Invacare Select IRC1705 Operator's Manual

Invacare Select IRC1705 Operator's Manual

Compressor nebulizer system
Hide thumbs Also See for Select IRC1705:

Advertisement

Operator's Manual
Invacare
Select
®
Compressor Nebulizer System
Model IRC 1705
Dealer: This manual MUST be given to the end user.
User: BEFORE using this product, read this manual and
save for future reference.
For more information regarding
Invacare products, parts, and
services, please visit
www.invacare.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Select IRC1705 and is the answer not in the manual?

Questions and answers

victoria Depew
January 16, 2025

Where do i get the .refills. for the compressor nebulizer system.model irc 1705

Summary of Contents for Invacare Select IRC1705

  • Page 1 Compressor Nebulizer System Model IRC 1705 Dealer: This manual MUST be given to the end user. User: BEFORE using this product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com...
  • Page 2 WARNING DO NOT use this product or any available optional  equipment without first completely reading and  understanding these instructions and any additional  instructional material such as owner’s manuals, service  manuals or instruction sheets supplied with this product or  optional equipment. If you are unable to understand the  warnings, cautions or instructions, contact a healthcare  professional, dealer or technical personnel before  attempting to use this equipment ‐ otherwise, injury or  damage may occur. Invacare products are specifically designed and  manufactured for use in conjunction with Invacare  accessories. Accessories designed by other manufacturers  have not been tested by Invacare and are not recommended  for use with Invacare products. Invacare Select Part No 1148073 ®...
  • Page 3: Table Of Contents

    Inspection .................... 15 SECTION 2—OPERATING INSTRUCTIONS ..16 Setup..................... 16 Treatment Instructions ..............19 SECTION 3—CLEANING AND DISINFECTING ... 20 Cleaning the Outer Case..............20 Cleaning the Nebulizer ..............21 Cleaning the Compressor ............... 23 Part No 1148073 Invacare Select ®...
  • Page 4: Register Your Product

    Register ONLINE at warranty.invacare.com Please have your model number and purchase date available to complete your registration. Any registration information you submit will only be used by Invacare Corporation and protected as required by applicable laws and regulations. Invacare Select Part No 1148073 ®...
  • Page 5: Special Notes

    Caution indicates a potentially hazardous situation which, if CAUTION not avoided, may result in property damage or minor injury or both. NOTICE The information contained in this document is subject to change without notice. Part No 1148073 Invacare Select ®...
  • Page 6: Important Safeguards

    IMPORTANT SAFEGUARDS IMPORTANT SAFEGUARDS WARNING IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product. CAUTION Federal law restricts this device to sale or rental by or on the  order of a physician, or any other practitioner licensed by the  law of the State in which he/she practices to use or order the  use of this device. “This appliance has a polarized plug (one blade is wider than  the other). As a safety feature, this plug will fit in a polarized  outlet only one way. If the plug does not fit fully in the outlet,  reverse the plug. If it still does not fit, contact a qualified  electrician. Do not attempt to defeat this safety feature.” WARNING When using electrical products, especially when children are  present, basic safety precautions should always be followed,  including the following: Invacare Select Part No 1148073 ®...
  • Page 7 IMPORTANT SAFEGUARDS READ ALL INSTRUCTIONS BEFORE USING DANGER FIRE RISK ‐ DO NOT use this product in an oxygen  enriched environment (i.e – oxygen tent). Risk of electric shock. DO NOT disassemble. Refer  servicing to qualified service personnel. To Reduce the Risk of Electrocution: ALWAYS unplug this product IMMEDIATELY after using. DO NOT use while bathing. DO NOT place or store product where it can fall or be  pulled into a tub or sink. DO NOT place in or drop into water or other liquid. DO NOT reach for a product that has fallen into water.  Unplug IMMEDIATELY. Part No 1148073 Invacare Select ®...
  • Page 8 IMPORTANT SAFEGUARDS WARNING To Reduce the Risk of Burns, Electrocution, Fire and Injury to Persons: This product should NEVER be left unattended when  plugged in. Close supervision is necessary when this product is used by,  on, or near children and physically challenged individuals. Use this product only for its intended use as described in this  manual and only with the prescribed medication. Use this  product only under physician’s direction. NEVER operate this product with a damaged cord or plug, if  it is not working properly, if it was dropped or damaged, or if  it was dropped into water. Call qualified service personnel for  examination and repair. Invacare Select Part No 1148073 ®...
  • Page 9 IMPORTANT SAFEGUARDS WARNING Keep the nebulizer tubing, cord, and unit away from heated  or hot surfaces, including space heaters, electric blankets,  stoves and similar electrical appliances. NEVER drop or insert objects into any opening or hose. NEVER use while sleeping or drowsy. NEVER block the air openings of the product or place it on a  soft surface, such as a bed or couch, where the air openings  may be blocked. Keep the air openings free from lint, hair and  the like. Blocked openings may cause the unit to shut down.  Contact dealer immediately. The compressor nebulizer system uses the surrounding air to  effectively administer the prescribed medication through the  nebulizer. The user and administrator should be aware of  surroundings so as not to have contaminated air filtered  through the compressor nebulizer system. (Examples can be  air contaminated with paint or cleaning fumes, weed killer,  car exhaust, etc.)  DO NOT use the compressor nebulizer system in these  environments. Otherwise, serious bodily injury may occur. This unit is oil‐less. DO NOT lubricate. Part No 1148073 Invacare Select ®...
  • Page 10 IMPORTANT SAFEGUARDS WARNING This equipment is not suitable for use in the presence of a  flammable anesthetic mixture with air or with nitrous oxide. When not administering treatment, keep out of the reach of  small children. Some of the parts of the nebulizer are small  enough for children to swallow. The tubing may present a  strangulation hazard NEVER touch the unit with wet hands. DO NOT pull on the power cord to unplug the unit from an  electrical outlet. DO NOT overload wall outlets. Maintenance MUST be performed by a qualified service  technician. SAVE THESE INSTRUCTIONS Invacare Select Part No 1148073 ®...
  • Page 11: Introduction

