  • Page 1 USB-1000 Series Multifunctional Data Acquisition Devices User 's Manual Rev: B Smacq Technologies. Co., Ltd Smacq.com Smacq.cn...
  • Page 2: Statement

    Smacq Technologies. Co., Ltd. shall not be liable for any errors or inaccurate descriptions that may be ⚫...
  • Page 3: Safety Requirements

    Safety requirements Warning: Only connect voltage within the specified range. If the voltage exceeds the specified range, it may cause equipment damage and even affect personal safety. The voltage range that can be connected to each port is detailed in the product specification section.
  • Page 4: Measurement Category

    Measurement category Warning: This device can only be used in measurement category I (CAT I). Do not use this device to connect signals or perform measurements in measurement categories II/III/IV. Measurement category description Measurement Category I (CAT I) refers to measurements taken on circuits that are not directly connected to the main power supply.
  • Page 5: Environment

    Environment Temperature Operation 0℃~55℃ Storage -40℃~85℃ Humidity Operation 5% RH~95% RH, non-condensing Storage 5% RH~95% RH, non-condensing Pollution level Highest altitude 2000m Pollution level description Pollution level 1: No pollution, or only dry non-conductive pollution occurs. This pollution level has no impact.
  • Page 6: Table Of Contents

    CONTENT Statement ............................1 Copyright ............................1 Trademark information ........................1 Other statements ..........................1 Contact us ............................1 Safety requirements........................... 2 Measurement category ........................3 Environment ............................4 1. Getting Started ..........................7 1.1. Product introduction ........................7 1.2. Function Diagram ........................8 1.3.
  • Page 7 5. Digital I/O ..........................23 5.1. Digital signal connect ........................ 23 6. Counter ............................24 6.1. Overview ..........................24 6.2. Event counter ........................... 24 6.3. Period/Positive/Negative Pulse Width Measurement ..............24 6.4. Connecting Counter Signals ....................... 25 7. Programming Instructions ......................26 7.1.
  • Page 8: Getting Started

    1. Getting Started This chapter describes the basic functions of USB-1000 Series Data Acquisition Device, as well as product specifications and precautions in the process of product unpacking. 1.1. Product introduction USB-1000 Series data acquisition device is the multifunctional data acquisition device based on high-speed USB2.0 interface.
  • Page 9: Function Diagram

    1.2. Function Diagram USB-1000 series data acquisition device functions schematic 1.3. Product specifications The following product specification parameters, unless otherwise stated, are acquired at the temperature of 25° C and the humidity of 40%, while the device is turned on for 20 minutes.
  • Page 10 Digital I/O Channel USB-1252A: 16-DI, 16-DO Ground reference DGND High level: 3.3~5V Digital input voltage Low level: 0~0.5V High level: 3.0-3.4V Digital output voltage Low level: 0~0.1V Digital output power-on status Low level Channel synchronization Counter Channel USB-1252A: 4-CT Resolution 32-bit Counter Measurement Edge count...
  • Page 11: Product Unpacking And Packing List

    If you find that the product comes in damaged after unpacking, please get in touch with us as soon as possible. Do not install damaged equipment on your devices. 2.3. Packing list Name Specification Description Quantity USB-1000 Series Multifunctional Data USB-1000 Series Acquisition Devices Include Attachments USB Cable USB cable/black/1.5 meters Wiring Terminals 10Pin/Green/3.81mm/pitch terminal block...
  • Page 12: Installation

    3. Installation This chapter describes signal connection and drive installation of USB-1000 series. 3.1. Connector signal pins distribution USB-1000 series signal pins distribution...
  • Page 13 1252A Signal pin allocation list Signal name NOTE(Single end) NOTE(Difference) Analog input 0 Analog input 0 + Analog input 1 Analog input 0 - Analog input 2 Analog input 1 + Analog input 3 Analog input 1 - Analog input 4 Analog input 2 + Analog input 5 Analog input 2 -...
  • Page 14: Usb Cable Reinforcement Design

    3.2. USB cable reinforcement design USB cable connectors are prone to be pulled off during operation. USB-1000 series data acquisition devices provide a cable reinforcement design, with which a strap can be used to fix the USB cable to the device to prevent the accidents.
  • Page 15: Drive Installation

    3.3. Drive installation Smacq USB-1000 series data acquisition device support Microsoft Windows XP, Windows 7, Windows 8/8.1, and Windows 10, including all the 32-bit and 64-bit versions. To install the driver for USB-1000devices, you need to turn off driver signature enforcement first. Here is an example step-by-step tutorial on how to install the driver in Windows 7.
  • Page 16: Analog Input (Ai)

    4. Analog Input (AI) This chapter describes measuring the relevant content of analog input signals on USB-1000 series data acquisition cards. AI here is short for Analog Input. 4.1. Circuit diagram Analog input circuit 4.2. Input Range Description ⚫ Single-Ended Mode For single-ended input mode, positive input is connected to Aix, and negative input to AGND or AI-sense.
  • Page 17: Description On Multi-Channel Scanning

    ⚫ Trigger Sources When USB-1000 series DAQ acquires signals via analog input, its trigger source can be set to trigger by software or trigger by the rising edge or falling edge of digital IO input channel DIN0...
  • Page 18: Signal Connection Mode

    4.4. Signal Connection Mode USB-1000 series data acquisition devices support analog input acquisition connection methods of non-grounding reference single-ended input (NRSE), differential input (DIFF),referenced Single- ended(RSE). The recommended connection methods for floating-ground and grounding signal sources list Input mode Floating grounded signal source...
  • Page 19: Floating Grounded Signal Source

