Page 1
⚑ Note: These instructions are provided as a reference tool for pump users and caregivers who are already familiar with the use of an insulin pump and with insulin therapy in general. Not all screens are shown. For more detailed information on the operation of the Tandem Mobi system, please refer to its user guide. ☚...
Page 2
☚ ☚ If the blood glucose (BG) is If the BG is less than the Target greater than the Target BG, BG, but greater than 70 mg/dL, the option to add a correction the option to reduce the bolus will appear. Tap Yes bolus amount will appear.
Page 3
☚ ☚ Tap Next to continue. To deliver an Extended Bolus, flip to the back to continue. ⚑ Note: If a calculated bolus is larger than the Otherwise, verify the dose and Max Bolus setting, the bolus will be reduced. You tap Confirm to continue.
Page 4
The Extended Bolus feature allows part of the bolus to be delivered now ☚ and part of it to be delivered slowly over a period of up to eight hours. This Tap the toggle next to can be helpful for Extend Food Bolus.
Page 5
☚ ☚ Tap 2 hr to select the length Tap Next to continue. of time that the bolus is to be delivered and then tap Done. ⚑ Note: When Control-IQ technology is enabled, the default and maximum limit is two hours. ☚...
Need help?
Do you have a question about the Mobi Bolus and is the answer not in the manual?
Questions and answers