage may occur. Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products. Invacare Select Part No 1148073 ®...
Register ONLINE at warranty.invacare.com Please have your model number and purchase date available to complete your registration. Any registration information you submit will only be used by Invacare Corporation and protected as required by applicable laws and regulations. Invacare Select Part No 1148073 ®...
Caution indicates a potentially hazardous situation which, if CAUTION not avoided, may result in property damage or minor injury or both. NOTICE The information contained in this document is subject to change without notice. Part No 1148073 Invacare Select ®...
IMPORTANT SAFEGUARDS IMPORTANT SAFEGUARDS WARNING IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product. CAUTION Federal law restricts this device to sale or rental by or on the order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device. “This appliance has a polarized plug (one blade is wider than the other). As a safety feature, this plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. Do not attempt to defeat this safety feature.” WARNING When using electrical products, especially when children are present, basic safety precautions should always be followed, including the following: Invacare Select Part No 1148073 ®...
Page 7
IMPORTANT SAFEGUARDS READ ALL INSTRUCTIONS BEFORE USING DANGER FIRE RISK ‐ DO NOT use this product in an oxygen enriched environment (i.e – oxygen tent). Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel. To Reduce the Risk of Electrocution: ALWAYS unplug this product IMMEDIATELY after using. DO NOT use while bathing. DO NOT place or store product where it can fall or be pulled into a tub or sink. DO NOT place in or drop into water or other liquid. DO NOT reach for a product that has fallen into water. Unplug IMMEDIATELY. Part No 1148073 Invacare Select ®...
Page 8
IMPORTANT SAFEGUARDS WARNING To Reduce the Risk of Burns, Electrocution, Fire and Injury to Persons: This product should NEVER be left unattended when plugged in. Close supervision is necessary when this product is used by, on, or near children and physically challenged individuals. Use this product only for its intended use as described in this manual and only with the prescribed medication. Use this product only under physician’s direction. NEVER operate this product with a damaged cord or plug, if it is not working properly, if it was dropped or damaged, or if it was dropped into water. Call qualified service personnel for examination and repair. Invacare Select Part No 1148073 ®...
Page 9
IMPORTANT SAFEGUARDS WARNING Keep the nebulizer tubing, cord, and unit away from heated or hot surfaces, including space heaters, electric blankets, stoves and similar electrical appliances. NEVER drop or insert objects into any opening or hose. NEVER use while sleeping or drowsy. NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air openings may be blocked. Keep the air openings free from lint, hair and the like. Blocked openings may cause the unit to shut down. Contact dealer immediately. The compressor nebulizer system uses the surrounding air to effectively administer the prescribed medication through the nebulizer. The user and administrator should be aware of surroundings so as not to have contaminated air filtered through the compressor nebulizer system. (Examples can be air contaminated with paint or cleaning fumes, weed killer, car exhaust, etc.) DO NOT use the compressor nebulizer system in these environments. Otherwise, serious bodily injury may occur. This unit is oil‐less. DO NOT lubricate. Part No 1148073 Invacare Select ®...
Page 10
IMPORTANT SAFEGUARDS WARNING This equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with nitrous oxide. When not administering treatment, keep out of the reach of small children. Some of the parts of the nebulizer are small enough for children to swallow. The tubing may present a strangulation hazard NEVER touch the unit with wet hands. DO NOT pull on the power cord to unplug the unit from an electrical outlet. DO NOT overload wall outlets. Maintenance MUST be performed by a qualified service technician. SAVE THESE INSTRUCTIONS Invacare Select Part No 1148073 ®...
INTRODUCTION INTRODUCTION The Invacare Select Compressor Nebulizer System is a portable medical device designed to efficiently deliver physician‐prescribed medication to the bronchial lung passages. The nebulizer system consists of the following components: • Compressor • Nebulizer • Air tubing • Mouthpiece Part No 1148073 Invacare Select ®...
In compliance with UL 1431 Personal Hygiene and Healthcare Appliances and CSA 22.2 Electromedical Equipment. Maximum Pressure: 36 psi Maximum Flow Rate: 8 to 11 L/min. Operating Flow Rate: 4.0 L/min. @ 16 psi operating pressure Invacare Select Part No 1148073 ®...
Page 13
50°F to 104°F (10°C to 40°C); 10 to 95% Altitude: 0 to 6000 ft (1825 m) above sea level Mode of Operation: Intermittent Use (30 min. ON / 30 min. OFF) NOTE: Product parameters and appearance are subject to change without notice for purposes of improvement. NOTE: Nebulization rate and particle size are dependent on the characteristics of the nebulizer chosen. Part No 1148073 Invacare Select ®...
