Instruction Manual pH 1100/ pH 2100 Bench pH/Ion Meter Technology Made Easy ... 68X090818 Rev. 3 07/03...
Page 3
If there are doubts in the use of the meter, please do not hesitate to contact the nearest Authorized Distributor or Customer Service Dept. for assistance. Eutech Instruments/ Oakton Instruments reserve the rights to change, make improvement and modify specifications without prior notice and cannot accept any responsibility for damage or malfunction to the instrument caused by improper use.
Instruction Manual pH 1100/ 2100 TABLE OF CONTENTS INTRODUCTION Introducing the Bench Meter Series Keypad Rear Panel Electrode Holder STARTING UP Back Panel Connections Powering up and powering down 2.2.1 Powering up 2.2.2 Powering down H CALIBRATION & MEASUREMENT pH calibration 3.1.1 Automatic Temperature Compensation (ATC) 3.1.2...
Page 5
Instruction Manual pH 1100/ 2100 10.2.9 pH calibration data reset setup program P1.7 10.3 mV setup mode 10.3.1 Entering mV setup mode 10.3.2 mV measurement alarm setup program P2.0 10.3.3 mV calibration due alarm setup program P2.1 10.3.4 Stability setup program P2.2 10.3.5 View mV calibration data setup program P2.3 10.3.6...
Instruction Manual pH 1100/ 2100 INTRODUCTION Introducing the Bench Meter Series The pH 1100/ 2100 bench meter is microprocessor-based which incorporates new ASIC (Application Specific Integrated Circuit). It is designed with convenience in mind and offers many advanced, user-friendly features. The meters are capable of storing and recalling up to 100 data sets in its non-volatile memory.
Instruction Manual pH 1100/ 2100 Keypad See Figure 1. A large membrane keypad with touch feedback makes the meter easy to use. Names and symbols describe the function button controls. Figure 1: Membrane Keypad Power meter On or Off ON/OFF Confirm selection in all modes of operations;...
Instruction Manual pH 1100/ 2100 Rear Panel See Figure 2. The pH 1100/ 2100 meters provide a complete set of input connections for the various accessories commonly used. Listed in the table below are details of the connections that you can make. Figure 2: Rear View of Meter Connector Function...
Instruction Manual pH 1100/ 2100 Electrode Holder The electrode holder is included in the same box as the meter. To attach the electrode holder to the meter: The electrode holder base attached to the bottom of the meter comes in the shipping position.
Instruction Manual pH 1100/ 2100 STARTING UP Attention: Do not get water on the BNC connector during operation. Avoid touching the connector with soiled or wet hands. Back Panel Connections Refer back to Figure 2. Make the necessary connections as mentioned in Section 1.3 Rear Panel, pg 8. Powering up and powering down 2.2.1 Powering up...
Instruction Manual pH 1100/ 2100 H CALIBRATION & MEASUREMENT pH calibration This meter is capable of up to 5 point calibration with selectable buffer sets (USA, NIST, Bf1, Bf2 or any 5 custom buffers of your choice). The meter retains stored pH calibrations even when turned off. For best accuracy, we recommend that you perform at least a 2-point calibration using buffers that bracket (one above and one below) the expected sample range.
Instruction Manual pH 1100/ 2100 3.1.2 Starting pH calibration Turn the meter ON. If the meter is not in the pH measurement mode, press EXIT and press >/CAL to select pH/Temp. The pH/Temp mode selected will flash. See Figure 13. Press ENTER to Select &...
Instruction Manual pH 1100/ 2100 3.1.3 Standard pH buffer calibration When entering the standard calibration mode, the first buffer calibration point for standard pH buffers is 7.000 or 6.860 ° 7.001pH 25.4 C (depending upon which pH buffer set is selected). The upper display shows the current reading, the lower display will show Auto Buf Scan: 7.000...
Instruction Manual pH 1100/ 2100 3.1.4 Custom pH buffer calibration When entering the custom calibration mode, the lower display ° ° 7.179pH 25.3 C 7.179pH 25.3 C will show Key In Std:. The upper display shows the pH buffer the electrode is dipped into.
Instruction Manual pH 1100/ 2100 3.1.5 Calibration error messages Standard pH Buffer Calibration: If the first buffer value of 7.000 or 6.860 is off by more than ±1.0, Error : the upper display will flash Error: and the lower display will read pH Offset Error pH Offset Error.
Instruction Manual pH 1100/ 2100 pH measurement This meter is capable of taking measurements with automatic or ° 7.179pH 25.3 C manual temperature compensation. Automatic Temperature Compensation (ATC) occurs when a temperature sensor is Jan 18 2002 08:46:25 plugged into the meter. If there is no temperature sensor plugged into the meter, the default setting is automatically 25°C Figure 29 (if the meter has never been manually set for temperature) or the...
