VISSCO JOHARI WinStim Plus-WS2UWB Instruction Manual

VISSCO JOHARI WinStim Plus-WS2UWB Instruction Manual

Combo therapy electro+ultrasound

Advertisement

Quick Links

WS2UWB
COMBO THERAPY
ELECTRO + ULTRASOUND

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the WinStim Plus-WS2UWB and is the answer not in the manual?

Questions and answers

Summary of Contents for VISSCO JOHARI WinStim Plus-WS2UWB

  • Page 1 WS2UWB COMBO THERAPY ELECTRO + ULTRASOUND...
  • Page 2 WinStim Plus-WS2UWB NO SITTING - Text with a “NO SITTING” indicator will explain possible Safety infractions that will potentially cause injury and equipment damage. NO STEPPING ON SURFACE - Text with a “NO STEPPING ON SURFACE” indicator will explain possible...
  • Page 3: Table Of Contents

    1.6 Adverse reactions INTRODUCTION DESCRIPTION ACCESSORIES SPECIFICATIONS CONTROLS AND FUNCTIONS LCD DESCRIPTION HOW TO OPERATE WINSTIM PLUS-WS2UWB 8.1 Preparing electrodes 8.2 Prerequisites 8.3 Preparing the skin 8.4 Starting the device 8.5 Selecting the waveform 8.6 Starting and stopping the treatment 8.7 Waveform details...
  • Page 4: Instruction For The User Of Winstim Plus Ws2Uwb

    The WinStim Plus-WS2UWB system is not designated to prevent the ingress of water liquids. Ingress of water or liquids could cause malfunction of internal components of the system and therefore create a...
  • Page 5 WinStim Plus-WS2UWB If problem persist, contact doctor. RMS 2 mA/cm Do not modify this equipment without authorization of the manufacturer Ÿ...
  • Page 6 WinStim Plus-WS2UWB WinStim Plus-WS2UWB is indicated to be used as under Russian, High Volt and EMS for: Relaxation of muscle spasms Prevention or retardation of disuse atrophy Increase local blood circulation Muscle re-education Maintaining or increasing range of motion Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis...
  • Page 7 WinStim Plus-WS2UWB WinStim Plus-WS2UWB should not be used in the following areas: This device must not be used by pregnant woman, patients with metallic prosthesis / implant-able devices, patients with cardiac problems.
  • Page 8 WinStim Plus-WS2UWB The WinStim Plus-WS2UWB may be susceptible to interference originating from short wave diathermy units operating in close proximity to it (aprrox. 3m). Avoid operating the WinStim Plus-WS2UWB adjacent to and simultaneously with operating short-wave devices. The WinStim Plus-WS2UWB should be kept out of the reach of children.
  • Page 9 WinStim Plus-WS2UWB Precaution while using Pen electrode Ensure the intensity is adjusted slowly and smoothly and not increased beyond the patient's tolerance. During treatment, do not lift electrodes off the skin without the intensity being turned down to zero first.
  • Page 10: Introduction

    It is clinical model with easy user interface and versatility to treat different body areas simultaneously. This aesthetically designed clinical model has 7 selectable modes and treatment parameters. The state of art WinStim Plus-WS2UWB is light weight, small in size, and battery powered which allows it to be easily moved any location for immediate use.
  • Page 11: Accessories

    WinStim Plus-WS2UWB 4. ACCESSORIES WinStim Plus-WS2UWB comes complete with all the necessary accessories and below is a list of items that are included: Length : 2.5 meter Electrode Sponge Pouch (Size : 8.5x8.5cm) Pen Electrode OUTPUT: 24V DC (3.75A) Instruction Manual / CD...
  • Page 12: Specifications

    WinStim Plus-WS2UWB 5. SPECIFICATIONS Length WS2UWB unit Width WS2UWB unit Height WS2UWB unit Standard Weight WS2UWB unit Power (WS2UWB unit) +24Volts DC, 3.75A Fuse Rating 4A/32V, Fast BLOW...
  • Page 13 WinStim Plus-WS2UWB INTERFERENTIAL...
  • Page 14 WinStim Plus-WS2UWB 0-37.5 0-37.5 0-37.5 Microcurrent waveform Output Mode : Electrodes or Probe Output Intensity : 0 – 1000.0μA Polarity Positive : Negative or Alternating Available on channels : 1, 2, 3, 4 Mode : CC / uC (Constant Current / Micro Current)
  • Page 15: Controls And Functions

