Do you have a question about the ROSEMOUNT 5300 and is the answer not in the manual?
Questions and answers
tiago clemente ribeiro
February 27, 2025
não estou conseguindo habilitar o simula de valores no rosemount 5300 series, mostra erro!
1 comments:
Mr. Anderson
February 27, 2025
The context does not provide information on enabling value simulation for the Emerson Rosemount 5300 series. However, it mentions how to disable simulation modes. To turn off the Advanced Simulation Mode, set Holding Register 3600 to 0. If you need to enable value simulation, refer to the device manual or configuration settings.
This answer is automatically generated
Tiago
February 27, 2025
Não estou conseguindo ativar o simulação no rosemount 5300
Wireless guided wave radar, rosemount 3308 series (196 pages)
Summary of Contents for Emerson ROSEMOUNT 5300
Page 1
Reference Manual 00809-0100-4530, Rev BA July 2009 Rosemount 5300 Series Superior Performance Guided Wave Radar www.rosemount.com...
Page 3
Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product. Within the United States, Emerson Process Management has two toll-free assistance numbers. Customer Central: 1-800-999-9307(7:00 a.m. to 7:00 p.m. CST) Technical support, quoting, and order-related questions.
Page 7
Viewing Input and Holding Registers......7-27 Removing the Transmitter Head ......7-28 Rosemount 5300 Series TOC-3...
Page 8
Rosemount 5300 Series SECTION 8 Safety Instrumented Systems (4-20 mA only) APPENDIX A Reference Data TOC-4 Changing a Probe ........7-29 Probe and Firmware Compatibility .
Equipment ratings and certifications are no longer valid on any products that have been damaged or modified without the prior written permission of Emerson Process Management. Any continued use of product that has been damaged or modified without prior written authorization is at the customer's sole risk and expense.
Rosemount 5300 Series MANUAL OVERVIEW This manual provides installation, configuration and maintenance information for the Rosemount 5300 Series Radar Transmitter. Section 2: Transmitter Overview • Theory of operation • Description of the transmitter • Process and vessel characteristics Section 3: Mechanical Installation •...
To expedite the return process outside of the United States, contact the nearest Emerson Process Management representative. Within the United States, call the Emerson Process Management Instrument and Valves Response Center using the 1-800-654-RSMT (7768) toll-free number. This center, available 24 hours a day, will assist you with any needed information or materials.
Return Material Authorization (RMA) number. The center will also ask for the process material to which the product was last exposed. Emerson Process Management Instrument and Valves Response Center representatives will explain the additional information and procedures necessary to return goods exposed to hazardous substance can avoid injury if they are informed of and understand the hazard.
Installation Procedure ......page 2-15 The Rosemount 5300 Series Radar Transmitter is a smart, two-wire continuous level transmitter based on Time Domain Reflectometry (TDR) principles.
Rosemount 5300 Series APPLICATIONS Figure 2-2. Application examples The Rosemount 5300 Series Radar Transmitter series is suited for aggregate (total) level measurements on most liquids, semi-liquids, solids, and liquid/liquid interfaces. Guided microwave technology offers the highest reliability and precision to...
Page 17
The Rosemount 5302 measures both level and interface level in a separator tank. Water The Rosemount 5300 Series is a good choice for underground tanks. It is installed on the top of the tank with the radar pulse concentrated near the probe.
Page 18
5303 probe, is the solution for solids, powders and granules. It measures independently of dust, angled surfaces etc. The Rosemount 5300 with Dynamic Vapor Compensation will automatically compensate for dielectric changes in high pressure steam applications and maintain the level accuracy.
M20, eurofast, minifast Radar Electronics The Rosemount 5300 Series Radar Transmitter has an aluminum or stainless steel (SST) transmitter housing containing advanced electronics and software for signal processing. SST housing is preferred for harsh environment applications, such as off-shore platforms or other locations where the housing can be exposed to corrodents, such as salt solutions and caustics.
Page 20
Rosemount 5300 Series Figure 2-4. Remote Housing components. Dual Compartment Housing Remote Housing allows for the transmitter head to be mounted separately from the probe. U-bolt Cable Remote Connection Reference Manual 00809-0100-4530, Rev BA July 2009 Clamping Brackets Bracket M50 nut...
TRANSMITTER Integral Display The Rosemount 5300 Series Radar Transmitter is loop-powered, and it uses the same two wires for both power supply and output signal. The output is a 4-20 mA analog signal superimposed with a digital HART Fieldbus signal.
Page 22
Rosemount 5300 Series Figure 2-6. F OUNDATION Fieldbus system architecture Host/DCS system (e.g. DeltaV 375 Field Communicator Rosemount 5400 Rosemount 5300 Note: Intrinsically safe installations may allow fewer devices per I.S. barrier due to current limitations. Maintenance ® Display Rosemount 5600...
(1) For overall level applications, a changing dielectric has no effect on the measurement. For interface measurements, a changing dielectric for the top fluid will degrade the accuracy of the interface measurement. (2) Limited measuring range. The following guidelines should be used to choose the appropriate probe for the Rosemount 5300 transmitter: Rigid Single Flexible Single Lead...
Rosemount 5300 Series Transition Zones Figure 2-7. Transition Zones Table 2-2. Transition Zones for different probe types and dielectric constants Dielectric Rigid Single Lead Constant Upper 4.3 in. (11 cm) Transition 6.3 in. (16 cm) Zone 2 in. (5 cm)
Foam Vapor Boiling Hydrocarbons The Rosemount 5300 Series has high sensitivity because of its advanced signal processing and high signal to noise ratio. This makes it able to handle various disturbances, however, the following circumstances should be considered before mounting the transmitter.
1.4 (1.25 if installed in a metallic (1)(2) bypass or stilling well) (1) The probe end projection software function will improve the minimum dielectric constant. Consult you local Emerson Process Management representative for details. (2) Measuring range may be lower depending on installation.
Page 27
However, characteristics may vary between the different applications. Flexible Single Lead 115 (35) 98 (30) 82 (25) 66 (20) 49 (15) 33 (10) 16 (5) Upper product dielectric constant Rosemount 5300 Series Lower product dielectric constant 2-13...
Emerson Process Management representative. Because the radar signal is transmitted along a probe, the Rosemount 5300 Radar transmitter is generally not affected by objects in the tank. Avoid physical contact with metallic objects when Twin Lead or Single Lead probes are used.
(see page 3-15) Wire the transmitter (see Section 4: Electrical Installation) Make sure covers and cable/conduit connections are tight Power up the transmitter Configure the transmitter (see Section 5: Configuration) Verify measurements Set the Write Protection Rosemount 5300 Series 2-15...
Page 30
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 2-16...
Probes covered with plastic and/or with plastic discs may generate an ignition-capable level of electrostatic charge under certain extreme conditions. Therefore, when the probe is used in a potentially explosive atmosphere, appropriate measures must be taken to prevent electrostatic discharge. Rosemount 5300 Series ). Please...
Page 32
Equipment ratings and certifications are no longer valid on any products that have been damaged or modified without the prior written permission of Emerson Process Management. Any continued use of product that has been damaged or modified without prior written authorization is at the customer's sole risk and expense.
For Remote Housing installation see Appendix D: Remote Mounting. The Rosemount 5300 Series has a threaded connection for easy mounting on a tank roof. It can also be mounted on a nozzle by using different flanges.
Page 34
Rosemount 5300 Series Table 3-1. Nozzle considerations The transmitter can be mounted in nozzles by using an appropriate flange. The nozzle sizes given in Table 3-1 show the recommended dimensions. For small nozzles, it may be necessary to increase the Upper Null Zone (UNZ) to reduce the measuring range in the upper part of the tank.
(d>8 in./200 mm), if the threaded version is used. Metal flange Ø>2 in./ D N50 Electromagnetic disturbances should be kept to a minimum since they may affect measurement performance. Rosemount 5300 Series Metal sheet Ø>8 in. /200 mm...
0.24 in. (6 mm) diameter Tensile strength is minimum 6519 lb (29 kN) Collapse load is maximum 7868 lb (35 kN) Keep the following in mind when planning installation of the Rosemount 5300 in solid applications: • There might be considerable down-pull forces on silo roofs caused by...
NOTE! To avoid disturbances from object near the pipe, metal-pipes are preferred, especially in applications with low dielectric constant. Rosemount 5300 Series Tensile load for 0.24 in. (6 mm) flexible single lead probe, lb (kN) Probe length 115 ft (35 m) Tank Ø=...
Page 38
Chambers larger than 6 in. (150 mm) can be used but provide no advantages for radar measurement. It is recommended that single probes are used with the Rosemount 5300 Series. Other probe types are more susceptible to build-up and are not recommended.
Page 39
When mounting in a Rosemount 9901 chamber, the probe length to use can be calculated with these formulas: Side-and-Side dimension: Probe length = Centre-to-Centre dimension + 19 in. (48 cm) Side-and-Bottom dimension: Probe length = Centre-to-Centre dimension + 4 in. (10 cm) Side-and-Side Side-and-Bottom dimension dimension Rosemount 5300 Series...
NOTE! The formulas are not valid when using Dynamic Vapor Compensation probes. A Rosemount 5300 Series transmitter is the perfect replacement for an existing displacer chamber. To simplify installation, proprietary flanges are offered to allow for using the same chambers.
20 in. (500 mm) concrete or plastic tank walls, rugged metal tank walls. (1) When measuring in low DC (around 1.4). For higher DC, the recommended free space is lower. Rosemount 5300 Series Flexible Twin 4 in. (100 mm) 3-11...
Rosemount 5300 Series Recommended Mounting Position for Liquids Figure 3-8. Mounting Position 3-12 Tank conditions are recommended to be carefully considered when finding the appropriate mounting position for the transmitter. The transmitter should be mounted so the influence of disturbing objects is reduced to a minimum.
Product build-up on the silo walls near the probe may interfere with measurements. Choose a mounting position where the probe is not in contact with, or close to, the product build-up Rosemount 5300 Series 1 in./100 in. (1 cm/m) to...
Figure 3-9. Ambient temperature vs. process temperature. -320 (-196) 3-14 When the Rosemount 5300 is installed in high temperature applications, consider the maximum ambient temperature. Tank insulation should not exceed 4 in. (10 cm). HTHP version Ambient Temperature °F (°C)
3. Tighten the bolts. 4. Loosen the nut that connects the transmitter housing to Flange the probe slightly. Probe 5. Rotate the transmitter housing so the cable entries/display face the desired direction. Gasket 6. Tighten the nut. Tank flange Rosemount 5300 Series 3-15...
Rosemount 5300 Series Figure 3-11. Tank connection with loose flange (“plate design”). Transmitter head Flange nut Bolts Probe Tank flange Threaded Connection Figure 3-12. Threaded tank connection. Tank connection Sealant on threads or Probe gasket (for BSP/G threads) 3-16 Transmitters delivered with Alloy probes featuring plate design are mounted as described below: 1.