Instruction Manual pH 1100/ 2100 3.2.3 Taking pH Measurements NOTE: Be sure to remove the electrode soaker bottle or ° 7.179pH 25.3 C protective rubber cap on the electrode before measurement. Jan 18 2002 08:46:25 Rinse the probe with deionized or distilled water before use Figure 35 to remove any impurities adhering to the probe body.
Instruction Manual pH 1100/ 2100 mV CALIBRATION AND MEASUREMENT mV Calibration Turn the meter ON. If the meter is not in the mV pH/Temp measurement mode, press EXIT and use >/CAL to select Select & Press ENTER mV. The mode selected will flash. See Figure 37. Press ENTER to confirm selection.
Instruction Manual pH 1100/ 2100 mV Measurement Turn the meter ON. If the meter is NOT in the mV pH/Temp measurement mode, press EXIT and use >/CAL to select mV. The mode selected will flash. See Figure 45. Press ENTER to Select &...
Instruction Manual pH 1100/ 2100 ION CALIBRATION AND MEASUREMENT (ONLY APPLICABLE FOR pH 2100) The pH 2100 meter reads ion concentration in ppm, ppt, mg/l, g/l, mmol/l and mol/l units. See Ion setup Program P3.0 (page 43) to select ion concentration units. Ion calibration This instrument is capable of up to 5-point custom ion calibration with a minimum of 2-point calibration.
Instruction Manual pH 1100/ 2100 If you are performing a 2-point calibration, press EXIT. The screen flashes Aver. Slope = (the average slope of the Aver. Slope = 49.2mV electrode) in the upper display and Sys Updating... in the lower Sys Updating.
Instruction Manual pH 1100/ 2100 Ion measurement The pH 2100 meter features two different modes of ion pH/Temp measurement: high resolution mode and low resolution mode. Select & Press ENTER These modes are user selectable in the Ion setup program P3.4. (page 47).
Instruction Manual pH 1100/ 2100 MEMORY FUNCTIONS Memory input This meter can store up to 100 sets of data in several combinations: pH 1100 meter: • pH, temperature, date and time • mV, temperature, date and time pH 2100 meter: •...
Instruction Manual pH 1100/ 2100 Memory recall This function recalls the previous readings stored in the memory. You can only access memory recall from the measurement mode. Memory recall is in “Last In First Out” (LIFO) order. There are two methods of viewing stored data in the memory recall function: manual and automatic. In manual memory recall, the user must manually scroll through the memory points using the ∆/MI and ∇/MR keys.
Instruction Manual pH 1100/ 2100 STABILITY INDICATOR The stability indicator signals when pH, mV or Ion readings are stable. ° The pH, mV or Ion annunciator will blink when reading are not stable. 7.546pH 25.4 C See Figure 68. Once the reading becomes stable the annunciator Jan 18 2002 08:46:30 becomes static and no longer blinks.
Instruction Manual pH 1100/ 2100 TEMPERATURE CALIBRATION Temperature Calibration For this calibration procedure, the ATC probe should be attached to the meter. ATC will appear in the upper display. See Figure 72. Dip the temperature probe into a solution of known temperature, °...
Instruction Manual pH 1100/ 2100 10 SETUP MODE The advanced setup mode lets you customize your meter’s preferences and defaults. The pH 2100 benchtop meter features 4 different setup modes (the pH 1100 meter features 3 modes): 1. pH/Temperature 2. mV 3.
Instruction Manual pH 1100/ 2100 10.1.2 mV setup submenus 10.1.3 Ion setup submenus (pH 2100 only) Press SETUP to enter setup mode from the Ion Press SETUP to enter setup mode. Use >/CAL to select mV, and then press ENTER. Use the ∆/MI and measurement mode.
Instruction Manual pH 1100/ 2100 10.1.4 Meter general configuration setup submenus Press SETUP to enter setup mode. Use >/CAL to select meter, and then press ENTER. Use the ∆/MI and ∇/MR keys to scroll through submenus. There are 8 submenus (programs) in the meter setup mode: Date/Time Setup:P4.0 Date/Time Setup:P4.0 Memory Clear Setup:P4.6...
Instruction Manual pH 1100/ 2100 10.2 pH/ temperature setup mode 10.2.1 Entering pH/temperature setup mode Setup Mode Activated Turn the meter ON. If the meter is not in the pH measurement Please wait. . . mode, press EXIT and use >/CAL to select pH/Temp. The mode selected will flash.
Instruction Manual pH 1100/ 2100 10.2.3 pH resolution setup program P1.1 Use this program to select pH resolution (0.01 or 0.001 pH). pH Resol. Setup: P1.1 Enter pH/Temperature setup mode as described in steps 1-4 in 10.2.1 Entering pH/temperature setup mode (page 30). Press ENT.
