Table of Contents INTRODUCTION ........................ 2 DOCTOR MODE ......................... 3 ‘D ’ ........................ 3 CCESSING OCTOR ‘D ’? ............... 4 HAT CAN BE ADJUSTED WITHIN COTOR PASSWORD MODE ......................9 ‘P ’ ......................9 CCESSING ASSWORD ‘P ’? ............... 10 HAT CAN BE ADJUSTED WITHIN ASSWORD DISTRIBUTOR MODE ......................
Introduction This manual is only to be used by certified DANA Insulin Pump trainers and healthcare professionals who manage patients on the DANA RS Insulin Pump and are required to alter settings within the insulin pump configuration. The information within this manual is supplementary to the IFU that is distributed with the DANA Insulin Pump.
Doctor Mode Accessing ‘Doctor Mode’ Within MAIN MENU, press and hold three buttons at the same time. The pump will request a PASSWORD, – enter 30XY and press (OK). Notice The XY of 30XY is derived from the manufacturing day of the pumps.
What can be adjusted within ‘Dcotor Mode’? PRESET BOLUS Pre-set bolus feature is for a patient who has fixed-carbohydrate meals for breakfast, lunch and dinner. Set and save 3 different insulin dosages for meal bolus of breakfast, lunch and dinner. On –...
Page 6
MELODY Melody feature is a reminder to check BG level post-prandial. Set the melody from 0-150 minutes in 10 minute intervals. Set the time to 0 (zero) to turn this reminder off. BOLUS BLOCK Bolus Block is a safety feature. Example if BOLUS BLOCK is set to 10 (minutes) it is impossible to enter a subsequent bolus for 10 minutes.
Page 7
DEC. RATIO (%) DEC ratio means decrement ratio of insulin. Often referred to as an hourly absorption ratio for insulin. For the DANA insulin pump to calculate for residual active insulin, the DEC RATIO to suit the individual patient needs to be set.
Page 8
BOLUS MAX This is a safety feature to set maximum limits of delivery for individual patients. BOLUS MAXIMUM can be set from 0 u to a maximum single delivery of 80 units. DAILY MAX This is a safety feature to set maximum limits of delivery for individual patients.
Page 9
BLOCK SENSITIVE Is a variable sensitivity for any occlusion within the pump or infusion sets. Occlusions occur due to crystallization, blood or infusion set kink age. The Block sensitivity can be adjusted between H(high) 2-3u, M(medium) 3-5u, or L(low) 5-8u of blocked delivery and is usually adjusted for total daily dose.
Password Mode Accessing ‘Password Mode’ Within MAIN MENU, press and hold the three buttons at the same time. The pump will request a PASSWORD, – enter 3571 as the following sequence (1→ 3→5→7) and press (OK). Notice Must input the last number (1) first, if not - cannot access ‘Distributor Mode’.
) at the same time. LOCK : 1234 This is the personalized user password set within the pump. The password is derived from the manufacturing date of the pumps. Refer to the User Manual of the DANA Diabecare RS pump.
Distributor Mode Accessing ‘Distributor Mode’ Within MAIN MENU, press and hold the three buttons at the same time. The pump will request a PASSWORD, – enter 8949 as the following sequence (4→9→8→9) and press (OK). Notice Must input the third number (4) first, or cannot access the ‘Distributor Mode’.
What can be adjusted within ‘Distributor Mode’? DELETE HISTORY This is a function to delete all history within the pump. USER PW This is a function to check the personalized user password set within the pump. In case of forgetting the password, the distributor can review and/or reset the password here.
Some of these advanced algorithms deliver many boluses (referred to as SMB or Small Micro Bolus). The Audio tune that the DANA RS pump announces at the finish of a bolus / could be considered annoying. Within the ENGINEERING MODE of the DANA RS pump –...