Mitsubishi Electric CP30DW Operation Manual page 6

Digital color printer
Hide thumbs Also See for CP30DW:
Table of Contents

Advertisement

Compliance with the following EMC guidance is assured only for the use in Europe (220 - 240 V).
Guidance and manufacturer's declaration - electromagnetic emissions
The Model CP30DW is intended for use in the electromagnetic environment specified below.
The customer or user of the Model CP30DW should assure that it is used in such an environment.
Emissions test
RF emissions
EN 55011
RF emissions
EN 55011
Harmonic emissions
EN 61000-3-2
Voltage fluctuations /
flicker emissions
EN 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The Model CP30DW is intended for use in the electromagnetic environment specified below.
The customer or user of the Model CP30DW should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
Electrical fast
transient/burst
EN 61000-4-4
Surge
EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
EN 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
EN 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
Class A
Complies
EN 60601 test
Compliance
level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input
±1 kV for input
/output lines
/output lines
±1 kV line(s) to
±1 kV line(s) to
line(s)
line(s)
±2 kV line(s) to
±2 kV line(s) to
earth
earth
< 5% U
< 5% U
T
(> 95% dip in U
)
(> 95% dip in U
T
for 0.5 cycle
for 0.5 cycle
40% U
40% U
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
70% U
70% U
T
(30% dip in v
)
(30% dip in U
T
for 25 cycles
for 25 cycles
< 5% U
< 5% U
T
(> 95% dip in U
)
(> 95% dip in U
T
for 5 sec.
for 5 sec.
3 A/m
3 A/m
Electromagnetic environment - guidance
The Model CP30DW uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause interference in nearby
electronic equipment.
The Model CP30DW is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Electromagnetic environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
T
typical commercial or hospital
)
T
environment. If the user of the Model
CP30DW requires continued operation
T
during power mains interruptions, it is
)
T
recommended that the Model CP30DW
be powered from an uninterruptible power
supply or a battery.
T
)
T
T
)
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.