Specifications Introduction; Timing Parameters; Common Specifications - HP 8112A Operating, Programming And Servicing Manual

50 mhz programmable pulse generator
Table of Contents

Advertisement

Specifications
Introduction

Timing Parameters

Common Specifications

All specifications in the following sections describe the warranted
performance of the instrument:
Timing parameters
n
Output parameters
n
Waveform characteristics
n
All specifications apply with a 50 52 load, after a 30 minute warm-up
period, and are valid for ambient temperature in the range 15OC to
35°C. Refer to the General Characteristics section of this chapter for
the performance derating factor to be used outside this temperature
range (within the specified operating range of 0°C to 55OC) .
All operating characteristics given in the following sections describe
typical performance figures which are non-warranted:
Trigger modes
n
Control modes
n
Output Modes
n
Inputs and Outputs
n
Additional features
n
General characteristics
n
Unless otherwise stated, specifications are quoted for 50% amplitude
in normal mode, fastest transitions.
Resolution
3 digits, best case 100 ps
Accuracy
Repeatability
Factor 4 better than accuracy
Jitter
max 0.2% of programmed value t100 ps
Specifications 2-1