3. Fasten the Tri-Clamp to the tank with a clamp. 4. Loosen the nut that connects the transmitter housing to the probe slightly. 5. Rotate the transmitter housing so the cable entries/display face the desired direction. Tri-Clamp 6. Tighten the nut. Clamp Rosemount 5300 Series 3-17...
Rosemount 5300 Series Bracket Mounting U-bolt M6 screw Pipe mounting (vertical pipe) Wall mounting 3-18 Bracket Clamping Brackets Connect the transmitter head on the probe Connect the transmitter head on the probe, making sure that the M50 nut is properly tightened.
2. If a centering disc is used, follow the instructions on page 3-26. 3. Update the transmitter configuration to the new probe length, see “Tank and Probe Geometry“ on page 5-4 and “Probe“ on page 5-21. Rosemount 5300 Series 3-19...
Page 50
Rosemount 5300 Series L>46.5 in. (1180 mm) 20.5 < L<46.5 in. (520 mm< L <1180 mm) 15.7< L<20.5 in. (400< L<520 mm) 3-20 Rigid Twin Lead The spacers are put closer together at the probe end. The maximum amount that can be cut is related to the ordering length L.
Page 51
Coaxial Centering piece Maximum shortening 23.6 in. (600 mm) Minimum probe length 15.7 in. (400 mm) Rosemount 5300 Series To cut a coaxial probe: NOTE! The HTHP coaxial probe must not be cut in field. 1. Insert the centering piece.
Rosemount 5300 Series Anchoring Ring Magnet 1.1 in. (28 mm) 3-22 In turbulent tanks, it may be necessary to fix the probe. Depending on the probe type, different methods can be used to guide the probe to the tank bottom. This may be needed to prevent the probe from hitting the tank wall or other objects in the tank, as well as preventing a probe from breaking.
Page 53
Reference Manual 00809-0100-4530, Rev BA July 2009 Drain Ø 0.3 in. (8 mm) Rosemount 5300 Series Coaxial probe. The Coaxial probe can be guided by a tube welded on the tank bottom. Tubes are customer supplied. Make sure that the probe can move freely in order to handle thermal expansion.
Page 54
Rosemount 5300 Series 3-24 Solid Applications Pull the probe rope through a suitable anchoring point, e.g. a welded eye and fasten it with two clamps. It is recommended that the probe is slack in order to prevent high tensile loads.
5.55 in. (141 mm) 8 in. 7.4 in. (188 mm) Mounting a Centering Disc on Flexible Single Probes Centering disc Tab washer Weight Bolt Tab washer Rosemount 5300 Series Pipe Schedule 40s,40 80s,80 2 in. 3 in. 2 in. 4 in. 4 in.
Page 56
Rosemount 5300 Series Table 3-9. Hole requirements on probes for mounting a centering disc. Figure 3-14. Use drill template for proper holes. Washer 3-26 Mounting a Centering Disc on Rigid Single Probes The mounting of a centering disc on a rigid single lead probe requires one hole (8 mm probe) or two holes (13 mm probe) at certain distances from the end of the lead.
Page 57
The clearance distance of 1 in. (25 mm) is selected with a dome shaped chamber bottom in mind, which may prevent the centering disk from reaching the bottom. Rosemount 5300 Series 1. Mount the centering disc at the end of the probe.
Page 58
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 3-28...
Process leaks could result in death or serious injury. Make sure that the transmitter is handled carefully. If the Process Seal is damaged, gas might escape from the tank if the transmitter head is removed from the probe. Rosemount 5300 Series ). Please...
High voltage that may be present on leads could cause electrical shock: Avoid contact with leads and terminals. Make sure the main power to the Rosemount 5300 transmitter is off and the lines to any other external power source are disconnected or not powered while wiring the gauge.
Use shielded twisted pair wiring for the Rosemount 5300 Series to comply with EMC regulations. The cables must be suitable for the supply voltage and approved for use in hazardous areas, where applicable. For instance, in the U.S., explosion-proof conduits must be used in the vicinity of the vessel.
Figure 4-2. Non-hazardous installations. Figure 4-3. Intrinsically Safe installations. Terminals in the transmitter housing provide connections for signal cables. The Rosemount 5300 transmitter is loop-powered and operates with the following power supplies: Hazardous approval Non-Hazardous Installations and Intrinsically Safe Installations...
Connecting the Transmitter Figure 4-4. Terminal compartment and external ground screw. To connect the Rosemount 5300 Series transmitter: 1. Make sure the power supply is switched off. 2. Remove the terminal block cover. 3. Pull the cable through the cable gland/conduit. Install wiring with a drip loop.
Rosemount 5300 Series Non-Intrinsically Safe Output Figure 4-5. Wiring diagram for non-intrinsically safe installations (HART). AMS Suite Power Supply With non-intrinsically safe power supply in non-hazardous installations or Explosion-proof/Flameproof installations, wire the transmitter as shown in Figure 4-5. NOTE! Make sure that the power supply is off when connecting the transmitter.
For maximum load resistance see Figure 4-3. The power supply voltage ranges from 16 Vdc to 30 Vdc. IS parameters Ui=30 V. Ii=130 mA. Pi=1 W. Ci=7.26 nF. Li=0. (1) See Section B: Product Certifications for more information. Rosemount 5300 Series Rosemount 5300 Series Radar Transmitter...
Power Requirements Connecting the Transmitter Terminals in the transmitter housing provide connections for signal cables. The Rosemount 5300 transmitter is powered over F standard fielbus power supplies. The transmitter operates with the following power supplies: Approval Type Power Supply (Vdc)
Page 67
Shield Wire Ground To protect the fieldbus segment from noise, grounding techniques for shield wire usually require a single grounding point for shield wire to avoid creating a ground loop. The ground point is typically at the power supply. Rosemount 5300 Series...
Page 68
Rosemount 5300 Series Figure 4-8. Rosemount 5300 Radar Transmitter field wiring. 4-10 Connecting Fieldbus Devices Integrated Power Conditioner and Filter Power Supply (The power supply, OUNDATION filter, first Fieldbus terminator, and Configuration configuration tool Tool are typically located in the control room.) *Intrinsically safe installations may allow fewer devices per I.S.
July 2009 Non-Intrinsically Safe Output Figure 4-9. Wiring for non-intrinsically safe power supply (Foundation Fieldbus). Rosemount 5300 Series Radar Transmitter With non-intrinsically safe power supply in Non-hazardous installations or Explosion-proof/Flameproof installations, wire the transmitter as shown in Figure 4-9. NOTE! Make sure that the power supply is off when connecting the transmitter.
Rosemount 5300 Series Intrinsically Safe Output Figure 4-10. Wiring diagram for intrinsically safe power supply (Foundation Fieldbus). Rosemount 5300 Series Radar Transmitter 4-12 When your power supply is intrinsically safe, wire the transmitter as shown in Figure 4-10. NOTE! Make sure the instruments in the loop are installed in accordance with intrinsically safe field wiring practices.
Figure 4-11. Wiring diagram for HART Tri-Loop. Rosemount 5300 Series Radar Transmitter The Rosemount 5300 transmitter outputs a HART signal with four process variables. By using the Model 333 HART Tri-Loop, up to three additional analog 4-20 mA outputs are provided. DIN Rail Mounted HART Tri-Loop Ch.
Rosemount 5300 Series 751 Field Signal Indicator Figure 4-12. Wiring diagram for a Rosemount 5300 transmitter with 751 Field Signal Indicator. 4-14 Rosemount 5300 Series Radar Transmitter Reference Manual 00809-0100-4530, Rev BA July 2009 Model 751 Field Signal Indicator Power supply...
Use the equipment only as specified in this manual. Failure to do so may impair the protection provided by the equipment. Do not perform any service other than those contained in this manual unless you are qualified. Rosemount 5300 Series ). Refer to...
LCD Configuration Advanced Configuration Configuration Tools The configuration of a Rosemount 5300 transmitter is normally a simple and straight-forward task. The complete configuration of a Rosemount 5300 transmitter includes Basic Configuration, Echo Tuning and Advanced Configuration. This section describes Basic Configuration.
Page 75
(part number 03095-5108-0001 for PCMCIA) is required. For Fieldbus communication you will also need the National OUNDATION Instruments Communication Manager software (see “Installing the RRM Software for FOUNDATION Fieldbus” on page 5-16). Rosemount 5300 Series based software package including Fieldbus OUNDATION...
Figure 5-2. Upper Reference Point Upper Reference Point This section describes basic configuration parameters for a Rosemount 5300 transmitter. Basic configuration is only needed for the 5300 Series transmitters which are not pre-configured at the factory. Factory configuration is normally specified in the Configuration Data Sheet.
Page 77
This parameter is pre-configured at the factory. It must be changed if the probe is shortened. Probe Type The transmitter is designed to optimize measurement performance for each probe type. This parameter is pre-configured at the factory. This value needs to be changed if the probe type is changed. Rosemount 5300 Series...
As a default standard, a Rosemount 5300 transmitter is able to track level changes of up to 1.5 in./s (40 mm/s). When the Rapid Level Changes check box is marked, the transmitter can track level changes of up to 8 in./s (200 mm/s).
Using only 3 strapping points results in a level-to-volume profile that is more angular than the actual shape. Using 10-15 of the points at the bottom of the tank yields a level-to-volume profile that is similar to the actual tank bottom. Rosemount 5300 Series...
Page 80
Rosemount 5300 Series Figure 5-4. Standard tank shapes Standard Tank Shapes Diameter Height Diameter Height Height Diameter Diameter Height Diameter Reference Manual 00809-0100-4530, Rev BA July 2009 Vertical Cylinder Vertical Cylinder tanks are specified by Diameter, Height, and Volume Offset.
(This parameter may be used if there are measurement problems in the upper part of the tank, see “Handling of Disturbances from Nozzle” on page C-4). The UNZ is equal to zero in the default configuration. Rosemount 5300 Series 20 mA Upper Range Value (URV) = 100% 4 mA Lower Range Value...
Page 82
Rosemount 5300 Series Table 5-2. Analog Output: Standard Alarm Values vs. Saturation Values Table 5-3. Analog Output: NAMUR-Compliant Alarm Values vs. Saturation Values 5-10 Alarm Mode The Alarm mode specifies the analog output state when there is a failure or a measurement error: High: the output current is set to the High Alarm Limit.
5. Restart the transmitter. HART command: [3, 2, 1, 1]. See also “Guided Setup” on page 5-19 for further information on configuration of the Rosemount 5300 transmitter. Rosemount 5300 Series Navigation Keys Enter Key...
Page 84
Rosemount 5300 Series Figure 5-7. HART Communicator Menu Tree corresponding to Device Revision 3. Process Variables 1 Primary Variable 2 2nd 3 3rd 4 4th 5 All variables 6 Signal Quality Metrics 7 Identification Setup 1 Process Variables 2 Setup...
