User's and Service Guide
Agilent Technologies 85050D
7 mm Economy Calibration Kit
This manual applies directly to 85050D calibration kits with serial number prefix
3101A.The calibration devices in this kit are individually serialized. Record the device
serial numbers in the table provided in this manual (see "Recording the Device Serial
Numbers" in Chapter 1.)
Agilent Part Number: 85050-90052
Printed in USA
Print Date: August 2008
Supersedes: December 2006
© Copyright 1994, 1996, 2002, 2006, 2008 Agilent Technologies, Inc. All rights reserved.
www.valuetronics.com

Advertisement

loading
Need help?

Need help?

Do you have a question about the 85050D and is the answer not in the manual?

Questions and answers

Summary of Contents for Agilent Technologies 85050D

  • Page 1 Agilent Technologies 85050D 7 mm Economy Calibration Kit This manual applies directly to 85050D calibration kits with serial number prefix 3101A.The calibration devices in this kit are individually serialized. Record the device serial numbers in the table provided in this manual (see “Recording the Device Serial Numbers”...
  • Page 2 FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
  • Page 3 • Go to www.agilent.com. • Enter the document’s part number (located on the title page) in the Search box. • Click Search. • Click on the hyperlink for the document. • Click the printer icon located in the tool bar. 85050D www.valuetronics.com...
  • Page 4 85050D www.valuetronics.com...
  • Page 5 85050D Kits with Serial Prefix 3027A........
  • Page 6 Introduction ..............6-2 A Standard Definitions Class Assignments and Standard Definitions Values are Available on the Web ..A-2 Contents-vi 85050D www.valuetronics.com...
  • Page 7 General Information 85050D 1- 1 www.valuetronics.com...
  • Page 8 Calibration Kit Overview Calibration Kit Overview The Agilent 85050D 7 mm calibration kit is used to calibrate Agilent network analyzers up to 18 GHz for measurements of components with 7 mm connectors. The standards in this calibration kit allow you to perform simple 1- or 2-port and TRM (thru–reflect–match) calibrations.
  • Page 9 Equipment Required but Not Supplied Connector cleaning supplies and various electrostatic discharge (ESD) protection devices are not supplied with the calibration kit but are required to ensure successful operation of the kit. Refer to Table 6-1 on page 6-2 for ordering information. 85050D www.valuetronics.com...
  • Page 10 _______________________________ Calibration Kits Documented in This Manual This manual applies to any 85050D calibration kit whose serial number is listed on the title page. If your calibration kit has a different serial number prefix, refer to the “Calibration Kit History”...
  • Page 11 This section describes calibration kits with serial number prefixes lower that the ones listed on the title page. 85050D Kits with Serial Prefix 3027A These calibration kits did not have the calibration definitions disk to support the Agilent 8510C network analyzer. The part numbers provided in this manual are the recommended replacement parts for these kits.
  • Page 12 Periodically remeasure the device and note any changes in its corrected response which can be attributed to the test system. With experience you will be able to see changes in the reference responses that indicate a need to perform the measurement calibration again. 1- 6 85050D www.valuetronics.com...
  • Page 13 Specifications 85050D 2- 1 www.valuetronics.com...
  • Page 14 During a measurement calibration, the temperature of the calibration devices must be stable and within the range shown in Table 2-1. IMPORTANT Avoid unnecessary handling of the devices during calibration because your fingers are a heat source. 85050D www.valuetronics.com...
  • Page 15 Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Agilent Technologies verifies the mechanical characteristics of the devices in the kit with special gaging processes and electrical testing.
  • Page 16 Specifications Mechanical Characteristics Figure 2-1 Connector Pin Depth 85050D www.valuetronics.com...
  • Page 17 When measuring pin depth, the measured value (resultant average of three or more measurements) is not the true value. Always compare the measured value with the observed pin depth limits in Table 2-2 to evaluate the condition of device connectors. 85050D www.valuetronics.com...
  • Page 18 Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities...
  • Page 19 Use, Maintenance, and Care of the Devices 3- 1 85050D www.valuetronics.com...
  • Page 20 Remove the short from the cable. Figure 3-1 shows a typical ESD protection setup using a grounded mat and wrist strap. Refer to Chapter 6 , “Replaceable Parts,” for information on ordering supplies for ESD protection. Figure 3-1 ESD Protection Setup 3- 2 85050D www.valuetronics.com...
  • Page 21 If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration. 3- 3 85050D www.valuetronics.com...
  • Page 22 Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Table 6-1, “Replaceable Parts for the 85050D Calibration Kit,” on page 6-2 for a part number for cleaning swabs.
