1.3 Contraindications CAUTION The i700 system is not intended to be used to create images of the The CAUTION symbol indicates safety information that, if ignored, internal structure of teeth or the supporting skeletal structure.
The user should check the accuracy of the Read the user guide results and assess each individual case. ▪ T he i700 system must be used in accordance with its accompanying The official mark of the Europe Certificate User Guide. Authorized representative in the European ▪...
Keep Dry Reusable Tip This way up Seven-layer stacking prohibited Small Tip Consult electronic user guide Stand by (IEC 60417-5009) Calibration Tool 1.6 i700 Components Overview Practice Model Item Appearance i700 Handpiece Wrist Strap i700 Handpiece Cover Desktop Cradle User Guide...
A discolored practice model may affect the results of the practice mode. Power Cord ▪ T he strap is designed specifically for the weight of the i700 and should not be used with other products. ▪ M edit Scan for Clinics is included in the USB drive.
③ C onnect the power cord to ④ C onnect the power cord to a the medical adapter. power source. You can connect i700 directly to a PC with the power delivery cable. 1.7.2 Additional Settings of i700 (With Power Hub and Medical Adapter) ⑤ C onnect the USB 3.0 cable’s A...
Page 9
LED on the upper part of the i700 handpiece turns blue. Wall Mount Holder Turning Off the i700 Press and hold the power button at the bottom of the i700 handpiece for 3 seconds. Intraoral Scanner i700...
USB 3.0 cable provided by Medit. Medit is not responsible for any problems caused by cables other than the USB 3.0 cable macOS Monterey 12 provided by Medit. Be sure to use only the USB 3.0 cable included in the package. User Guide...
③ Select the installation path. ④ R ead the “License Agreement” carefully before checking “I agree to T he installation will not be processed while the i700 system is the License terms and conditions.” and then click “Install.” connected to a PC. Please be sure to disconnect the i700 USB 3.0 cable from the PC before the installation.
Medit Scan for Clinics User Guide W e recommend performing a calibration periodically. You can Please refer to the User Guide of Medit Scan for Clinics: Medit Scan for set the calibration period via Menu > Settings > Calibration Clinics > Menu > User Guide.
U se a sterilized and non-abrasive cloth to dry the mirror and the ▪ S canner tips can be re-sterilized up to 150 times and thereafter tip gently. must be disposed of as described in the disposal section. Intraoral Scanner i700...
Make sure to use alcohol that is free of impurities or ④ P ut the cover on the front of the i700 handpiece after cleaning it may stain the mirror. You can use either ethanol or propanol the filter.
R oHS, Restriction of the Use of Certain Hazardous Substances in CAUTION Electrical and Electronic Equipment. (2011/65/EU) ▪ D o not clean the i700 handpiece when the device is turned on W EEE, Waste Electrical and Electronic Equipment Directive. as the fluid may enter the scanner and cause malfunction. (2012/19/EU) ▪...
The i700 system is a high-precision optical medical device. Acquaint ▪ D o not block the air vents located at the rear of the i700 system. If all the following safety and operating instructions before the the equipment overheats, the i700 system may malfunction or stop installation, usage, and operation of i700.
Y ou should be familiar with the safe use of the i700 system, as ▪ I f the i700 system is dropped on the floor or if the unit is impacted, detailed in this User Guide. it must be calibrated before use. If the instrument is unable to ▪...
▪ H andling and replacing the tip. ▪ D o not connect the i700 system to a regular power strip or ▪ U sing the i700 system on patients. extension cord, as these connections are not as safe as grounded ▪...
T he i700 system projects a bright light from its tip during concentrations. scanning. ▪ T here is a risk of explosion if you use the i700 system near ▪ T he bright light projected from the tip of the i700 is not harmful flammable anesthetics.
The i700 system is intended for use in the electromagnetic ▪ Guidance 1 environment as specified below. The customer or the user of the i700 T he i700 system is intended for use in the electromagnetic system should ensure that it is used in such an environment.
Page 21
Dips, Short environment. If the modulation modulation Resistance to 0.5/1 cycles 0.5/1 cycles Magnetic Interruptions, user of the i700 magnetic fields was 70% Uт (30% dip 70% Uт (30% dip Fields in the and Voltage system requires 65 A/m 65 A/m tested and applied in Uт) for...
Page 22
▪ Guidance 3 power rating of the transmitter in watts (W) according to the T he i700 system is intended for use in the electromagnetic transmitter manufacturer, environment specified below. The customer or the user of the User Guide...
Page 23
RFID 2450; Portable RF communications equipment should be used no 217 Hz LTE Band 7 closer than 30 cm (12 inches) to any part of the i700 system. 5100 – Pulse Otherwise, degradation of the performance of this equipment WLAN...
WARNING Rating , 3 A ▪ U se of the i700 adjacent to or on other equipment must be Classifications for Protection Against Class I, Type BF Applied Parts avoided as it may result in improper operation. If this use is...
Page 25
9F, 10F, 13F, 14F, 16F, 8, Yangpyeong-ro 25-gil, Yeongdeungpo-gu, Transport 20 – 80% relative humidity Seoul, 07207, Republic of Korea Humidity Condition (non-condensing) Tel: +82-02-2193-9600 Air Pressure 620 – 1,200 hPa Contact for Product Support Emission limits per environment Email: support@medit.com Environment Hospital environment Tel: +82-02-2193-9600 Intraoral Scanner i700...