System Requirements Help In RRM The Rosemount Radar Master (RRM) is a user-friendly software tool that allows the user to configure the Rosemount 5300 Series transmitter. Choose either of the following methods to configure a Rosemount 5300 Series transmitter with RRM: •...
Rosemount 5300 Series Installing the RRM software for HART communication RRM communicates with the transmitter No communication with the transmitter 5-14 To install the Rosemount Radar Master: 1. Insert the installation CD into your CD-ROM drive. 2. If the installation program is not automatically started, choose Run from the Windows Start bar.
4. Click the right mouse button on the selected COM port and choose Properties. 5. Select the Port Settings tab and click the Advanced button. 6. Drag the Receive Buffer and Transmit Buffer slides to 1. 7. Click the OK button. 8. Restart the computer. Rosemount 5300 Series 5-15...
Rosemount 5300 Series Installing the RRM Software for FOUNDATION Fieldbus 5-16 To install the Rosemount Radar Master for F communication: 1. Start by installing the National Instruments Communication Manager software. See National Instruments manual (Getting started with your PCMCIA-FBUS and the NI-FBUS™ software) for more information.
RRM program is installed. Units can also be changed as follows: 1. From the View menu, choose the Application Preferences option. 2. Select the Measurement Units tab. 3. Choose the desired units for Length, Level Rate, Volume, and Temperature. Rosemount 5300 Series 5-17...
Analog Output Echo Curve Advanced 5-18 Use the Setup function if you are already familiar with the configuration process for the Rosemount 5300 Series transmitter or for changes to the current settings: Reference Manual 00809-0100-4530, Rev BA July 2009 1. Start the RRM software.
Sequence) and F Fieldbus parameters are also shown. OUNDATION The Guided Setup is useful if you are unfamiliar with the Rosemount 5300 Series transmitter. 1. Start the Guided Setup Start RRM. It automatically presents a list of available transmitters. Select the desired transmitter.
Rosemount 5300 Series Device Properties General Information 5-20 3. Check the device properties The first window in the configuration wizard presents general information stored in the transmitter database such as device model, serial number, probe type, probe length, communication protocol and device address.
Probe Type from the list. The Rosemount 5300 Series transmitter automatically makes some initial calibrations based on the chosen Probe Type. The following Probe Types are available: •...
Level value offset error. When specifying the Tank Height, keep in mind that this value is used for all level and volume measurements performed by the Rosemount 5300 Series transmitter. The Tank Height must be set in linear (level) units, such as feet or meters, regardless of primary variable assignment.
By default, the Upper Product Dielectric parameter is about 2. Set the Upper Product Dielectric Constant to a value that corresponds to the current product. Rosemount 5300 Series 5-23...
Page 96
Rosemount 5300 Series 5-24 Rosemount Radar Master (RRM) includes tools to estimate the dielectric constant of the current product: • The Dielectric Chart lists the dielectric constant of a large number of products. The Dielectric Chart can be opened with one of the following methods: - Choose the View>Dielectric Constant Chart menu option...
Tank height/length (not for spherical tanks) • Volume Offset: use this parameter if you do not want zero volume and zero level to match (for example if you want to include a volume below the zero level) Rosemount 5300 Series 5-25...
Rosemount 5300 Series Analog Output (HART) Finish Configuration Wizard 5-26 9. Configure the Analog Output HART command: [2,1, 6]. Analog Output is not available for F Typically, the Primary Variable (PV) is configured to be Product Level, Interface Level, or Volume. Other variables like Product Distance, Interface Distance, Upper Product Thickness, etc.
Trim Near Zone is described further in “Handling of Disturbances from Nozzle” on page C-4. Probe End Projection is described further in “Probe End Projection” on page C-10. Vapor Compensation is described further in “Dielectric Constant Settings” on page C-14. Choose the Tank Material if recommended. Rosemount 5300 Series 5-27...
Rosemount 5300 Series Restart the Transmitter View Measured Values Backup 5-28 13. Restart the transmitter When the transmitter is configured, it should be restarted to make sure that all configuration changes are properly activated and the transmitter performs as expected. It may take up to 60 seconds after the restart button is pressed until measurement values are updated.
Page 101
5300 in a similar tank since the file can be directly uploaded to a new device • restoring the configuration, if for any reason, configuration data is lost or accidentally modified making the device inoperable The Configuration Report window appears automatically when the backup is completed. Rosemount 5300 Series 5-29...
BASIC CONFIGURATION USING AMS SUITE (HART) 5-30 The Rosemount 5300 transmitter can be configured by using the AMS Suite software. 1. Start the AMS Device Manager and make sure that the transmitter connects. 2. In the Device Connection View, click the right mouse button on the transmitter icon.
July 2009 BASIC CONFIGURATION USING DELTAV The Rosemount 5300 Series supports DD Methods for DeltaV to facilitate transmitter configuration. The following description shows how to use DeltaV with the AMS application to configure the Rosemount 5300 Series. The corresponding F Fieldbus commands are also shown.
Page 104
Probe Type from the list. The Rosemount 5300 Series transmitter automatically makes some initial calibrations based on the chosen Probe Type. The following probes are available: •...
Page 105
Fieldbus parameter: OUNDATION TRANSDUCER 1100>GEOM_TANK_HEIGHT TRANSDUCER 1100>MOUNTING_TYPE TRANSDUCER 1100>PIPE_DIAMETER TRANSDUCER 1100>NOZZLE_HEIGHT 12. Select the Environment tab. 13. Normally, the Measurement Mode does not need to be changed. The transmitter is pre-configured according to the specified model. Rosemount 5300 Series 5-33...
Page 106
Rosemount 5300 Series 5-34 Submerged is used for applications where the probe is fully submerged in liquid. In this mode, the transmitter ignores the upper product level. See “Section 7: Interface Measurements with Fully Submerged Probes” for more information. NOTE! Only use Submerged for applications where the interface is measured with a fully submerged probe.
Page 107
Calculation Method: Fieldbus parameter: OUNDATION TRANSDUCER 1300>VOL_VOLUME_CALC_METHOD Diameter: Fieldbus parameter: OUNDATION TRANSDUCER 1300>VOL_IDEAL_DIAMETER Tank Length: Fieldbus parameter: OUNDATION TRANSDUCER 1300>VOL_IDEAL_LENGTH Volume Offset: Fieldbus parameter: OUNDATION TRANSDUCER 1300>VOL_VOLUME_OFFSET See “Volume Configuration” on page 5-7 for more information. Rosemount 5300 Series 5-35...
Rosemount 5300 Series, refer to the F Block Manual (Document No. 00809-0100-4783). A Rosemount 5300 Series transmitter is shipped with a blank tag and a temporary address (unless specifically ordered with both) to allow a host to automatically assign an address and a tag.
Resource Block The Resource block contains diagnostic, hardware, electronics, and mode handling information. There are no linkable inputs or outputs to the Resource Block. Rosemount 5300 Series 5-37...
Page 110
Level Transducer Block, Appendix G: Register Transducer Block, Appendix H: Advanced Configuration Transducer Block, Appendix I: Resource Transducer Block and Appendix J: Analog-Input Block. Function Block Summary The following function blocks are available for the Rosemount 5300 Series: • Analog Input (AI) •...
CHANNEL Select the channel that corresponds to the desired sensor measurement. The Rosemount 5300 measures Level (channel 1), Distance (channel 2), Level Rate (channel 3), Signal Strength (channel 4), Volume (channel 5), Internal Temperature (channel 6), Upper Product Volume (channel 7), Lower Product...
Page 112
Rosemount 5300 Series 5-40 Indirect Square Root Select indirect square root when the desired output is an inferred measurement based on the transmitter measurement and the relationship between the sensor measurement and the inferred measurement is square root (e.g. level).
Page 113
OUT_SCALE that are supported by the device. The supported units are: Display Display ft/s in/m Display °C °F Display Display Gallon ImpGall Rosemount 5300 Series Description meter centimeter millimeter feet inch Description meter per second meter per hour feet per second inch per minute Description...
Rosemount 5300 Series Application Example 1 Figure 5-12. Situation Diagram Table 5-9. Analog Input Function Block Configuration for a typical Level Transmitter Figure 5-13. Analog Input Function Block Diagram for a typical Level Transmitter 5-42 Radar Level Transmitter, Level Value A level transmitter is measuring the level in a 33 ft.
Table 5-10 lists the appropriate configuration settings, and Figure 5-15 illustrates the correct function block configuration. Parameter Configured Values L_TYPE Indirect XD_SCALE 0 to 14 m OUT_SCALE 0 to 100% CHANNEL CH1: Level Level Measurement - percent OUT_D AI Function Block Rosemount 5300 Series 0 to 100% 5-43...
Rosemount 5300 Series Application Example 3 Figure 5-16. Situation Diagram Table 5-11. Analog Input Function Block Configuration for a Level and Interface Transmitter Figure 5-17. Analog Input Function Block Diagram for a Level and Interface Transmitter 5-44 Radar Level Transmitter, Product Level and Interface Level Value A level transmitter is measuring the product level and the interface level in a 33 ft.
HART burst signal into three additional 4-20 mA analog signals. To set the Rosemount 5300 transmitter up for the HART Tri-Loop: 1. Make sure the Rosemount 5300 transmitter is properly configured. 2. Assign transmitter variables Primary Variable, Secondary Variable etc.
Page 118
Rosemount 5300 Series Figure 5-18. Tri-Loop wiring. 5-46 7. Configure Tri-Loop Channel 1: a. Assign variable: Tri-Loop HART command [1,2,2,1,1]. Make sure that the SV, TV, and QV match the configuration of the 5300 transmitter. b. Assign units: Tri-Loop HART command [1,2,2,1,2]. Make sure that the same units are used as for the 5300 transmitter.
To change the poll address using the Rosemount Radar Master (RRM) software: 1. Choose the Setup>General option. 2. Select the Communication tab. 3. Set the desired address (between 1 and 15 for multidrop operation). 4. Click the Store button to save the new address. Rosemount 5300 Series 5-47...
Page 120
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 5-48...
Equipment ratings and certifications are no longer valid on any products that have been damaged or modified without the prior written permission of Emerson Process Management. Any continued use of product that has been damaged or modified without prior written authorization is at the customer's sole risk and expense.
Using the Display Panel Figure 6-1. The 5300 Display Panel. The Rosemount 5300 transmitter uses an optional Display Panel for presentation of measurement data. When the transmitter is switched on, the Display Panel presents information such as transmitter model, measurement...
3. Select the variables to be shown on the display panel. The LCD will alternate between the selected items. The available LCD parameters are listed in Table 6-1 on page 6-6. 4. Click the Store button to save the LCD settings in the transmitter database. Rosemount 5300 Series...
