Basic Build-Up Of The Pm2528 - Philips PM2528 Operating Manual

Automatic rms multimeter
Hide thumbs Also See for PM2528:
Table of Contents

Advertisement

38
4.
PRINCIPLE OF OPERATION
The PM2528 is subdivided into four functional parts: Analog Section, Analog Digital Converter, Control
Section and Display Keyboard Section. The basic build-up of the PM2528 is shown in fig. 13.
INTERFACE
'
(optioNaty
|
va
OH
ANALOG
ANALOG
CONTROL
DISPLAY/
'
i
SECTION
DIGITAL
SECTION
KEYBOARD
!
7
©
1
CONVERTER
SECTION
DC ATTENUATOR
ADC ANALOG
CONTROL SECTION
DISPLAY SECTION
OC AMPLIFIER
ADC CONTROL
'BP RO35
DISPLAY
i
AC ATTENUATOR
COUNTER
4k ROM
DISPLAY CONTROL,
AC AMPLIFIER
'ANALOG CONTROL
POWER SUPPLY
YBOARO SECTION
NICHES
OPTIONAL :
KEYBOARD CONTROL
j
PEAK DETECTOR
PARALLEL OUTPUT,
ton
HF caRD
(c0).
ANALOG OUTPUT
sra7ig
Fig. 13. Basic build-up of the PM2528.
4.1,
ANALOG SECTION
In the analog section ail quantities to be measured are converted into voltages. These voltages are supplied to
the Analog Digital Converter (ADC) via the ADC input control.