    WinStim Plus-WS2UWB 6. CONTROLS AND FUNCTIONS WINSTIM-WS2UWB is an innovative touch screen desktop device providing Electronic Stimulation through four channels. (For Service Purpose Only...
  • Page 16: Lcd Description

    WinStim Plus-WS2UWB 7. LCD DESCRIPTION 1. Channel 1 - 2 : Provides Electrical Stimulation in Channel 1 & 2 4. Selectable Programs Icons.
  • Page 17 WinStim Plus-WS2UWB Library Button : for treatment details...
  • Page 18 WinStim Plus-WS2UWB Safety Indication : - 93.75 mA Warning 800 us HIGH ELECTRICAL CHARGE, DEVICE SHOULD BE USED UNDER THE CONTINUOUS SUPERVISION OF A PHYSICIAN OR LICENSED PRACTITIONER Cancel If the phase duration is 800usec. and intensity 93.75 a popup message will appear 93.75 mA...
  • Page 19 WinStim Plus-WS2UWB 8. HOW TO OPERATE WELLSTIM Instruction For Use Carbon Electrode Un-pack Carbon Electrode, Lead Wires & electrode pads. Soak the electrode pads in water & then Connect the lead wires in the electrodes. rinse out the excess water.
  • Page 20 WinStim Plus-WS2UWB The desired mode on the WinStim Plus-WS2UWB should be selected and set. Keep the WinStim Plus-WS2UWB on a stable surface.
  • Page 21 WinStim Plus-WS2UWB NOTE: Do not connect WinStim Plus-WS2UWB to a power supply rated differently than that described aboce. Avoid operating the WinStim Plus-WS2UWB adjacent to and simultaneously with operating short-wave device, as it may be susceptible to interference originating from short-wave diathermy units.
  • Page 22 WinStim Plus-WS2UWB...
  • Page 23 WinStim Plus-WS2UWB...
  • Page 24 WinStim Plus-WS2UWB 8.7.1.1 Interferentiral Waveform Channel 1-2 Eight Carbon Electrodes carbon 8.7.1.2 IFC Premodulated 2P Four Carbon Electrodes...
  • Page 25 WinStim Plus-WS2UWB carbon 8.7.2 Russian Waveform Four Carbon Electrodes carbon...
  • Page 26 WinStim Plus-WS2UWB 8.7.3 TENS Waveform Four Carbon Electrodes...
  • Page 27 WinStim Plus-WS2UWB carbon...
  • Page 28 WinStim Plus-WS2UWB carbon carbon...
  • Page 29 WinStim Plus-WS2UWB carbon SAFETY INDICATION : a) In CC mode, a pop-up message will appear at phase duration 800usec & intensity 93.75mA. If you press ok then you have exceeded safety limit and in this case a sign will appear on running channel b) In CV mode same thing will happen at phase duration 500usec &...
  • Page 30 WinStim Plus-WS2UWB carbon 8.7.5 Short Pulse Waveform Eight Carbon Electrodes...
  • Page 31 WinStim Plus-WS2UWB carbon...
  • Page 32 WinStim Plus-WS2UWB carbon...
  • Page 33 WinStim Plus-WS2UWB SAFETY INDICATION : a) In CC mode, a pop-up message will appear at phase duration 800usec & intensity 93.75mA. If you press ok then you have exceeded safety limit and in this case a sign will appear on running channel b) In CV mode same thing will happen at phase duration 500usec &...
  • Page 34 WinStim Plus-WS2UWB 8.7.7 Monophasic Waveform Four Carbon Electrodes carbon SAFETY INDICATION : A pop-up message will appear at phase duration 3000usec & intensity 25mA. If you press ok then you have exceeded safety limit and in this case a sign will appear on running channel...
  • Page 35 WinStim Plus-WS2UWB carbon SAFETY INDICATION : A pop-up message will appear at phase duration 3000usec & intensity 50mA. If you press ok then you have exceeded safety limit and in this case a sign will appear on running channel carbon NOTE: If the polarity reversal is on then a sharp sensation will be felt at every 7.5 minutes due to...
  • Page 36 WinStim Plus-WS2UWB 8.7.8 Galvanic Waveform Four Carbon Electrodes carbon...
  • Page 37 WinStim Plus-WS2UWB 8.7.9 Surged Waveform Four Carbon Electrodes carbon...
  • Page 38 WinStim Plus-WS2UWB SAFETY INDICATION : A pop-up message will appear at phase duration 3000usec & intensity 25mA. If you press ok then you have exceeded safety limit and in this case a sign will appear on running channel carbon SAFETY INDICATION : A pop-up message will appear at phase duration 3000usec &...
  • Page 39 WinStim Plus-WS2UWB Micro-current Detail Screen a) Press Microcurrent 2P sub-waveform Note: The Channel selection screen will appear. 2) Select desired channel/Group Note : The Microcurrent 2P Detail screen will appear 3) Now you can select the desired parameters such as Time, Pulse polarity, Frequency etc.
  • Page 40 WinStim Plus-WS2UWB b) Press Microcurrent 4P sub waveform Note: The Channel selection screen will appear 2) Select desired Channel/ group will show the following Note: The Microcurrent 4P Detail screen will appear 3) Now you can select the desired parameters such as Time, Pulse polarity, Frequency etc.
  • Page 41 WinStim Plus-WS2UWB 8.7.11 Triangular Waveform Four Carbon Electrodes carbon...
  • Page 42: D Curves