Page 124
Rosemount 5300 Series Figure 6-3. The LCD tab in the AMS Suite configuration window lets you configure parameters to be presented on the Display Panel. Using AMS The LCD tab in the Configure/Setup window lets you specify which variables to view on the Display Panel screen: 1.
Page 125
LCD Unit. The LCD will alternate between the selected display parameters. The available LCD parameters are listed in Table 6-1 on page 6-6. 5. Click the OK button to save the LCD settings in the transmitter database. Rosemount 5300 Series...
Page 126
Rosemount 5300 Series Table 6-1. LCD parameters and presentation on display Parameter Level Distance Level Rate Signal Strength Volume Internal Temperature Analog Output Current Percent Range Interface Level Interface Distance Interface Level Rate Interface Signal Strength Volume Lower Volume Upper...
To view measurement data such as Level, Signal Strength, etc. in Rosemount Radar Master, choose the Tools>Device Display option and select the Level tab: To view the Analog Output signal, choose the Tools>Device Display option and select the Analog Out tab: Rosemount 5300 Series...
Rosemount 5300 Series Viewing Measurement Data in AMS Suite Figure 6-7. Presentation of measurement data in AMS Suite. To view measurement data such as Level, Signal Strength, etc. in the AMS Suite: 1. Select the transmitter icon in the AMS Suite Device Connection View window.
Viewing Measurement Data in DeltaV Figure 6-8. Presentation of measurement data in DeltaV for the Rosemount 5300. Rosemount 5300 Series 1. Click the right mouse button on the transmitter icon and choose the Properties option. 2. Select the Transducer1100 block.
Page 130
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 6-10...
Information that raises potential safety issues is indicated by a warning symbol ( refer to the following safety messages before performing an operation preceded by this symbol. Rosemount 5300 Series ). Please...
High voltage that may be present on leads could cause electrical shock. Avoid contact with leads and terminals. Make sure the main power to the Rosemount 5300 Transmitter is off and the lines to any other external power source are disconnected or not powered while wiring the gauge.
When the surface echo is found, the next pulse below the product surface and with a signal strength above the Interface Threshold, is considered as the Interface. The following amplitude thresholds are used for the Rosemount 5300 Series transmitter: Reference Threshold - amplitude threshold for detection of the Reference pulse.
Page 134
Rosemount 5300 Series Surface Threshold - amplitude threshold for detection of the Product level peak. The surface threshold is designed as a number of individually adjustable amplitude threshold points, the Amplitude Threshold Curve (ATC), see “Amplitude Threshold Curve” on page 7-7.
Rosemount 5300 Series Figure 7-3. Example 2: surface threshold is too low. Disturbing echo misinterpreted as product surface Figure 7-4. Echo Curve after surface threshold was adjusted Probe End Projection Actual surface Distance, m By adjusting the surface threshold the product surface is properly detected as...
Disturbing object Distance, m In addition to using the ATC, the Rosemount 5300 transmitter supports alternative methods to filter out disturbances at the top of the tank, such as the Trim Near Zone and the Upper Null Zone functions, which can be used to handle disturbances from narrow nozzles or nozzles with rough edges.
Rosemount 5300 Series INTERFACE PULSE NOT FOUND Figure 7-6. Echo Curve plot indicating that the amplitude threshold for the interface peak is too high. In interface applications where the bottom product has a relatively low dielectric constant (<40), or if the signal is attenuated in the upper product, the amplitude of the reflected signal is relatively low and difficult for the transmitter to detect.
Signal Quality Metrics can give an indication of how good the surface signal is compared to the noise. See Signal Quality Metrics page C-18 in Appendix C: Advanced Configuration. Rosemount 5300 Series Interface Threshold Surface Threshold...
Rosemount 5300 Series USING THE ECHO CURVE ANALYZER Using the Rosemount Radar Master Figure 7-8. The Echo Curve function is a useful tool for signal analysis. Device Config Tools 7-10 The Echo Curve Analyzer in Rosemount Radar Master (RRM) shows the measurement signal amplitude from the top to the bottom of the tank.
Measure and Learn The Measure and Learn function in RRM automatically creates an Amplitude Threshold Curve (ATC) which is used by the Rosemount 5300 transmitter to find the surface pulse. The ATC is adapted to the shape of the measurement signal as described in “Disturbance Echo Handling”...
Rosemount 5300 Series The View/Record ModeTab Figure 7-10. An Echo Curve plot in View/Record mode. 7-12 The current level Measurement Output from the device is presented with an arrow at the top of the plot. Normally, the measurement output points directly at the surface echo peak,...
The Echo Curve plot also shows an ATC if available. See “Threshold Settings” on page 7-13 for information on how to create an ATC with the Measure and Learn function. Threshold Settings To adjust the amplitude thresholds: Rosemount 5300 Series 7-13...
Page 144
Rosemount 5300 Series 7-14 1. Select HART command [2, 5, 2]. Fieldbus parameter: OUNDATION TRANSDUCER 1300>PROBE_END_THRESH TRANSDUCER 1300>REFERENCE_THRESH TRANSDUCER 1300>INTERFACE_THRESH TRANSDUCER 1300>FULL_TANK_THRESH_OFFSET The different threshold options appear on the display: 2. Open the desired option. For example, choosing option 2 Surface Threshold Settings displays the following screen: 3.
Figure 7-11. Interface Level measurements in a full chamber. The Rosemount 5300 series has a measurement option which makes it possible to handle interface measurements when the probe is fully submerged into the upper product, see Figure 7-11, and only the interface level is detected by the transmitter.
Rosemount 5300 Series ANALOG OUTPUT CALIBRATION 7-16 This function lets you calibrate the Analog Output by comparing the actual output current with the nominal 4 mA and 20 mA currents. Calibration is done at the factory and normally the transmitter does not need to be recalibrated.
Enter the desired value in the Calibration Distance field and click the Store button. Level calibration 1. Measure the actual Product Level. 2. Adjust the Tank Height so the product level measured by the transmitter corresponds to the actual product level. Rosemount 5300 Series 7-17...
Page 148
Rosemount 5300 Series Figure 7-12. Distance and Level calibration 7-18 Reference Point Reference Point Distance Reference Gauge Height Reference Manual 00809-0100-4530, Rev BA July 2009 Level...
4. Enter the update rate. An update rate of for example 10 seconds means that the plot will be updated every 10 seconds. 5. Click the Start Log button. The logging will proceed until stopped by clicking the Stop Log button. Rosemount 5300 Series 7-19...
Rosemount 5300 Series BACKING UP THE TRANSMITTER CONFIGURATION Figure 7-14. It is recommended that the transmitter configuration is stored in a backup file. Figure 7-15. The configuration backup file can be viewed in a word processor program. 7-20 Use this RRM option to make a backup copy of the configuration parameters in the transmitter database.
The report compares a specified backup file with the default transmitter configuration. To open the Configuration Report, choose the Tools>Configuration Report menu option: Information on probe type, software versions, software and hardware configuration, and unit code is presented. Rosemount 5300 Series 7-21...
Rosemount 5300 Series RESET TO FACTORY SETTINGS Figure 7-17. The Reset to Factory Settings window in RRM. TRANSDUCER1100 7-22 This function resets all or a specific part of the holding registers to factory settings. NOTE! It is recommended that a backup of the configuration is done before the factory reset.
To open the Diagnostics window in RRM, choose the Diagnostics option from the Tools menu: To view the Diagnostics window in AMS Suite, click the right mouse button on the desired transmitter, and choose the Device Diagnostics option: Rosemount 5300 Series 7-23...
Page 154
Rosemount 5300 Series TRANSDUCER1100 Figure 7-20. The Status window in DeltaV shows various status information. 7-24 DeltaV 1. In the DeltaV Explorer, select the desired transmitter icon, and click the right mouse button on the Transducer 1100 block icon: 2. Choose the Status option.
HART Command: [3, 2, 1, 3]. DeltaV: 1. In the DeltaV Explorer, select the desired transmitter icon, and click the right mouse button on the Transducer 1300 block icon: 2. Choose the Simulation Mode option. Rosemount 5300 Series Simulation Mode 7-25...
2. Choose the Unlock/Lock Device option. In RRM, service functions for advanced users are available for the Rosemount 5300 Series transmitter. Setting RRM into Service Mode, all the Service menu options in RRM are enabled. The default password for enabling the Service Mode is “admin”.
3. Choose the View Input/Holding Registers option from the Service menu. 4. Click the Read button. To change a Holding register value, just type a new value in the corresponding Value field. The new value is not stored until the Store button is clicked. Rosemount 5300 Series 7-27...
Rosemount 5300 Series REMOVING THE TRANSMITTER HEAD FLANGE VERSION Put the protection plug here! THREADED VERSION Put the protection plug here! 7-28 1. Loosen the nut that connects the transmitter housing to the Process Seal. Carefully lift the transmitter head.
Dynamic Vapor Compensation function is supported by the transmitter head. Compatibility Between Probe Types and Firmware Versions Standard HP / HTHP without R2 marking Rosemount 5300 Series Probe Type HP / HTHP / C with HTHP with both R2 marking only R2 and VC marking...
Rosemount 5300 Series Check Firmware and Probe Version 7-30 1. Check the manufacturing date on the transmitter head label. Manufacturing Date Before 080618 (YYMMDD) 2. Check the R2, VC marking on the probe. NOTE! In RRM, the software revision number can be checked either in Device Explorer, or at the bottom of the RRM window, as shown in the screenshot below.
Reference Manual 00809-0100-4530, Rev BA July 2009 Changing the Probe Rosemount 5300 Series Transmitter head Process seal Probe 1. Loosen the nut. 2. Remove the transmitter head from the old probe. 3. On the new probe, make sure that the protection plug is removed and the upper surface of the Process Seal is clean.
Page 162
Rosemount 5300 Series Figure 7-23. Product level lowered beneath the Near Zone. Table 7-2. Near Zone definitions depending on software releases and probe type 7-32 Near Zone 5300 GWR Series Near Zone Firmware Version Rigid Probes Firmware version earlier than 1.A4 Firmware version 1.A4 or...
• Surface pulse and Interface pulse are very close • Wrong Probe Type set • Bad Reference Threshold value Rosemount 5300 Series Action • Check the power supply • Check the cables for serial data communication • View the Diagnostic window, see “Diagnostics”...
Page 164
SW Error/ Display Error/ Analog Out Error (1) A malfunctioning display panel may only be replaced by service personnel at the Emerson Process Management Service Department. A display must not be replaced when the transmitter is in operation. (2) If the 5300 transmitter has been exposed to temperatures outside the specified limits, the device may stop its normal operation.
Write protection jumper on the display is enabled. The factory default configuration is used. Probe is not detected. Rosemount 5300 Series Action Contact Emerson Process Management Service Department. See Warning Messages for details. See Error Messages for details. Turn off the simulation mode in RRM, select Tools, Simulation Mode, and click the Stop button.
