HP 84811A Operating And Service Manual page 14

Peak power sensor
Table of Contents

Advertisement

HP 84811A
e.
Remove and discard the Caution label on
the connector barrel. (A new one is supplied.)
f. Place the bulkhead flange in a vise.
g.
Using soft-jaw pliers, such as Utica 529-
10C, loosen the RF connector assembly by
grasping the connector barrel.
ከ. Remove the diode module, and associated
contacts.
L Apply new contacts, supplied presoldered
in the replacement kit, to the new diode. Insert
the new diode and contacts into the bulkhead.
Page 13
j. Reassemble by reversing steps a through
d, f and g.
CAUTION
The four allen cap screws must not be
torqued to greater than 13 in.-ozs. The
connector barrel should be torqued to 30
in.-lbs. maximum.
k.
Apply the new Caution label, in the same
location as the label removed in step e.
l
Apply the new Correction Factor label,
over the existing label.
Check for proper operation of the Peak Power
Sensor either by use or by performing the Per-
formance Tests in the Operating and Service
Manual.