SUPPLEMENTAL INFORMATION SUPPLEMENTAL INFORMATION Compressor If the unit shuts down frequently, you may have a low voltage situation. Low voltage can also be suspected when: • The motor does not get up to full power or speed. • Home fuses or circuit breakers activate when starting compressor. • Lights dim or remain dim when compressor is started. • Other motor operated appliances fail to operate properly or too many motor operated appliances are on same circuit. Extension Cords If you must use an extension cord, use only a cord having the same or higher electrical rating as the device being connected. Make certain the extension cord is in good condition without any exposed wires or hazardous contacts, and the extension cord wire is NOT LESS than 18 gauge for 25 feet. Invacare Select Part No 1148073 ®...
SECTION 1—HANDLING SECTION 1—HANDLING Shipping Instructions The shipping container has been designed to assure protection of this device. If this device is to be reshipped by common carrier, it should be packed in the same carton. Unpacking 1. Check for any obvious damage to the carton or its contents. If damage is evident, DO NOT use. Notify carrier/Invacare or your local Invacare dealer. 2. Remove all the loose packing from the carton. 3. Carefully remove all the components from the carton. NOTE: Retain carton and packing materials for use in shipping or storing the unit until use is required. Inspection 1. Examine exterior for damage. Inspect all components. Part No 1148073 Invacare Select ®...
SECTION 2—OPERATING INSTRUCTIONS SECTION 2—OPERATING INSTRUCTIONS WARNING DO NOT operate unit without filter. Avoid operating this product in a dusty environment. Otherwise, premature compressor wear could occur. NOTE: Clean and disinfect the nebulizer and mouthpiece before using the device for the first time. Refer to Cleaning and Disinfecting on page 20. Setup NOTE: For this procedure, refer to FIGURE 2.1 on page 17. 1. Place the unit on a table or other flat, stable surface. Be sure you can easily reach the controls when you are seated. CAUTION Make sure the power switch is in the off “O” position. 2. Gently twist the nebulizer to separate it into two sections. Invacare Select Part No 1148073 ®...
Page 17
SECTION 2—OPERATING INSTRUCTIONS 3. Be sure the pick‐up funnel is in place in the bottom section of the nebulizer. CAUTION DO NOT add over 6 ml of medication to the medication cup.The maximum capacity of the nebulizer medication cup is 6 ml. 4. Add the prescribed amount of medication to the cup of the nebulizer using an eyedropper, pre‐measured dose or ampoule. Eyedropper Nebulizer Cup FIGURE 2.1 Adding Medication 5. Gently twist the top and bottom sections together to assemble the nebulizer. Be sure the two sections fit securely. 6. Attach one end of the air tubing to the air tubing connector located on the front of the compressor. Part No 1148073 Invacare Select ®...
Page 18
SECTION 2—OPERATING INSTRUCTIONS 7. Attach the other end of the air tubing to the base of the nebulizer. NOTE: Ensure nebulizer tubing is firmly attached to compressor before use. WARNING In the case of high humidity, condensation may form inside the internal tubing. Run the unit for two minutes before connecting the nebulizer to the tubing. 8. Attach the mouthpiece to the top section of the nebulizer. Refer to the nebulizer manual for additional instructions. NOTE: Before starting a cold unit, let the unit reach room temperature in a heated area. 9. With the power switch off, plug the power cord into an appropriate electrical outlet. CAUTION This device is for intermittent use: 30 min. ON / 30 min. OFF. Failure to operate within these limits could damage the device beyond repair. Invacare Select Part No 1148073 ®...
SECTION 2—OPERATING INSTRUCTIONS Treatment Instructions NOTE: For this procedure, refer to FIGURE 2.2. 1. Turn the unit on, pressing the ON/OFF switch to the “I” position. 2. Place mouthpiece between your teeth, close your mouth and tilt the nebulizer toward your FIGURE 2.2 Using the mouth. Nebulizer 3. As aerosol begins to flow, inhale deeply, slowly, and hold breath for ten seconds. Then exhale slowly through the mouthpiece. NOTE: If you need to rest during a treatment, be sure the nebulizer remains in an upright position; otherwise, the medication may spill. 4. When treatment is complete, turn the unit off by setting the switch to the “O” position and unplug the unit from the electrical outlet. 5. Before storing supplies between treatments, clean and dry the nebulizer and accessories. Refer to Cleaning and Disinfecting on page 20. Part No 1148073 Invacare Select ®...
SECTION 3—CLEANING AND DISINFECTING SECTION 3—CLEANING AND DISINFECTING Following the cleaning instructions after each use will prevent any remaining medication in the nebulizer cup from drying and causing the device to nebulize ineffectively. It is recommended that the nebulizer and the mouthpiece be thoroughly rinsed with hot tap water after each use and cleaned with a mild detergent after the last treatment of the day. The same pieces MUST also be disinfected daily by soaking in a vinegar/water solution or a commercially available disinfectant. If your physician or respiratory therapist specifies a different cleaning procedure, follow their instructions. Cleaning the Outer Case WARNING Unplug unit before cleaning. DO NOT submerge in water to clean. Clean outer case with damp cloth. Invacare Select Part No 1148073 ®...