Rosemount 5300 Series Errors Table 7-5. Error messages. Message RAM error FPROM error Database (Hreg) error Microwave Module error LCD error Modem error Analog out error Internal temperature error Other hardware error Measurement error Configuration error Software error 7-36 Error messages that may be displayed on the Integral Display, on a 375 Handheld Communicator, in AMS, or in RRM are shown in Table 7-5.
Analog Output alarms. In RRM, warnings are indicated in the Diagnostics window. Description See Diagnostics (RRM: Tools>Diagnostics) for further information on a warning message. See also “Diagnostics” on page 7-23. Rosemount 5300 Series Action 7-37...
Rosemount 5300 Series Measurement Status Table 7-7. Measurement status. Message Full tank Empty tank Probe missing Seal contaminated Reference pulse calculated Reference pulse invalid DeltaF not at setpoint Tank signal clip warning No surface echo found Predicted level No reference echo...
Page 169
Reference echo present Sudden level jump detected Nearzone echo present Nonlinear gain used Nearzone measurement Rosemount 5300 Series Description The simulation mode is active. The given measurements are simulated. The advanced simulation mode is active. The given lower volume value is invalid.
Rosemount 5300 Series Interface Status Table 7-8. Interface status. Message Interface not OK Interface not found Can't measure interface on horizontal probe Can't handle max possible interface thickness Interface thickness close to max range Interface set to max thickness No interface echo found. The upper...
The Primary variable is out of range. The configured span is too small. The given Upper and Lower Sensor Limits are invalid. Rosemount 5300 Series Action Check Error and Warning Messages to find the reason for the Alarm. No action needed.
Rosemount 5300 Series LCD ERROR MESSAGES Figure 7-24. The 5300 Display Panel displaying an error message. Table 7-11. Error messages displayed on the 5300 Display Panel. 7-42 Error Message An error in the gauge data memory (RAM) has been RAM FAIL detected during the startup tests.
LED for presentation of error messages. Table 7-12. LED error codes. For Rosemount 5300 transmitters without display, a flashing Light Emitting Diode (LED) is used for presentation of error messages. Flashing LED In normal operation, the LED flashes once every other second. When an error occurs, the LED flashes a sequence that corresponds to the Code number followed by a five second pause.
July 2009 Recommended Action 1. Restart processor 2. Check display connection 3. Call service center 1. Restart processor 2. Check Rosemount 5300 cable 3. Call service center 1. Restart processor 2. Call service center 1. Restart processor 2.Call service center 1.
Simulate Active: Simulation is enabled and the block is using a simulated value in its execution Input Failure/Process Variable has Bad Status: The hardware is bad, or a bad status is being simulated Power Up Out of Service: The actual mode is out of service Rosemount 5300 Series 7-45...
Page 176
Rosemount 5300 Series Table 7-19. Troubleshooting the AI block Symptom Bad or no level readings (Read the AI “BLOCK_ERR” parameter) OUT parameter status reads UNCERTAIN and substatus reads EngUnitRangViolation Mode will not leave OOS Process and/or block alarms will not work Features...
Electrical shock could cause death or serious injury. • Use extreme caution when making contact with the leads and terminals. Rosemount 5300 Series...
Page 178
Equipment ratings and certifications are no longer valid on any products that have been damaged or modified without the prior written permission of Emerson Process Management. Any continued use of product that has been damaged or modified without prior written authorization is at the customer's sole risk and expense.
Refer to the 5300 FMEDA report for failure rate data, assessment details, and assumptions regarding failure rate analysis. Table 8-1 lists the versions of the Rosemount 5300 Series transmitter that have been considered for the hardware assessment, to which this section applies.
Use a HART-compliant master, such as Rosemount Radar Master or 375 Field Communicator, to communicate with and verify configuration of the Rosemount 5300. A full review of configuration methods is available in Section 5: Configuration. These instructions are applicable to the 5300 Prior-Use option with any differences noted.
The required proof test intervals are dependant on the configuration of the transmitter and the process environment. The Rosemount 5300 is designed to have a 5-year proof test interval assuming it represents the typical 35% of the SIF PFD of the operator/owner of the system to determine the sufficient time interval and verify it is followed.
Page 182
Rosemount 5300 Series If the overfill and empty tank protection function cannot be tested by a controlled filling to the response height, suitable simulation of the level or of the physical measuring effect, e.g. by shorting the probe, must be used to make the level sensor respond.
Feedback can be submitted electronically at www.emersonprocess.com/rosemount/safety (Contact Us). The Rosemount 5300 must be operated in accordance with the functional and performance specifications provided in Appendix A: Reference Data. The FMEDA report includes failure rates. The full report is accessible at www.emersonprocess.com/rosemount/safety/PriorUse.htm.
Page 184
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009...
Appendix A SPECIFICATIONS General Product Rosemount 5300 Series Guided Wave Radar Level and Interface Transmitter; Model 5301, Liquid Level or Interface Transmitter (interface available for fully submerged probe). Model 5302 Liquid Level and Interface Transmitter. Model 5303 Solids Level Transmitter.
Page 186
Rosemount 5300 Series Display / Configuration (continued) Fieldbus Link Master (LAS). OUNDATION Class (Basic or Link Master) Fieldbus AI-block: 30 ms. PID-block: 40 ms. OUNDATION Block Execution Time ARTH-, ISEL-, OSPL-block: 65 ms. CHAR-block: 75 ms. Fieldbus OUNDATION Instantiation Conforming F ITK 5.0.
Page 187
OUNDATION (5) PFA is a fluoropolymer with properties similar to PTFE. (6) 1 mm PTFE cover. (7) Final rating may be lower depending on flange and O-ring selection. Rosemount 5300 Series Fieldbus: -58°F to 140°F (-50°C to 60°C). OUNDATION ±...
Rosemount 5300 Series Process Temperature and Pressure Rating Figure A-1. Maximum process temperature and pressure diagram for standard tank connections. The tank connection consists of a tank seal, a flange BSP/G threads . See “Ordering Information” on page A-21). Flange mating face dimensions follow ANSI B 16.5, JIS B2220, and EN 1092-1 standards for blind flanges.
Page 189
PFA/PTFE for HP/C, and ceramics for HTHP. Ceramic spacers allow for usage in applications with higher temperature. The following table gives the temperature ranges for tank seal with different O-ring material (applicable for the Standard Tank Connection): Rosemount 5300 Series HTHP tank connection Temperature °F (°C)
Rosemount 5300 Series Table A-1. Temperature range for different tank seal material. Flange Connection Rating Bolting material ANSI Stainless steel SA193 Soft (1a) with min. B8M C1.2 thickness 1.6 mm. EN 1515-1/-2 group Soft (EN 1514-1) with 13E0, A4-70. min. thickness 1.6 mm.
For Alloy C-276 HTHP/HP probes with flange plate design is available up to Class 600/PN 63. When the Rosemount 5300 is installed in high or low temperature applications, it is important that the maximum/minimum ambient temperature is considered. Nozzle insulation for the HTHP version should not exceed 4 in.
7.4 (188.5) Alloy probe and protective plate HTHP/HP Plate Design (Option for Alloy versions) 15.6 (397.5) Dimensions are in inches (millimeters). Rosemount 5300 Series PTFE covered probe and protective plate The PTFE and Alloy probes are designed with a protective plate.
Page 194
Rosemount 5300 Series Figure A-7. Rigid Single Lead Probe with Tri-Clamp Connection 3.4 (87) 3.6 (92) 10.1 (257.5) 10 ft. (3 m) for Ø 0.31 (8) 20 ft. (6 m) for Ø 0.51 (13) 10.1 (257.5) A-10 7.1 (180) ½...
Page 195
1.1 (27) Ø 0.31 (8) or 0.51 (13): SST and Alloy probes Ø 0.47 (12): PTFE covered probe Dimensions are in inches (millimeters). Rosemount 5300 Series NPT 1/1½/2 inch 5.2 (133) 7.4 (188.5) NPT 1½, G 1½ inch HTHP/HP/C version...
Page 196
Rosemount 5300 Series Figure A-9. Flexible Single Lead Probe with Flange Connection 7.1 (180) 3.4 (87) 3.6 (92) 10.1 (257.5) 164 feet (50 m) Ø 0.16 (4): SST probe Ø 0.24 (6): SST probe Ø 0.28 (7): PTFE covered probe HTHP/HP/C version 15.6 (397.5)
Page 197
1.10 (28): 6 mm SST probe The PTFE covered probe is designed with a protective plate. Dimensions are in inches (millimeters). Rosemount 5300 Series Short weight (option W2) 2 (50): 4 mm SST probes 1.5 (37.5): 4 mm SST probes...
Page 198
Rosemount 5300 Series Figure A-11. Flexible Single Lead with threaded connection NPT 1/1½/2 inch 7.1 (180) 3.4 (87) 3.6 (92) 10.1 (257.5) 164 feet (50 m) 5.5 (140): 4 and 6 mm SST probes 17.1 (434): PTFE covered probe 0.86 (22): 4 mm SST probe...
Page 199
5.2 (133) ½ - 14 NPT Optional adapters: M20x1.5, eurofast and minifast HTHP/HP Plate Design (Option for Alloy versions) 15.6 (397.5) Dimensions are in inches (millimeters). Rosemount 5300 Series 7.4 (188.5) The Alloy probes are designed with a protective plate. A-15...
Page 200
Rosemount 5300 Series Figure A-13. Coaxial probe with threaded connection 10.1 (257.1) 20 feet (6 m) 10.1 (257.1) 1 in.: s52 1½ in.: s60 20 feet (6 m) A-16 NPT 1/1½/2 inch 7.1 (180) ½ - 14 NPT Optional 3.4 (87)
Page 201
10 feet (3 m) Ø 0.24 (6) Ø 0.31 (8) 1.0 (26) Flange 5.2 (133) Dimensions are in inches (millimeters). Rosemount 5300 Series NPT 1½ / 2 inch 5.2 (133) 7.4 (188.5) 1½ in.: s52 2 in.: s60 7.4 (188.5) A-17...