    INTRODUCTION INTRODUCTION The Invacare Select Compressor Nebulizer System is a  portable medical device designed to efficiently deliver  physician‐prescribed medication to the bronchial lung  passages. The nebulizer system consists of the following  components: • Compressor • Nebulizer • Air tubing • Mouthpiece Part No 1148073 Invacare Select ®...
  • Page 12: Typical Product Parameters

    In compliance with UL 1431 Personal Hygiene and Healthcare Appliances and CSA 22.2 Electromedical Equipment. Maximum Pressure: 36 psi Maximum Flow Rate: 8 to 11 L/min. Operating Flow Rate: 4.0 L/min. @ 16 psi operating pressure Invacare Select Part No 1148073 ®...
  • Page 13 50°F to 104°F (10°C to 40°C); 10 to 95% Altitude: 0 to 6000 ft (1825 m) above sea level Mode of Operation: Intermittent Use (30 min. ON / 30 min. OFF) NOTE: Product parameters and appearance are subject to change  without notice for purposes of improvement. NOTE: Nebulization rate and particle size are dependent on the  characteristics of the nebulizer chosen. Part No 1148073 Invacare Select ®...
  • Page 14: Supplemental Information

    SUPPLEMENTAL INFORMATION SUPPLEMENTAL INFORMATION Compressor If the unit shuts down frequently, you may have a low voltage  situation.  Low voltage can also be suspected when: • The motor does not get up to full power or speed. • Home fuses or circuit breakers activate when starting  compressor. • Lights dim or remain dim when compressor is started. • Other motor operated appliances fail to operate properly  or too many motor operated appliances are on same  circuit. Extension Cords If you must use an extension cord, use only a cord having the  same or higher electrical rating as the device being connected.  Make certain the extension cord is in good condition without  any exposed wires or hazardous contacts, and the extension  cord wire is NOT LESS than 18 gauge for 25 feet. Invacare Select Part No 1148073 ®...
  • Page 15: Section 1-Handling

    SECTION 1—HANDLING SECTION 1—HANDLING Shipping Instructions The shipping container has been designed to assure  protection of this device. If this device is to be reshipped by common carrier, it should  be packed in the same carton. Unpacking 1. Check for any obvious damage to the carton or its  contents. If damage is evident, DO NOT use. Notify  carrier/Invacare or your local Invacare dealer. 2. Remove all the loose packing from the carton. 3. Carefully remove all the components from the carton. NOTE: Retain carton and packing materials for use in shipping or  storing the unit until use is required. Inspection 1. Examine exterior for damage. Inspect all components. Part No 1148073 Invacare Select ®...
  • Page 16: Section 2-Operating Instructions