Instruction Manual pH 1100/ 2100 10.2.5 pH Measurement Alarm Setup Program P1.3 Use this program to select alarm for low and high pH pH Meas. Alarm :P1.3 measurements. Press ENT. to Select Enter pH/Temperature setup mode as described in steps 1-4 Figure 87 in 10.2.1 Entering pH/temperature setup mode (page 30).
Instruction Manual pH 1100/ 2100 10.2.6 pH Calibration Due Alarm Setup program P1.4 pHCalDue Alarm : P1.4 Use this program to set a reminder when the next calibration is due. Choose from 1 to 365 days from last calibration. Press ENT. to Select Enter pH/Temperature setup mode as described in steps 1-4 in Figure 91...
Instruction Manual pH 1100/ 2100 10.2.7 Stability Setup Program P1.5 Use this program to turn on or off the pH stability indicator and to set level of stability (low, medium or high). Stability definitions High stability: reading is stabilized quickly, but reading is less accurate. Medium stability: reading stability is averaged between high and low stability.
Instruction Manual pH 1100/ 2100 10.2.8 View pH calibration data setup program P1.6 Use this program to view the following: • Last pH calibration date/time • pH buffer set selected • pH calibration temperature and temperature compensation • pH offset value pH Cal Data :P1.6 •...
Instruction Manual pH 1100/ 2100 10.2.9 pH calibration data reset setup program P1.7 pH Cal Reset :P1.7 Press ENT. to Select Use this program to erase all pH calibration data. Figure 105 Enter pH/Temperature setup mode as described in steps 1-4 in Section 10.2.1 Entering pH/temperature setup mode (page 30).
Instruction Manual pH 1100/ 2100 10.3 mV setup mode 10.3.1 Entering mV setup mode Turn the meter ON. If the meter is not in the mV measurement Setup Mode Activated mode, press EXIT and use >/CAL to select mV. The mode Please wait.
Instruction Manual pH 1100/ 2100 10.3.2 mV measurement alarm setup program P2.0 Use this program to select alarm for low and high mV measurements. mV Meas. Alarm :P2.0 Enter mV setup mode as described in steps 1-4 in Section Press ENT. to Select 10.3.1 Entering mV setup mode (page 37).
Instruction Manual pH 1100/ 2100 10.3.3 mV calibration due alarm setup program P2.1 Use this program to set a reminder when the next calibration is mVCalDue Alarm :P2.1 due. Choose from 1 to 365 days from last calibration. Press ENT. to Select Enter mV setup mode as described in steps 1-4 in Section 10.3.1 Entering mV setup mode (page 37).
Instruction Manual pH 1100/ 2100 10.3.4 Stability setup program P2.2 Use this program to turn on or off the mV stability indicator and to set level of stability (low, medium or high). Stability Setup : P2.2 Press ENT. to Select Stability definitions Figure 119 High stability:...
Instruction Manual pH 1100/ 2100 10.3.5 View mV calibration data setup program P2.3 Use this program to view the following: • Last mV calibration date/time • mV offset value mV Cal Data :P2.3 Enter mV setup mode as described in steps 1-4 in Section Press ENT.
Instruction Manual pH 1100/ 2100 10.3.6 mV calibration data reset setup program P2.4 Use this program to erase all mV calibration data. mV Cal Reset :P2.4 Press ENT. to Select Enter mV setup mode as described in steps 1-4 in Section 10.3.1 Entering mV setup mode (page 37).
Instruction Manual pH 1100/ 2100 10.4 Ion setup mode (pH 2100 meter only) 10.4.1 Entering ion setup mode Setup Mode Activated Turn the meter ON. If the meter is not in ion measurement Please wait. . . mode, press EXIT and use >/CAL key to select ion. The mode selected will flash.
Instruction Manual pH 1100/ 2100 10.4.3 Ion measurement alarm setup program P3.1 Use this program to select alarm for low and high ion measurements. Ion Meas. Alarm :P3.1 Press ENT. to Select Enter ion setup mode as described in steps 1-4 in Section 10.4.1 Entering ion setup mode (page 43).
Instruction Manual pH 1100/ 2100 10.4.4 Ion calibration due alarm setup program P3.2 Use this program to set a reminder when the next calibration is due. Choose from 1 to 365 days from last calibration. Enter ion setup mode as described in steps 1-4 in Section 10.4.1 IonCalDue Alarm :P3.2 Entering ion setup mode (page 43).
Instruction Manual pH 1100/ 2100 10.4.5 Stability setup program P3.3 Use this program to turn on or off the ion stability indicator and to set level of stability (low, medium or high). Stability definitions Stability Setup : P3.3 High stability: Press ENT.