Page 202
Rosemount 5300 Series Figure A-15. Flexible Twin Lead G 1½ inch 7.1 (180) 3.4 (87) 3.6 (92) 10.2 (259.5) 1.1 (27) 164 feet (50 m) Ø 0.16 (4) Ø 0.16 (4) 3.5 (90) 1.4 (35) 7.1 (180) 3.4 (87) 3.6 (92) 10.2 (259.5)
Page 203
Figure A-16. Bracket mounting. 5.2 (133) Pipe mounting (vertical pipe) Wall mounting Pipe diameter max 2.5 in. (64 mm) Dimensions are in inches (millimeters). Rosemount 5300 Series Pipe mounting (horizontal pipe) 2.2 (57) 0.3 (7) 0.8 (20) Hole pattern wall mounting 2.8 (70)
Page 204
Rosemount 5300 Series Figure A-17. Remote housing. 5.2 (133) 6.9 (175) Standard Variant 12.4 (315) HTHP/HP/C Variant A-20 7.4 (188.5) 5.2 (133) Dimensions are in inches (millimeters). Reference Manual 00809-0100-4530, Rev BA July 2009 7 (180) (87) (92) 3, 6, 9 ft. (1, 2, or 3 m)
3A, 3B, 4A, 5A, 5B 4A and 5A 4A and 5A 3A, 3B, 4A Rosemount 5300 Series Probe Type 3A, 3B, 4A, 4U, 4V, 5A and 5B 3A, 3B, 4A, 5A, and 5B 3A, 3B, 4A, 5A, 5B (only SST)
Page 206
Rosemount 5300 Series Code Probe Type Rigid Twin Lead Flexible Twin Lead with weight Coaxial (for level measurement) Coaxial, perforated. For level and interface measurement, or easier cleaning. Rigid Single Lead (8 mm) (10) Rigid Single Lead (13 mm) Vapor Single Rigid Probe for 2 in. pipes Vapor Single Rigid Probe for 3-4 in.
Page 209
(25) Available for probe types 2A, 4A, 4B, and 5A, except for HTHP. (26) Not available for Cryogenic probe. Example Model String: 5301-H-A-1-S-1-V-1A-M-002-05-AA-I1-M1C1. E-002-05, means 2 ft and 5 in. probe length. M-002-05, means 2.05 m Rosemount 5300 Series parameter listed in Appendix B: Product Certifications). Outer Diameter 1.8 in.
Page 210
Rosemount 5300 Series Model Code 5303, Level for Solids Model Product Description 5303 Guided Wave Solids Level Transmitter Code Signal Output 4-20 mA with HART communication fieldbus OUNDATION Code Housing Material Polyurethane-covered Aluminum Stainless Steel, Grade CF8M (ASTM A743) Code Conduit / Cable Threads ½...
Page 211
1 in. NPT thread 2 in. NPT thread 1 ½ in. BSP (G 1 ½ in.) thread 1 in. BSP (G 1 in.) thread Rosemount 5300 Series Probe Type 5A, 5B, 6A, 6B, standard temperature and pressure All for standard temperature and pressure...
Page 212
Rosemount 5300 Series Code Hazardous Locations Certifications No Hazardous Locations Certifications ATEX Flameproof NEPSI Flameproof FM Explosion-proof CSA Explosion-proof IECEx Flameproof ATEX Intrinsic Safety ATEX FISCO Intrinsic Safety NEPSI Intrinsic Safety NEPSI FISCO Intrinsic Safety FM Intrinsic Safety and Non-Incendive...
Page 213
(11) The standard alarm setting is high. (12) Certificate includes all pressure retaining wetted parts. Example Model String: 5303-H-A-1-S-1-V-6A-M-025-50-AA-I1-M1C1. E-025-05, means 25 ft and 5 in. probe length. M-025-50, means 25.5 m. Rosemount 5300 Series parameter listed in Appendix B: Product Certifications). A-29...
Rosemount 5300 Series SPARE PARTS Spare parts list Transmitter head Model 5301/5302/5303 Model Product Description 5301 Guided Wave Radar Liquid Level or Interface Transmitter (interface available for fully submerged probe) 5302 Guided Wave Radar Liquid Level and Interface Transmitter 5303...
Page 215
HART Diagnostics Suite (1) Not available with Flame/Explosionproof approvals (E1, E3, E5, E6, E7, KA, KB, KC, and KD) (2) Requires Foundation Fieldbus signal output. (3) Probes are intrinsically safe. (4) The standard alarm setting is high. Rosemount 5300 Series A-31...
Page 216
Rosemount 5300 Series Spare parts list Probe Model 5301/5302/5303 Model Product Description 5309 Spare probe Code Signal Output Not Applicable Code Housing Material Not Applicable Code Conduit / Cable Threads Not Applicable Code Operating Temperature and Pressure - 15 psig (-1bar) to 580 psig (40 bar) @ 302 °F (150 °C) High Temp / High Pressure 2940 psi @ 752 °F and 5000 psi @ 100 °F (203 bar @ 400 °C and...
Page 217
DN50, PN250. HTHP / HP units DN80, PN16 DN80, PN40 DN80, PN64 . HTHP / HP units Rosemount 5300 Series Process Connection Probe Lengths Flange / 1.5 in. Thread Min: 2 ft 5 in. (0.76 m). Max: 13 ft (4 m) Flange / 1 in.
Page 218
Rosemount 5300 Series (14) EN (DIN) Flanges (continued) DN80, PN100. HTHP / HP units DN80, PN160. HTHP / HP units DN80, PN250. HTHP / HP units DN100, PN16 DN100, PN40 DN100, PN64. HTHP / HP units DN100, PN100. HTHP / HP units DN100, PN160.
Page 219
(21) Available for SST and Alloy C-276 probes, type 2A, 4A, 4B, and 5A. Same disc material as probe material. (22) Available for SST probes, type 2A, 4A, 4B, 5A and 6A except for HTHP. (23) Certificate includes all pressure retaining wetted metal parts. Rosemount 5300 Series (19) (20) Outer Diameter 1.8 in.
Page 220
Rosemount 5300 Series Spare parts list Other Spares and Accessories Rosemount 5300 Series Code Process Connection - Size/Type (consult factory for other process connections) Other spares 03300-7001-0002 Weight kit Flexible Twin lead 03300-7001-0003 Weight kit Flexible 4 mm Single lead...
Page 221
(2) To order a centering disc in a different material, consult the factory. (3) 1½ in. NPT threaded connection (RA) is required. Rosemount 5300 Series A-37...
Page 222
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 A-38...
Before connecting a HART-based communicator in an explosive atmosphere, make sure the instruments in the loop are installed in accordance with intrinsically safe or non-incendive field wiring practices. Do not remove the transmitter cover in explosive atmospheres when the circuit is alive. Rosemount 5300 Series ). Please...
Rosemount 5300 Series EU CONFORMITY Failure to follow safe installation and servicing guidelines could result in death or serious injury: Make sure the transmitter is installed by qualified personnel and in accordance with applicable code of practice. Use the equipment only as specified in this manual. Failure to do so may impair the protection provided by the equipment.
Figure B-1. Approval Label ATEX HART model The Rosemount 5300 Series Guided Wave Radar Level Transmitter that has the following label attached has been certified to comply with Directive 94/9/EC of the European Parliament and the Council as published in the Official Journal of the European Communities No.
Page 226
Rosemount 5300 Series Figure B-2. Approval Label ATEX Foundation Fieldbus model The following information is provided as part of the label of the transmitter: • Name and address of the manufacturer (Rosemount). • CE Conformity Marking • Complete model number •...
Page 227
Impact and friction hazards need then to be considered according to EN 50284 clause 4.3.1 when the transmitter exposed to the exterior atmosphere of the tank is made with light metal alloys and used in category II 1 G. Rosemount 5300 Series +60 °C) =380 mA, P...
Figure B-4. Approval Label ATEX HART model The Rosemount 5300 Series Guided Wave Radar Level Transmitter that has the following label attached has been certified to comply with Directive 94/9/EC of the European Parliament and the Council as published in the Official Journal of the European Communities No.
Page 229
Impact and friction hazards need then to be considered according to EN 50284 clause 4.3.1 when the transmitter exposed to the exterior atmosphere of the tank is made with light metal alloys and used in category II 1 G. Rosemount 5300 Series ° < +60...
Rosemount 5300 Series NATIONAL SUPERVISION AND INSPECTION CENTER FOR EXPLOSION PROTECTION AND SAFETY OF INSTRUMENTATION (NEPSI) APPROVALS Figure B-6. Approval Label National Supervision and Inspection Center for Explosion Protection and Safety of Instrumentation (NEPSI) HART Model E3 Flameproof: Ex iad IIC T4 (- 40 °C < T <...
Page 231
= 300 mA, P = 1.5 W, C Special Conditions for Safe Use (X): Refer to Certificates: GYJ081080X for Ex ia IIC T4 and GYJ081130X for Ex iad IIC T4. Rosemount 5300 Series E3/FIELDBUS I3/FIELDBUS < + 60 °C) < + 60 °C)
Page 232
Rosemount 5300 Series Figure B-8. Approval Label National Supervision and Inspection Center for Explosion Protection and Safety of Instrumentation (NEPSI) FISCO Model B-10 The following information is provided as part of the label of the transmitter: • Ex ia IIC T4 (- 50 °C < T •...
Figure B-9. Approval Labels Factory Mutual (FM) HART model The Rosemount 5300 Series Guided Wave Radar Level Transmitter that have the following labels attached have been certified to comply with the requirements of the approval agencies noted. Intrinsically Safe for Class I, II, III, Division 1, Groups A, B, C, D, E, F and G.
Page 234
Rosemount 5300 Series Figure B-10. Approval Labels Factory Mutual (FM) Foundation Fieldbus model Figure B-11. Approval Labels Factory Mutual (FM) FISCO model B-12 Intrinsically Safe for Class I, II, III, Division 1, Groups A, B, C, D, E, F and G.
Dust-Ignition proof for Class II/III, Division 1, Groups E, F and G with intrinsically safe connections to Class I, II, III, Div 1, Groups B, C, D, E, F and G. Temperature code T4. Ambient temperature limits: -50 Seal not required. Rosemount 5300 Series ° ° C to + 70 ° °...
Rosemount 5300 Series CANADIAN STANDARDS ASSOCIATION (CSA) APPROVAL Intrinsic Safety Figure B-14. Approval Label Canadian Standards Association (CSA) HART model B-14 This product meets the Dual Seal Requirements of ANSI/ISA 12.27.01-2003. Dual seal annunciation The breach of the secondary seal is annunciated via product leakage from the antenna’s vents.
Page 237
Temperature code T4. FISCO model: U =17.5 Vdc, I =0 nF, L =0 H. Control Drawing: 9240 030-937. Ambient temperature limits: -50 Rosemount 5300 Series =30 Vdc, I =300 mA, P =1.3 W, ° ° C to + 60 =380 mA, P =5.32 W,...
Rosemount 5300 Series Explosion Proof Figure B-17. Approval Label Canadian Standards Association (CSA) HART model Figure B-18. Approval Label Canadian Standards Association (CSA) Foundation Fieldbus model B-16 Cert. no. 1514653. E6 Explosion-proof with internal Intrinsically Safe Circuits [Exia]. Class I, Div. 1, Groups B, C and D.
