8 VERIFICATION OF THE PULSE DATA
The calibration and verifi cation of the pulse data requires a specialist test and
measurement laboratory for which the IEC standard sets out certain minimal
requirements.
Teseq uses the following instruments for calibration purposes:
Oscilloscope with an analogue bandwidth of min. 1 GHz
Coaxial measurement adapter MD 101 Pellegrini-target as per
IEC / EN 61000-4–2 (2001) or MD 103 according to latest draft
20 dB attenuator covering the range from dc to 12.4 GHz
SUCOFLEX-HF-coaxial cable
High voltage DC voltmeter (Ri >30 GΩ)
The instruments are periodically recalibrated in accordance with the require-
ments of ISO 17025.
Calibration installation with oscilloscope in a Faraday cage.
31