    WinStim Plus-WS2UWB 9. S/D CURVES Four Carbon Electrodes Back Store Library Start carbon...
  • Page 43: Ultrasound

    WinStim Plus-WS2UWB 10. ULTRASOUND Start Head Start Library Store Warming...
  • Page 44 WinStim Plus-WS2UWB...
  • Page 45: Combo

    WinStim Plus-WS2UWB 11. COMBO WinStim Plus-WS2UWB’s Combo mode allows the user to select & use Ultrasound therapy in combination with electrical muscle stimulation. In this mode of therapy, the head of the Ultrasound Applicator becomes one half of the electrical circuit. An electrode attached to the Black Electrode cable tagged COMBO (See Fig.
  • Page 46 WinStim Plus-WS2UWB Operational Mode Constant Current Duty Cycle Start Head Start Library Store Warming Paste the carbon electrode on the selected treatment are as per the practitioner’s advice. Operational Mode Constant Current Duty Cycle Start Head Start Library Store Warming...
  • Page 47 WinStim Plus-WS2UWB 1 Red Electrode cable tagged COMBO...
  • Page 48: Settings

    WinStim Plus-WS2UWB 12. SETTINGS Options available are shown below : Default values for Estim options are set to OFF as shown below :...
  • Page 49: Back Cover Assembly

    WinStim Plus-WS2UWB 13. BACK COVER ASSEMBLY Back cover is mounted to cover the output treatment panel. Back panel is to be push in the corresponding holes in top enclose part cover. Carefully place back cover on top part in a way that it is positioned correctly. As shown in below Picture Correct position of Back compartment cover is shown following picture.
  • Page 50: Troubleshooting

    WinStim Plus-WS2UWB 14. TROUBLESHOOTING...
  • Page 51: Maintenance

    WinStim Plus-WS2UWB 15 MAINTENANCE...
  • Page 52: Suggested Electrode Placement Chart

    WinStim Plus-WS2UWB 16. SUGGESTED ELECTRODE PLACEMENT CHART Caution : The device should only be operated under supervision of a registered medical practitioner. Electrode placements shown in this library/encyclopedia are only for reference purposes. Actual electrode placement may be vary according to pain or patient’s condition...
  • Page 53: Warranty

    WinStim Plus-WS2UWB Instruction Manual 17 WARRANTY This product warranty extends to the original consumer /purchaser of the product. Warranty duration This product is warranted to the original consumer for a period of one (1) year from the original purchase date.
  • Page 54 WinStim Plus-WS2UWB Instruction Manual The distributor cannot assume responsibility for loss or damage during shipment. For your protection, carefully pack the product for shipment and insure it with the carrier. Ensure that you return the unit and accessories related to your problem and also that you indicate full return address.
  • Page 55 Marketed By Vissco Healthcare Pvt Ltd. 1208, Dalamal Tower, Free Press Journal Road, Nariman Point, Mumbai- 400021, India Tel.: +91-22-43330300/ 43330333 Email: customercare@vissco.com Web.: www.vissco.com Manufactured By Johari Digital Healthcare Ltd. G-582-584, EPIP, Boranada-Salawas Road Jodhpur - 342012, Rajasthan, India TOLL FREE: 1800-102-8684 Email : info@joharidigital.com Web.: www.joharidigital.com...