Removing the case creates a serious electrical shock hazard. All disassembly and maintenance of this unit MUST be done by a qualified technician. Cleaning the Nebulizer WARNING To prevent possible risk of infection due to contamination, follow these instructions: • Clean the nebulizer before initial operations. • Clean the nebulizer after each aerosol treatment. • ALWAYS use fresh cleaning solution. • DO NOT towel dry the nebulizer components after cleaning. NOTE: Your physician and/or an Invacare dealer may specify certain cleaning procedures. If so, follow their recommendations. If not, clean according to the following instructions: 1. Remove the nebulizer and mouthpiece from the tubing. Part No 1148073 Invacare Select ®...
Page 22
SECTION 3—CLEANING AND DISINFECTING 2. Disassemble nebulizer chamber by turning counterclockwise and separating. 3. Fill two plastic containers or bowls: A. Washing: Hot water/detergent solution. B. Soaking: Hot water/vinegar solution (one part vinegar to three parts water). 4. Thoroughly clean the nebulizer and mouthpiece in hot water/detergent solution. 5. Remove from the water/detergent solution and rinse with clear hot tap water. 6. Soak in hot water/vinegar solution for 30 minutes. 7. Rinse with hot tap water again and air dry thoroughly. NOTE: If using medical disinfectant cleaners, follow manufacturer’s instructions carefully. 8. Keep the outer surface of the tubing dust‐free by wiping regularly. NOTE: The nebulizer air tubing does not have to be washed internally because only filtered air passes through the tubing. Invacare Select Part No 1148073 ®...
SECTION 3—CLEANING AND DISINFECTING Cleaning the Compressor CAUTION Unplug unit before cleaning. DO NOT submerge in water to clean. DO NOT clean using abrasive cleaners. 1. Clean the casing of the main unit daily by using a soft cloth moistened with water or a mild detergent. 2. Wipe the casing and immediately dry using a soft clean cloth. Cleaning the Mouthpiece 1. Boil the mouthpiece in water for a maximum of five minutes or autoclave at a maximum of 121° C for ten minutes. 2. Allow the mouthpiece to air dry. 3. Once dry, the mouthpiece may be stored inside the nebulizer top. Part No 1148073 Invacare Select ®...
Removing the case creates a serious electrical shock hazard. All disassembly and maintenance of this unit MUST be done by a qualified technician. Routine Care To keep your device in the best condition and protect the unit from damage, follow these directions: • DO NOT subject the device or the parts to any strong shocks such as dropping the device on the floor. • DO NOT store the device in extreme hot or cold temperatures, high humidity or under direct sunlight. Storage Store the repackaged device in a dry and safe area. DO NOT place other objects on top of the repackaged device. Invacare Select Part No 1148073 ®...
SECTION 4—MAINTENANCE Nebulizer Change The nebulizer is designed for single patient use. The nebulizer MUST be replaced after a long period of inactivity, or when it is obstructed by dry medication, dust, etc. Reusable nebulizers should last from six months to one year with general use. Refer to the nebulizers owner’s manual. Disposable nebulizers should last 30 days. For disposable nebulizers, it is recommended that an extra nebulizer be kept at all times. Refer to the disposable nebulizer’s owner’s manual. Filter Change 1. The filter MUST be replaced after approximately 30 hours of use or when it turns grey. 2. Open the filter holder to replace the filter with a new one. CAUTION DO NOT use cotton or any other material. DO NOT wash or clean the filter. Use only Invacare filters. DO NOT operate unit without an air filter. Part No 1148073 Invacare Select ®...
1. Make sure the plug is firmly fitted to the wall socket. switch ON. 2. Try plugging into another wall socket that is known to work. 3. Contact your Invacare dealer. Device does not 1. Make sure that the ends of the air tubing are fitted nebulize, or it tightly onto the main unit and onto the nebulizer.
Page 27
SECTION 5—TROUBLESHOOTING NOTES Part No 1148073 Invacare Select ®...
Page 28
SECTION 5—TROUBLESHOOTING NOTES Invacare Select Part No 1148073 ®...
Invacare warrants this product when purchased new and unused to be free from defects in materials and workmanship for a period of 3 years from date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty.
Page 30
LIMITED WARRANTY Upon receiving notice of an alleged defect in a product, Invacare will issue a serialized return authorization. It shall be the responsibility of the purchaser to pack the product(s) or part(s) in an appropriate manner to avoid shipping damage and return the product(s) or part(s), at the purchaser’s expense, to either...
Page 31
PURSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. SOME STATES DO NOT ALLOW THE EXCLUSION OR...