    SECTION 2—OPERATING INSTRUCTIONS SECTION 2—OPERATING INSTRUCTIONS WARNING DO NOT operate unit without filter. Avoid operating this product in a dusty environment.  Otherwise, premature compressor wear could occur. NOTE: Clean and disinfect the nebulizer and mouthpiece before  using the device for the first time. Refer to Cleaning and  Disinfecting on page 20. Setup NOTE: For this procedure, refer to FIGURE 2.1 on page 17. 1. Place the unit on a table or other flat, stable surface. Be sure  you can easily reach the controls when you are seated. CAUTION Make sure the power switch is in the off “O” position. 2. Gently twist the nebulizer to separate it into two sections. Invacare Select Part No 1148073 ®...
  • Page 17 SECTION 2—OPERATING INSTRUCTIONS 3. Be sure the pick‐up funnel is in place in the bottom section  of the nebulizer. CAUTION DO NOT add over 6 ml of medication to the medication  cup.The maximum capacity of the nebulizer medication cup  is 6 ml. 4. Add the prescribed amount of medication to the cup of the  nebulizer using an eyedropper, pre‐measured dose or  ampoule. Eyedropper Nebulizer Cup FIGURE 2.1 Adding Medication 5. Gently twist the top and bottom sections together to  assemble the nebulizer. Be sure the two sections fit  securely. 6. Attach one end of the air tubing to the air tubing connector  located on the front of the compressor. Part No 1148073 Invacare Select ®...
  • Page 18 SECTION 2—OPERATING INSTRUCTIONS 7. Attach the other end of the air tubing to the base of the  nebulizer. NOTE: Ensure nebulizer tubing is firmly attached to compressor  before use. WARNING In the case of high humidity, condensation may form inside  the internal tubing. Run the unit for two minutes before  connecting the nebulizer to the tubing. 8. Attach the mouthpiece to the top section of the nebulizer.  Refer to the nebulizer manual for additional instructions. NOTE: Before starting a cold unit, let the unit reach room  temperature in a heated area. 9. With the power switch off, plug the power cord into an  appropriate electrical outlet. CAUTION This device is for intermittent use: 30 min. ON / 30 min. OFF.  Failure to operate within these limits could damage the  device beyond repair. Invacare Select Part No 1148073 ®...
  • Page 19: Treatment Instructions

    SECTION 2—OPERATING INSTRUCTIONS Treatment Instructions NOTE: For this procedure, refer to FIGURE 2.2. 1. Turn the unit on,  pressing the ON/OFF  switch to the “I” position. 2. Place mouthpiece  between your teeth, close  your mouth and tilt the  nebulizer toward your  FIGURE 2.2 Using the mouth. Nebulizer 3. As aerosol begins to flow, inhale deeply, slowly, and hold  breath for ten seconds. Then exhale slowly through the  mouthpiece. NOTE: If you need to rest during a treatment, be sure the nebulizer  remains in an upright position; otherwise, the medication may spill. 4. When treatment is complete, turn the unit off by setting  the switch to the “O” position and unplug the unit from  the electrical outlet. 5. Before storing supplies between treatments, clean and dry  the nebulizer and accessories. Refer to Cleaning and  Disinfecting on page 20. Part No 1148073 Invacare Select ®...
  • Page 20: Section 3-Cleaning And Disinfecting

    SECTION 3—CLEANING AND DISINFECTING SECTION 3—CLEANING AND DISINFECTING Following the cleaning instructions after each use will  prevent any remaining medication in the nebulizer cup from  drying and causing the device to nebulize ineffectively. It is recommended that the nebulizer and the mouthpiece be  thoroughly rinsed with hot tap water after each use and  cleaned with a mild detergent after the last treatment of the  day. The same pieces MUST also be disinfected daily by  soaking in a vinegar/water solution or a commercially  available disinfectant. If your physician or respiratory  therapist specifies a different cleaning procedure, follow their  instructions. Cleaning the Outer Case WARNING Unplug unit before cleaning.  DO NOT submerge in water to clean. Clean outer case with  damp cloth. Invacare Select Part No 1148073 ®...
  • Page 21: Cleaning The Nebulizer