Instruction Manual pH 1100/ 2100 10.4.6 Ion mode setup program P3.4 Use this program to select high or low ion resolution mode. High Ion Mode Setup : P3.4 resolution has ±500 mV value, low has ±1850 mV value. Press ENT. to Select Enter ion setup mode as described in steps 1-4 in Section Figure 144 10.4.1 Entering ion setup mode (page 43).
Instruction Manual pH 1100/ 2100 10.4.7 View ion calibration data setup program P3.5 Ion Cal Data :P3.5 Use this program to view the following: Press ENT. to Select • Last ion calibration date/time • Calibration points with absolute mV values Figure 146 •...
Instruction Manual pH 1100/ 2100 10.4.8 Ion calibration data reset setup program P3.6 Use this program to erase all ion calibration data. Ion Cal Reset :P3.6 Enter ion setup mode as described in steps 1-4 in Section Press ENT. to Select 10.4.1 Entering ion setup mode (page 43).
Instruction Manual pH 1100/ 2100 10.5 Meter general configuration setup mode 10.5.1 Entering meter general setup mode Turn the meter ON. The meter setup mode can be entered Setup Mode Activated from any measurement mode—pH, mV or ion (pH 2100 meter Please wait.
Instruction Manual pH 1100/ 2100 10.5.3 Backlight setup mode P4.1 Use this program to turn backlight on or off and set duration time (0.1 to 10 minutes). BackLight Setup:P4.1 Enter meter setup mode as described in steps 1-4 in Section Press ENT.
Instruction Manual pH 1100/ 2100 10.5.4 Data log setup mode P4.2 Use this program select automatic or manual data logging. Data Log Setup :P4.2 Automatic data logging can be set at timed intervals from 10 seconds to 23 hours, 59 minutes, 59 seconds. Press ENT.
Instruction Manual pH 1100/ 2100 10.5.6 Communication setup mode P4.4 Use this program to set parameters of serial communication. Tx/Rx Parameter:P4.4 Press ENT. to Select Enter meter setup mode as described in steps 1-4 in Section 10.5.1 Entering meter setup mode (page 50). Figure 175 Use ∆/MI key to scroll to program Tx/Rx Parameter: P4.4.
Instruction Manual pH 1100/ 2100 10.5.8 Memory clear setup mode P4.6 Mem Clear Setup :P4.6 Clears all pH, mV and Ion (pH2100 only) memory points stored Press ENT. to Select in memory. Figure 181 Enter meter setup mode as described in steps 1-4 in Section 10.5.1 Entering meter setup mode (page 50).
For best results, always keep the bulb wet, preferably in pH storage solution. Other pH buffers or tap water are also acceptable. Do not store in de-ionized water. Eutech Instruments/ Oakton Instruments offers a complete range of electrode storage and cleaning solutions.
Instruction Manual pH 1100/ 2100 12 RS 232 COMMUNICATION This meter will send data to a printer and computer through the RS-232 port located on the back panel of the meter. See Figure 188. This meter uses a 9 pin serial connection for RS-232 communication. 12.1 Using with printer If you are going to use your meter to send data to a printer, you must first configure your meter to match the settings for your printer (consult printer manual for proper printer operation).
Instruction Manual pH 1100/ 2100 13 ADDITIONAL INFORMATION 13.1 pH and Temperature The pH electrode is affected by temperature changes. Automatic Temperature Compensation (ATC) compensates for these temperature changes. Some solutions show an increase while others a decrease in pH with the same temperature change.
Instruction Manual pH 1100/ 2100 15 TROUBLESHOOTING & ERROR MESSAGES TROUBLESHOOTING Problem Possible Cause Solution Power ON but no display AC adapter not connected properly Connect AC adapter properly Unstable Reading Electrode submersion in sample is Place electrode deeper into sample insufficient Broken electrode Replace electrode...
1100/ 2100 17 ACCESSORIES Consult your Authorized Distributors for these items and other range of specialized pH electrodes or Ion Selective Electrodes. Item Description Eutech Instruments Oakton Instruments Order Code Order Code Glass-body combination pH electrode (refillable) with 1m cable length...
1100/ 2100 18 WARRANTY Eutech Instruments/ Oakton Instruments supplies this bench meter with a 3-year warranty and 6-month warranty for against manufacturing defects from the date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warranty period, please return, freight prepaid, and correction will be made without charge.
Page 64
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our website listed below: Oakton Instruments Eutech Instruments Pte Ltd Distributed by: P.O Box 5136, Blk 55, Ayer Rajah Crescent, Vernon Hills, IL60061, USA #04-16/24 Singapore 139949...