HP 84811A
e.
Remove and discard the Caution label on
the connector barrel. (A new one is supplied.)
f. Place the bulkhead flange in a vise.
g.
Using soft-jaw pliers, such as Utica 529-
10C, loosen the RF connector assembly by
grasping the connector barrel.
ከ. Remove the diode module, and associated
contacts.
L Apply new contacts, supplied presoldered
in the replacement kit, to the new diode. Insert
the new diode and contacts into the bulkhead.
Page 13
j. Reassemble by reversing steps a through
d, f and g.
CAUTION
The four allen cap screws must not be
torqued to greater than 13 in.-ozs. The
connector barrel should be torqued to 30
in.-lbs. maximum.
k.
Apply the new Caution label, in the same
location as the label removed in step e.
l
Apply the new Correction Factor label,
over the existing label.
Check for proper operation of the Peak Power
Sensor either by use or by performing the Per-
formance Tests in the Operating and Service
Manual.