    Removing the case creates a serious electrical shock hazard. All disassembly and maintenance of this unit MUST be done by a qualified technician. Cleaning the Nebulizer WARNING To prevent possible risk of infection due to contamination,  follow these instructions: • Clean the nebulizer before initial operations. • Clean the nebulizer after each aerosol treatment. • ALWAYS use fresh cleaning solution. • DO NOT towel dry the nebulizer components after  cleaning. NOTE: Your physician and/or an Invacare dealer may specify  certain cleaning procedures. If so, follow their recommendations. If  not, clean according to the following instructions: 1. Remove the nebulizer and mouthpiece from the tubing. Part No 1148073 Invacare Select ®...
  • Page 22 SECTION 3—CLEANING AND DISINFECTING 2. Disassemble nebulizer chamber by turning  counterclockwise and separating. 3. Fill two plastic containers or bowls: A. Washing: Hot water/detergent solution. B. Soaking: Hot water/vinegar solution (one part vinegar  to three parts water). 4. Thoroughly clean the nebulizer and mouthpiece in hot  water/detergent solution. 5. Remove from the water/detergent solution and rinse with  clear hot tap water. 6. Soak in hot water/vinegar solution for 30 minutes. 7. Rinse with hot tap water again and air dry thoroughly. NOTE: If using medical disinfectant cleaners, follow  manufacturer’s instructions carefully. 8. Keep the outer surface of the tubing dust‐free by wiping  regularly. NOTE: The nebulizer air tubing does not have to be washed  internally because only filtered air passes through the tubing. Invacare Select Part No 1148073 ®...
  • Page 23: Cleaning The Compressor

    SECTION 3—CLEANING AND DISINFECTING Cleaning the Compressor CAUTION Unplug unit before cleaning. DO NOT submerge in water to  clean. DO NOT clean using abrasive cleaners. 1. Clean the casing of the main unit daily by using a soft cloth  moistened with water or a mild detergent. 2. Wipe the casing and immediately dry using a soft clean  cloth. Cleaning the Mouthpiece 1. Boil the mouthpiece in water for a maximum of five  minutes or autoclave at a maximum of 121° C for ten  minutes. 2. Allow the mouthpiece to air dry. 3. Once dry, the mouthpiece may be stored inside the  nebulizer top. Part No 1148073 Invacare Select ®...
  • Page 24: Section 4-Maintenance

    Removing the case creates a serious electrical shock hazard. All disassembly and maintenance of this unit MUST be done by a qualified technician. Routine Care To keep your device in the best condition and protect the unit  from damage, follow these directions: • DO NOT subject the device or the parts to any strong  shocks such as dropping the device on the floor. • DO NOT store the device in extreme hot or cold  temperatures, high humidity or under direct sunlight. Storage Store the repackaged device in a dry and safe area. DO NOT  place other objects on top of the repackaged device. Invacare Select Part No 1148073 ®...
  • Page 25: Nebulizer Change

    SECTION 4—MAINTENANCE Nebulizer Change The nebulizer is designed for single patient use. The nebulizer  MUST be replaced after a long period of inactivity, or when it  is obstructed by dry medication, dust, etc. Reusable  nebulizers should last from six months to one year with  general use. Refer to the nebulizers owner’s manual.  Disposable nebulizers should last 30 days. For disposable  nebulizers, it is recommended that an extra nebulizer be kept  at all times. Refer to the disposable nebulizer’s owner’s  manual. Filter Change 1. The filter MUST be replaced after approximately 30 hours  of use or when it turns grey. 2. Open the filter holder to replace the filter with a new one. CAUTION DO NOT use cotton or any other material. DO NOT wash or clean the filter. Use only Invacare filters. DO NOT operate unit without an air filter. Part No 1148073 Invacare Select ®...
  • Page 26: Section 5-Troubleshooting

    1. Make sure the plug is firmly fitted to the wall socket. switch ON. 2. Try plugging into another wall socket that is known to work. 3. Contact your Invacare dealer. Device does not 1. Make sure that the ends of the air tubing are fitted nebulize, or it tightly onto the main unit and onto the nebulizer.
  • Page 27 SECTION 5—TROUBLESHOOTING NOTES Part No 1148073 Invacare Select ®...
  • Page 28 SECTION 5—TROUBLESHOOTING NOTES Invacare Select Part No 1148073 ®...
  • Page 29: Limited Warranty

    Invacare warrants this product when purchased new and unused to be free from defects in materials and workmanship for a period of 3 years from date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty.
  • Page 30 LIMITED WARRANTY Upon receiving notice of an alleged defect in a product, Invacare will issue a serialized return authorization. It shall be the responsibility of the purchaser to pack the product(s) or part(s) in an appropriate manner to avoid shipping damage and return the product(s) or part(s), at the purchaser’s expense, to either...
  • Page 31 PURSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. SOME STATES DO NOT ALLOW THE EXCLUSION OR...
  • Page 32 One Invacare Way identified by the symbols ™ and ®. All Elyria, Ohio USA trademarks are owned by or licensed 44036-2125 to Invacare Corporation or its 800-333-6900 subsidiaries unless otherwise noted. © 2008 Invacare Corporation In Florida 2101 E. Lake Mary Blvd.