Sencore ReZolver LC103 Operation And Application Manual page 25

Capacitor & inductor analyzer
Hide thumbs Also See for ReZolver LC103:
Table of Contents

Advertisement

Aluminum Electrolytics
Capacit~
1.5V
3.0V
6.0V
10V
15V
20V
25V
35V
50V
100V
200V
300V
400V
500V
600V 1000V
<1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
3.5
4.9
6.1
7.0
7.8
8.6
11
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
3.5
4.9
6.1
7.0
7.8
8.6
11
1.5
1.0
1.0
1.0
1.0
1.0
1.0
1.4
1.0
2.0
4.3
6.1
7.4
8.6
9.6
11
14
2.2
1.0
1.0
1.0
1.0
1.0
1.0
1.8
1.2
2.0
5.2
7.3
9.0
10
12
13
16
3.3
1.0
1.0
1.0
1.0
1.0
1.5
2.2
2.0
3.0
6.4
9.0
11
13
14
16
20
4.7
1.0
1.0
1.0
1.0
1.5
2.0
2.6
2.5
3.5
7.6
11
13
15
17
19
24
6.8
1.0
1.0
1.0
1.5
2.0
2.5
3.0
3.0
6.5
9.1
13
16
18
20
22
29
10
1.6
1.6
1.6
2.0
2.5
3.0
4.0
5.0
7.8
11
16
19
22
25
27
35
15
2.2
2.2
2.2
2.5
3.0
4.0
7.0
5.0
9.6
14
19
23
27
30
33
43
22
2.8
2.8
2.8
3.0
5.0
9.5
10
10
12
16
23
28
33
37
40
52
33
3.4
3.4
3.4
5.0
7.5
15
10
11
14
20
28
35
40
45
49
64
47
4.0
4.0
4.0
10
10
15
15
16
17
24
34
42
48
54
59
76
68
5.0
5.0
5.0
15
15
20
15
17
20
29
41
50
58
65
71
91
100
10
10
10
15
20
20
17
21
25
35
49
61
70
78
86
111
150
15
15
15
20
20
19
21
25
30
43
61
74
86
96
105
136
220
20
20
20
20
20
23
26
31
37
52
73
90
104
116
127
164
330
20
20
20
20
25
28
32
38
45
64
90
110
127
142
156
201
470
24
24
24
24
29
34
38
45
54
76
107
131
152
170
186
240
680
29
29
29
29
35
41
46
54
65
91
129
158
183
204
224
289
1000
35
35
35
35
43
49
55
65
78
111
157
192
221
247
271
350
1500
43
43
43
43
53
61
68
80
96
136
192
235
271
303
332
429
2200
52
52
52
52
64
73
82
97
116
164
232
284
328
367
402
519
3300
64
64
64
64
78
90
101
119
142
201
284
348
402
450
492
636
4700
76
76
76
76
93
107
120
142
170
240
339
416
480
537
588
759
6800
91
91
91
91
112
129
144
171
204
289
408
500
577
645
707
913
10000
111
111
111
111
136
157
175
207
247
350
495
606
700
783
857
1107
15000
136
136
136
136
166
192
214
254
303
429
606
742
857
959
1050
1356
22000
164
164
164
164
201
232
260
307
367
519
734
899
1038
1161
1272
1642
33000
201
201
201
201
246
284
318
376
450
636
899
1101
1272
1422
1557
2011
47000
240
240
240
240
294
339
379
449
537
759
1073
1314
1518
1697
1859
2399
68000
289
289
289
289
353
408
456
540
645
913
1291
1581
1825
2041
2236
2886
100000
350
350
350
350
429
495
553
655
783
1107
1565
1917
2214
2475
2711
3500
150000
429
429
429
429
535
606
678
802
959
1356
1917
2348
2711
3031
3320
4287
200000
495
495
495
495
606
734
783
971
1100
1570
2210
2710
3130
3500
3830
5191
Table
4
- Maximum
allowable
leakage for aluminum electrolytics per EIA
standards.
Leakage
charts
The fo llowing
leakage
charts
list
the
maximum
amount of
al lowable
le akag
e for the
most
common
aluminum
e
lectrolytics
and
dipped
solid
tantalum
capacitors. These
charts a re
duplicated
on
th
e
PULL
CHART
below
th
e
LC103. Good
capacitors
(as far as leakage
is
concerned) will
measure lower
than the amou nts shown
in the
Leakage
Charts.
v\Then
m
easuring
leakage, you do not
need
to wait
for the
readings
to drop to
zero
or to
its
lowest point
The
capacitor
is
good
for
any
leakage reading
which
is
lower
than
the
amount
shown
in the
chart.
Leakage
values shown
in
Table 4 for
alumin um
electrolytic
capacitors
are the
worst-case
conditions,
as
specified
by the
22
Electronic Industries
Association
(EIA)
standard
RS-395. The
values are
determined
by
the
formulas:
L
=
0.05
x CV (for CV
products less
than
1000)
or L
=
6 x
square
root of CV (for CV
products greater
than
1000.
(The
CV product is equal
to the
capacitance value
multiplied
by
the
voltage rating).
T he tantalum capacitor
leakage values
listed
in
Table 5
are
fo
r
the
most
common type
of
tantalu
m capacitors - dipped
solid,
type
3.3. These
values are specified
by
EIA
standard
RS-228B,
following
the
formula:
L
=
0.35
x square
root
of
CV. In
a few app lications outside of consumer
service,
tantalum
capac itors other
than
type 3.3
may
be
encountered.
Refer
to the
manufacturers specifications for
the
m aximum
allowable leakage
for
th
ese special capacitor
types.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ReZolver LC103 and is the answer not in the manual?

Questions and answers

Table of Contents