  • Page 23 Always completely dry a connector before you reassemble or use it. 4. Inspect Inspect the connector to make sure that no particles or residue remain. Refer to “Visual Inspection” on page 3-3. 3- 5 85050D www.valuetronics.com...
  • Page 24 Use, Maintenance, and Care of the Devices Gaging Connectors Gaging Connectors The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only. See Table 6-1 on page 6-2 for part number information. They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.
  • Page 25 Refer to “Pin Depth” on page 2-3 for definitions of protrusion and recession. Figure 3-2 Reading the Connector Gage 3- 7 85050D www.valuetronics.com...
  • Page 26 Compare the average reading with the pin depth specifications listed in Table 2-2 on page 2-5. Remove the device from the gage. 3- 8 85050D www.valuetronics.com...
  • Page 27 Use, Maintenance, and Care of the Devices Gaging Connectors Figure 3-3 Gaging 7 mm Connectors 3- 9 85050D www.valuetronics.com...
  • Page 28 Refer to Table 6-1 on page 6-2 for replacement part number and ordering information. Table 3-1 Torque Wrench Information Connector Type Torque Setting Torque Tolerance 7 mm 136 N-cm (12 in-lb) ± 13.6 N-cm (± 1.2 in-lb) 3- 10 85050D www.valuetronics.com...
  • Page 29 This is especially true when several devices are connected together. Figure 3-4 Wrench Positions 2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-5. Figure 3-5 Using the Torque Wrench 3- 11 85050D www.valuetronics.com...
  • Page 30 • Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. 3- 12 85050D www.valuetronics.com...
  • Page 31 Performance Verification 85050D 4- 1 www.valuetronics.com...
  • Page 32 These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Agilent Technologies. 85050D www.valuetronics.com...
  • Page 33 A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Agilent Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Agilent Technologies. Refer to “Contacting Agilent” on page 5-5.
  • Page 34 Performance Verification Recertification 85050D www.valuetronics.com...
  • Page 35 Troubleshooting 85050D 5- 1 www.valuetronics.com...
  • Page 36 Troubleshooting Troubleshooting Process Troubleshooting Process If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1. 85050D www.valuetronics.com...
  • Page 37 Troubleshooting Troubleshooting Process Figure 5-1 Troubleshooting Flowchart 85050D www.valuetronics.com...
  • Page 38 If you need additional information, see “Contacting Agilent” on page 5. Returning a Kit or Device to Agilent If your kit or device requires service, contact Agilent Technologies for information on where to send it. See “Contacting Agilent” on page 5 for contact information. Include a...
  • Page 39 If you do not have access to the Internet, please contact your Agilent field engineer. NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period. 85050D www.valuetronics.com...
  • Page 40 Troubleshooting Contacting Agilent 85050D www.valuetronics.com...
  • Page 41 Replaceable Parts 85050D 6- 1 www.valuetronics.com...
  • Page 42 • items not included in the calibration kit that are either required or recommended for successful operation of the kit To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Agilent Technologies. See “Contacting Agilent” on page 5-5.
  • Page 43 Replaceable Parts Introduction Table 6-1 Replaceable Parts for the 85050D Calibration Kit Description Qty Per Kit Agilent Part Number ESD heel strap 9300-1308 a. See “Printing Copies of Documentation from the Web” on page -iii. b. See the 8510C On-Site Service Manual for instructions on using the disk.
  • Page 44 Replaceable Parts Introduction 85050D www.valuetronics.com...
  • Page 45 Standard Definitions 85050D www.valuetronics.com...
  • Page 46 Agilent vector network analyzers. It also provides some examples of how to set up a new calibration kit and how to modify an existing calibration kit definition file. To download a free copy, go to www.agilent.com and enter literature number 5989-4840EN in the Search window. 85050D www.valuetronics.com...
  • Page 47 See temperature effects of temperature calibration definitions when to perform electrical specifications contacting Agilent Technologies calibration definitions electrostatic discharge, See ESD disk, part number environmental contents entering regulations calibration kit calibration kit requirements...
  • Page 48 3-10 observed limits sliding loads manual protrusion temperature part number recession torque wrench 3-10 printing copies -iii typical values traceability mating plane surfaces preventive maintenance standard definitions cleaning procedures downloading from Agilent Web connector cleaning site Index-2 85050D www.valuetronics.com...
  • Page 49 3-10 traceability of device specifications troubleshooting, flowchart user’s and service guide part number verification performance temperature visual inspection warranty, documentation wear connector 85050D Index-3 www.valuetronics.com...
  • Page 50 Index Index-4 85050D www.valuetronics.com...