60079-0 clause 8.1.2 when the transmitter exposed to the exterior atmosphere of the tank is made with light metal alloys and used in Zone 0. (1) Other temperature restrictions may apply, please refer to “Specifications” on page A-1. Rosemount 5300 Series =1 W, C =7.26 nF, L =0 mH.
Page 240
Rosemount 5300 Series Figure B-20. Approval Label IECEx Foundation Fieldbus model B-18 Intrinsically Safe Ex ia IIC T4 (T -50 °C to +60 °C). IECEx NEM 06.0001X. Fieldbus: U =30 V, I OUNDATION =0 mH. Installation Drawing: 9240 030-938.
Page 241
Impact and friction hazards need then to be considered according to IEC 60079-0 clause 8.1.2 when the transmitter exposed to the exterior atmosphere of the tank is made with light metal alloys and used in Zone 0. Rosemount 5300 Series =5.32 W, C =0 nF,...
Rosemount 5300 Series Flameproof Figure B-22. Approval Labels IECEx HART B-20 E7 Flameproof Ex iad IIC T4 (T :-40° C to +70 °C). IECEx NEM 06.0001X. =250 VAC Conditions of Certification The intrinsically safe circuits do not withstand the 500V AC test as specified in EN 50020 clause 6.4.12.
Page 243
Impact and friction hazards need then to be considered according to IEC 60079-0 clause 8.1.2 when the transmitter exposed to the exterior atmosphere of the tank is made with light metal alloys and used in Zone 0. Rosemount 5300 Series for Zone 1 and 4 cm B-21...
Page 245
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 Figure B-24. System Control Drawing for hazardous location installation of intrinsically safe FM approved apparatus. B-23...
Page 246
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 Figure B-25. Installation Drawing for hazardous location installation of intrinsically safe CSA approved apparatus. B-24...
Page 247
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 Figure B-26. Installation Drawing for hazardous location installation of intrinsically safe ATEX and IECEx approved apparatus. B-25...
Page 248
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 B-26...
Before connecting a HART-based communicator in an explosive atmosphere, make sure the instruments in the loop are installed in accordance with intrinsically safe or non-incendive field wiring practices. Do not remove the transmitter cover in explosive atmospheres when the circuit is alive. Rosemount 5300 Series ). Please...
Page 250
Rosemount 5300 Series Failure to follow safe installation and servicing guidelines could result in death or serious injury: Make sure the transmitter is installed by qualified personnel and in accordance with applicable code of practice. Use the equipment only as specified in this manual. Failure to do so may impair the protection provided by the equipment.
OUNDATION TRANSDUCER 1100>GEOM_OFFSET_DIST The Distance Offset is also available in RRM: a. Click the Tank icon under Device Config/Setup in the RRM workspace. b. In the Tank window, select the Geometry tab. c. Click the Advanced button. Rosemount 5300 Series...
Near Zone (referred to as the region between 0-3.3 ft. (0-1 m) below the Upper Reference Point) is normally somewhat limited. However, the Rosemount 5300 transmitter is equipped with a software functionality that minimizes the Upper Transition Zone. The factory setting is normally sufficient and doesn’t need to be repeated after installation.
Page 253
Nozzle diameter < 3 in. for 13 mm single rigid 1. Click Trim Near Zone. 2. Select Trim near zone and click OK. 3. Click OK in the appearing dialog. 4. Wait 1 minute. 5. Restart the transmitter. Rosemount 5300 Series...
Page 254
Rosemount 5300 Series NOTE! The Trim Near Zone function can be reset by selecting Reset near zone trimming to factory settings in the previous dialog. In 375 Field Communicator, the Trim Near Zone is reached with the [2, 1, 7, 2] HART sequence (if the command is recommended) or from the [2, 7, 1] HART sequence.
4. Type the desired value in the Upper Null Zone field. 5. Click the OK button. Now the Upper Null Zone is stored in the transmitter memory. Upper Reference Point Reference Gauge Height Upper Null Zone Disturbance Distance, m Rosemount 5300 Series Upper Null Zone Product Level...
Figure C-6. Threshold settings in RRM. Measurement with the Rosemount 5300 is based on the fact that the radar signal pulses are reflected by the product surface and the interface between two liquids. Various signal amplitude thresholds are used to separate the measurement signal from disturbing echoes and noise.
Page 257
By default, this value is 0 (function is not used). This is because highly contaminated or condensed tank-seals, and turbulent or boiling surfaces, which can cause splashing, may cause function to be triggered. Rosemount 5300 Series Interface Threshold Surface Threshold...
A product with very low reflectivity is characterized by its low dielectric constant. Such a product reflects relatively small amounts of the microwave energy emitted from the Rosemount 5300 transmitter, and most of it is transmitted through the product. Consequently, situations may occur where the surface pulse is invisible to the transmitter at long measuring ranges.
Page 259
DC is allowed to differ from the initial product DC value. If the estimation goes outside this limit, a warning will be generated. Used Product DC: This is the estimated product dielectric constant that the device will use for Probe End Projection. Rosemount 5300 Series C-11...
DC. This will force the device to use the configured initial product DC. Measurement with the Rosemount 5300 is based on the fact that the radar signal pulses are reflected by the product surface. Different parameters are used to track the measurement signal to achieve a reliable and stable measurement.
Page 261
The default value is 2 seconds. Note that increasing the damping value will also prolong the system response time. The damping value can also be changed from the Output, General tab. Rosemount 5300 Series C-13...
For applications with a varying pressure and/or temperature, certain models of the Rosemount 5300 Series have a built-in function that automatically compensates for varying vapor dielectric constants. See “Dynamic Vapor Compensation” on page C-15.
Temperature ( Dielectric constant Pressure The standard version of a Rosemount 5300 Series level transmitter can be configured for static compensation of vapor by manually entering the dielectric constant of vapor, see “Dielectric Constant Settings” on page C-14. For applications with a varying pressure and/or temperature, certain models...
Page 264
Vapor Compensation is supported. C-16 The Rosemount 5300 uses a reference reflector mounted on the probe at a certain distance to estimate the dielectric constant of the vapor. The transmitter knows where the reference reflector pulse should have been if there were no vapor present.
Page 265
> 6.6 ft (2 m) (see X in figure below), to dynamically compensate up to level 100%. The minimum measuring range for this functionality is 12 in. (300 mm). Rosemount 5300 Series VC marking on tank seal C-17...
Rosemount 5300 Series Figure C-15. Minimum requirements. Installation Setup C-18 Level: 100% Measuring range: min 12 in. / 300 mm Level: 0% If a 5300 Series transmitter is ordered from Rosemount together with a 9901 Chamber, these space requirements are met. If an existing chamber is used, which does not meet these space requirements, a spool piece can be added.
Page 267
0.75 ft. (0.23 m) distance and probes longer than 6.6 ft. (2 m) will have the reflector end at 1.6 ft. (0.5 m). 6. Cycle power to finish. In 375 Field Communicator, the calibration can be done by HART command [2, 7, 3]. Rosemount 5300 Series C-19...
Rosemount 5300 Series SIGNAL QUALITY METRICS Figure C-17. Echo curve showing surface peak amplitude, noise peak amplitude, and surface threshold. C-20 Signal Quality Metrics indicates the surface signal integrity compared to the noise. It can be used to schedule maintenance to clean the probe or detect and monitor turbulence, boiling, foam and emulsions.
If Signal Quality Metrics is not supported or disabled, the Signal Quality and Surface / Noise Margin will always be set to 0. To view Signal Quality Metrics in RRM, choose the Tools > Device Display option and select the Signal Quality Metrics tab. Rosemount 5300 Series Diagnostics Suite C-21...
Page 270
Rosemount 5300 Series Figure C-20. Configure Transmitter Variables for Signal Quality Metrics. Signal Quality Surface / Noise Margin C-22 Signal Quality Metrics can be shown on the LCD panel. See “Operation” on page 6-1. Signal Quality Metrics can be assigned to Transmitter Variables (SV, TV, or QV).
- M6 screw (3 pcs) - M6 nut (4 pcs) Remove the transmitter head from the probe Remove the transmitter head from the probe by unscrewing the M50 nut. M50 nut Mount the probe in the tank Rosemount 5300 Series...
Page 272
Rosemount 5300 Series U-bolt M6 screw M50 nut Mount the bracket to the pole Mount the bracket to the pole, making sure the distance between the probe and bracket does not exceed the length of the remote connection. 1. Put the two U-bolts through the holes of 2.
Page 273
July 2009 REMOTE CONNECTION, FIELD RETROFIT To upgrade a currently installed Rosemount 5300 transmitter with a remote housing, the embedded software needs revision 2.A2 or later, and the Spare Part Kit (03300-3001-000X) must be ordered. The Rosemount 5300 transmitter head must be configured for remote housing to work.
Rosemount 5300 Series REMOTE HOUSING CONFIGURATION When using remote housing, the remote connection length should be configured. If the Remote Housing is ordered with a transmitter, it is configured in the factory. In Rosemount Radar Master, select Setup > Tank.
375 HART Sequence: [2, 4, 1, 7]. Select 3 Other. Enter the analog output level representing the low alarm current. Press Enter and click Verify that the analog output current is correct. Click Abort to end loop test. Verify that the Current output is restored to the original mode. Rosemount 5300 Series...
Page 276
Rosemount 5300 Series Figure E-1. Range Values. Upper Range Value Lower Range Value 20mA 5. Perform a two-point calibration check of the transmitter by applying level to two points on the probe within the measuring range. Verify that the current output corresponds to the level input values using a known reference measurement.
RRM: Set Alarm Mode AOut 1 to Low Current. Click Store to save changes. Click Loop test... and enter the Current AOut 1 value representing the low alarm current. Click Start and verify that the output current is correct. Click Stop to end loop test. Rosemount 5300 Series...
Page 278
Rosemount 5300 Series 5. Restore the Alarm mode to the original mode used in the loop. Verify that the analog output current is correct. RRM: Set Alarm Mode AOut 1 to original mode. Click Store to save changes. Verify that the output current is correct.
AMS: Select Configure / Setup, Analog Output from the menu. Click Loop Test... Select Other and enter the mA value representing the high Analog Output Level and follow the instructions. Verify that the output current is correct. Rosemount 5300 Series...
Page 280
Rosemount 5300 Series 4. Using loop test, enter the mA value representing the low alarm mode. Verify that the analog output current is correct using the reference meter. This step tests for possible quiescent current related failures. AMS: Select Configure / Setup, Analog Output from the menu. Click Loop Test...
The transducer block contains the actual measurement data, including a level and distance reading. Channels 1–16 are assigned to these measurements (see Figure F-1). The transducer block includes information about sensor type, engineering units, and all parameters needed to configure the transmitter. Rosemount 5300 Series Channel Channel Channel Channel...