    4.5. Floating grounded signal source A floating grounded signal source is not connected in any way to the building ground system but, it has an isolated ground-reference point. Common floating grounded signal sources are transformers, thermocouples, battery equipment, optical isolate, and isolation amplifier output. An instrument or device with isolated output is a floating-ground signal source.
  • Page 20 Differential input mode with single bias resistor For a floating signal source with a large internal resistance, you can use the differential input mode with two bias resistors. The fully balanced bias resistor connection in this way can provide a slightly better noise suppression, but it can reduce the load on the signal source and result in gain errors.
  • Page 21: Non-Grounded Reference Single-Ended Mode (Nrse) Connection

    ⚫ Non-grounded reference single-ended mode (NRSE) connection When both the following conditions are true, you can connect the float signal using a non-grounded reference single-end mode: ◼ The input signal voltage is higher than 1V ◼ The length of the cable connecting signals to the acquisition device is lower than 3 meters If the signal does not meet the above conditions, it is recommended to use a differential mode connection to ensure better signal integrity.
  • Page 22: Grounded Signal Source

    4.6. Grounded Signal Source The grounded signal source is a signal source connected to the building ground. If the computer is connected to the same power supply as the signal source, the source is already connected to a common ground point relative to the device. Instruments and equipment connected to the building power supply system while with non-isolated outputs belong to this type of signal source.
  • Page 23: Use A Non-Grounded Reference Single-Ended Mode (Nrse)

    ⚫ Use a non-grounded reference single-ended mode (NRSE) When all of the following conditions are true, you can connect the float signal using a non-grounded reference single-ended mode: The input signal voltage is higher than 1V ◼ The cable connecting the signal to the acquisition device is less than 3 meters ◼...
  • Page 24: Digital I/O

    5. Digital I/O This chapter introduces the digital I/O signal acquisition on USB-1000 series data acquisition cards. 16-channel digital input and 16-channel digital output, DGND is digital I/O reference grounded signal. The digital input DIN channel is compatible with 0~5V level signal. 2~5V is determined as high level, and 0~0.5V as low level.
  • Page 25: Counter

    6. Counter 6.1. Overview USB-1000 series DAQ device has 4 counter channels, with input signal voltage up to 5V. GND is the ground-reference signal of counter channels. The counter can be configured as any of the following three functions: Event counter ◼...
  • Page 26: Connecting Counter Signals

    6.4. Connecting Counter Signals The connection diagram for counter channel signal. Counter Signal Connection...
  • Page 27: Programming Instructions

    FindUSBDAQ() int_stdcall FindUSBDAQ() Find the USB-1000 series DAQ device connected to the computer. parameters Note Return value the number of USB-1000 series DAQ devices connected to the computer. OpenDevice() int_stdcall OpenDevice(int DevIndex) Turn on the specified device. parameters Note DevIndex the index number of the DAQ device, 0 is the number of the first device.
  • Page 28: Analog Input Related Functions

    Set the sampling rate of AI channel of the DAQ device. When the USB-1000 series DAQ device works in multiple AI channels, the sampling rate per channel = the set sampling rate/the number of channels used. For example, given the set sampling rate is 200kS/s, and 4 channels are used, then the sampling rate of each channel is 50kS/s.
  • Page 29: Digital Io Related Functions

    SetSoftTrig() int_stdcall SetSoftTrig(int DevIndex, unsigned char Trig) Set software trigger. parameters Note DevIndex The index number of the DAQ device, 0 is the number of the first device. Trig The software trigger switch, 0 indicates Off, 1 indicates On. Error Codes Return value 0 indicates error free.
  • Page 30: Read Data Control Functions

    ClearCounter() int_stdcall ClearCounter(int DevIndex, unsigned char CtrNum) Return the counter to zero. parameters Note DevIndex The index number of the DAQ device, 0 is the number of the first device. CtrNum The index number of counter channels. 0~3 correspond to Ct0~Ct3. If 0x0f is set, it indicates 4 counters operating simultaneously.34 Error Codes Return value...
  • Page 31 GetDioIn() unsigned int_stdcall GetDioIn(int DevIndex) Read the value of digital IO channel DIN. parameters Note DevIndex The index number of the DAQ device, 0 is the number of the first device. Return value It returns the value of digital IO channel DIN, the least 16 bits correspond to the values of DIN0~DIN15.
  • Page 32: Error Code

    If a negative is returned in function operation, it means that an operation error occurred. Refer to Table 4 below for detailed error codes. Table4. Error Codes Error Code Description No USB-1000 series DAQ device connected to the computer is found. Index of DAQ device out of bound. DAQ device firmware error. DAQ device closed.
  • Page 33: After Sales Service And Warranty

    8. After sales service and warranty Smacq Technologies. Co., Ltd. promises that its products are under warranty. If the product malfunctions during normal use, we will provide free repair or replacement of parts for the user. For detailed warranty instructions, please refer to the warranty instructions inside the packaging box.
  • Page 34: Ordering Information

    9. Ordering Information Main Equipment Model Notes USB-1252A 16-AI(Single-500kSa/s, Multipath-200kSa/s), 4-CT, 16-DI, 16-DO Standard accessories Model Notes USB connection cable, 1.5 meters, USB-A type to USB-B type USB Cable TB10-3.81 10-bit, 3.81mm pitch terminal block Optional accessories Model Notes SDIN 35mm DIN rail mounting bracket CHF-100B Current sensor, 100A, DC~20kHz, output ±...
  • Page 35: Document Revision History

    10. Document Revision History Date Edition Remarks First release 2016.08.30 Rev: A 2017.09.25 Rev: B Update USB-1252A model information...