Parameter ST_REV TAG_DESC STRATEGY ALERT_KEY MODE_BLK BLOCK_ERR UPDATE_EVT Each input has an assigned channel which can be linked to the AI block. The channels for the Rosemount 5300 Series are the following: Channel Number CHANNEL_RADAR_LEVEL CHANNEL_RADAR_ULLAGE CHANNEL_RADAR_LEVELRATE CHANNEL_RADAR_SIGNAL_STRENGTH CHANNEL_RADAR_VOLUME CHANNEL_RADAR_INTERNAL_TEMPERATURE...
Page 283
(0°means that probe is mounted vertically). This parameter is used for User Defined probes only. Please contact Emerson Process Management Service Department for more information This parameter is used for User Defined probes only. Please...
Page 284
STATS_ATTEMPTS STATS_FAILURES Index Number This parameter is used for User Defined probes only. Please contact Emerson Process Management Service Department for more information This parameter is used for User Defined probes only. Please contact Emerson Process Management Service Department for more information This parameter is used for User Defined probes only.
Page 285
MEAS_WARNING CONFIG_WARNING VAPOR_COMP_STATUS Table F-3. Probe Type Rosemount 5300 Series Index Number The total number of timed out A/D board message attempts This is the maximum thickness of the upper product that the device is able to measure through. If the thickness of the...
Page 286
Rosemount 5300 Series Table F-4. Device Mode Table F-5. Environment Table F-6. Presentation Table F-7. LCD Parameters VALUE Normal operation Spare Restart device Set to factory default database Trim Near Zone Value of Bit Number ENV_ENVIRONMENT 0x00000004 0x00000008 0x10000000 Value of...
Page 287
Reference Manual 00809-0100-4530, Rev BA July 2009 Table F-8. Product Dielectrical Range Table F-9. Measurement Mode Rosemount 5300 Series Value of Bit Number LCD_PARAMETERS 0x00010000 0x00020000 0x00040000 VALUE 1.4 - 1.9 (e.g. liquified gas, plastics) 1.9-2.5 (e.g. Oil Based) 2.5-4 (e.g. Oil Based) 4-10 (e.g.
Rosemount 5300 Series SUPPORTED UNITS Unit Codes Table F-10. Length Table F-11. Level Rate Table F-12. Temperature Table F-13. Signal Strength Table F-14. Volume Table F-15. Time Table F-16. Percent Value Display 1010 1012 1013 1018 1019 Value Display 1061...
Gain control Unsupported Combination of Functions Probe Missing Interface Measurement Failure Hardware Write Protected Rosemount 5300 Series Corrective action Replace gauge Verify device installation and configuration Check ambient temperature. If ambient temperature ok replace the device. Check volume configuration Load default database to the device and...
Page 290
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 F-10...
Register Transducer Block The Register Transducer Block allows access to Database registers and Input registers of the Rosemount 5300 transmitter. This makes it possible to read a selected set of register directly by accessing the memory location. The Register Transducer Block is only available with advanced service.
Page 293
DB_REG_10_TYPE DB_REG_10_FLOAT DB_REG_10_INT_DEC RM_COMMAND RM_DATA RM_STATUS Rosemount 5300 Series Index Description Number If the register contains a float value, it should be displayed here If the register contains a DWORD value, and dec is chosen, it should be displayed here...
Page 294
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009...
Transducer Block The Advanced Configuration Transducer Block contains functions for the advanced configuration of the Rosemount 5300 transmitter. It includes functions, such as amplitude threshold settings for filtering of disturbing echoes and noise, simulation of measurement values, Empty Tank Handling for optimizing measurements close to the tank bottom, and strapping table for volume measurements.
Page 297
USE_STATIC_PRODUCT_DC CALCULATE_SIGNAL_QUALITY_M ETRICS USE_VAPOR_COMPENSATION SPEC_CONFIG_PEP SPEC_CONFIG_VC Rosemount 5300 Series Index Description Number Use this parameter to add a volume to each calculated volume value. The volume can for instance correspond to a sump volume that you wish to add to your calculation.
Page 298
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009...
The time the resource will wait for confirmation of receipt of a report before trying again. Retry will not happen when CONFIRM_TIME=0. Used to select the block execution method for this resource. The Rosemount 5300 supports the following: Scheduled: Blocks are only executed based on the function block schedule.
Page 300
Rosemount 5300 Series Index Parameter Number CYCLE_TYPE DD_RESOURCE DD_REV DEFINE_WRITE_LOCK DETAILED_STATUS DEV_REV DEV_STRING DEV_TYPE DIAG_OPTION DISTRIBUTOR DOWNLOAD_MODE FAULT_STATE FAILED_ACTIVE FAILED_ALM FAILED_ENABLE FAILED_MASK FAILED_PRI FB_OPTION FEATURES FEATURES_SEL FINAL_ASSY_NUM FREE_SPACE FREE_TIME GRANT_DENY HARD_TYPES HARDWARE_REV HEALTH_INDEX ITK_VER LIM_NOTIFY MAINT_ACTIVE Identifies the block execution methods available for this resource.
Page 301
4 Restart processor – does a warm start of CPU. State of the function block application state machine. Number of EEPROM blocks that have been modified since last burn. This value will count down to zero when the configuration is saved. Rosemount 5300 Series Description...
Page 302
Rosemount 5300 Series Index Parameter Number SAVE_CONFIG_NOW SECURITY_IO SELF_TEST SET_FSTATE SHED_RCAS SHED_ROUT SIMULATE_IO SIMULATE_STATE ST_REV START_WITH_DEFAULTS STRATEGY SUMMARY_STATUS TAG_DESC TEST_RW UPDATE_EVT WRITE_ALM WRITE_LOCK WRITE_PRI XD_OPTION Allows the user to optionally save all non-volatile information immediately. Status of security switch. Instructs resource block to perform self-test. Tests are device specific.
A maintenance alarm indicates that the device or some part of the device needs maintenance soon. If the condition is ignored, the device will eventually fail. There are five parameters associated with MAINT_ALARMS, they are described below. Rosemount 5300 Series...
Page 304
Rosemount 5300 Series MAINT_ENABLED The MAINT_ENABLED parameter contains a list of conditions indicating that the device or some part of the device needs maintenance soon. Below is a list of the conditions: 1. Configuration Error 2. Configuration Warning 3. Simulation Mode 4.
HI_HI_PRI • LO_PRI • LO_LO_PRI RECOMMENDED_ACTION The RECOMMENDED_ACTION parameter displays a text string that will give a recommended course of action to take based on which type and which specific event of the PlantWeb alerts are active. Rosemount 5300 Series...
Figure J-2 on page J-4 illustrates the internal components of the AI function block, and Table J-1 lists the AI block parameters and their units of measure, descriptions, and index numbers. Rosemount 5300 Series OUT_D...
Page 308
Rosemount 5300 Series Table J-1. Definitions of Analog Input Function Block System Parameters Index Parameter Number ACK_OPTION None ALARM_HYS Percent ALARM_SEL None ALARM_SUM None ALERT_KEY None BLOCK_ALM None BLOCK_ERR None CHANNEL None FIELD_VAL Percent GRANT_DENY None HI_ALM None HI_HI_ALM None...
Page 309
This measure is to prevent devices that went through simulation in the staging process from being installed with simulation enabled. With simulation enabled, the actual measurement value has no impact on the OUT value or the status. Rosemount 5300 Series Description...
Page 310
Rosemount 5300 Series Figure J-2. Analog Input Function Block Schematic Figure J-3. Analog Input Function Block Timing Diagram DAMPING Analog Measurement Access HI_HI_LIM Analog HI_LIM Meas. LO_LO_LIM LO_LIM CHANNEL ALARM_HYS LOW_CUT Cutoff Filter Convert SIMULATE L_TYPE FIELD_VAL IO_OPTS OUT_SCALE XD_SCALE NOTES: OUT = block output value and status.
NOTE! Low Cutoff is the only I/O option supported by the AI block. You can set the I/O option in Manual or Out of Service mode only. Rosemount 5300 Series – * XD_SCALE values ...
Rosemount 5300 Series BLOCK ERRORS Table J-2. BLOCK_ERR Conditions MODES Table J-2 lists conditions reported in the BLOCK_ERR parameter. Condition Condition Name and Description Number Other Block Configuration Error: the selected channel carries a measurement that is incompatible with the engineering units selected in XD_SCALE, the L_TYPE parameter is not configured, or CHANNEL = zero.
(such as diagnostics and system alerts). Alarm conditions of priority 3 to 7 are advisory alarms of increasing priority. 8-15 Alarm conditions of priority 8 to 15 are critical alarms of increasing priority. Rosemount 5300 Series Priority Description...
Rosemount 5300 Series Status Handling ADVANCED FEATURES Normally, the status of the PV reflects the status of the measurement value, the operating condition of the I/O card, and any active alarm condition. In Auto mode, OUT reflects the value and status quality of the PV. In Man mode, the OUT status constant limit is set to indicate that the value is a constant and the OUT status is Good.
CHANNEL Select the channel that corresponds to the desired sensor measurement. The Rosemount 5300 measures Level (channel 1), Distance (channel 2), Level Rate (channel 3), Signal Strength (channel 4), Volume (channel 5), Internal Temperature (channel 6), Upper Product Volume (channel 7), Lower Product...
Page 316
Rosemount 5300 Series Table J-4. Length Table J-5. Level Rate J-10 Indirect Square Root Select indirect square root when the desired output is an inferred measurement based on the transmitter measurement and the relationship between the sensor measurement and the inferred measurement is square root (e.g.
Page 317
Reference Manual 00809-0100-4530, Rev BA July 2009 Table J-6. Temperature Table J-7. Signal Strength Table J-8. Volume Rosemount 5300 Series Display °C °F Display Display Gallon ImpGall Description Degree Celsius Degree Fahrenheit Description millivolt Description Cubic meter Liter Cubic inch...
Page 318
Reference Manual 00809-0100-4530, Rev BA Rosemount 5300 Series July 2009 J-12...
Page 319
AI Block ....5-38 Block Errors ....7-45 www.rosemount.com Rosemount 5300 Series Index BLOCK_ERR AI Block .
Page 322
Reference Manual Rosemount 5300 Series 00809-0100-4530, Rev BA July 2009 Index-4...
Page 324
Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc. HART is a registered trademark of the HART Communication Foundation. PlantWeb is a registered trademark of one of the Emerson Process Management group of companies. AMS Suite is a trademark of Emerson Process Management.
Need help?
Do you have a question about the ROSEMOUNT 5300 and is the answer not in the manual?
Questions and answers
não estou conseguindo habilitar o simula de valores no rosemount 5300 series, mostra erro!
The context does not provide information on enabling value simulation for the Emerson Rosemount 5300 series. However, it mentions how to disable simulation modes. To turn off the Advanced Simulation Mode, set Holding Register 3600 to 0. If you need to enable value simulation, refer to the device manual or configuration settings.
This answer is automatically generated
Não estou conseguindo ativar o simulação no rosemount 5300