Section V
5-25.
HIGH FREQUENCY CALIBRATION.
5-26.
AMPLIFIER GAIN ADJUSTMENT .
The Test Oscillator used in this
procedure should be calibrated
at the end of its output cable.
a.
Connect test setup illustrated in Figure 5-2.
b.
Set Model 3400A RANGE switch and Test
Oscillator output attenuator to
c.
Adjust Test Oscillator output frequency for
400 HZj output voltage for 90% of full scale
as indicated on Model 3400A meter.
d.
Switch the Test Oscillator monitor switch to
expand mode and set a convenient reference
level.
e.
Change Test Oscillator frequency to 10 MHz.
Adjust Test Oscillator output voltage to main
tain reference level set in step d.
l.
Adjust C405 (10 MHz ADJUST) on
for 90% full scale as indicated on the Model
3400A meter.
Replace right side cover;
if
readjust C405
replacing the cover.
g.
Vary oscillator between 3 and 10 MHz; main
taining reference level set in step d.
ModeJ 3400A meter reading varies below 85%
or above 95% of full scale. repeat step f
until optimum response is obtained between
3 and 10 MHz.
5-27.
INPUT ATTENUATOR ADJUSTMENT.
The Test Oscillator used in this
procedure should
at the end of its output cable.
a.
Connect test setup illustrated in Figure 5-2.
b.
Set Model 3400A RANGE switch and Test
Oscillator output attenuator to 1 volt position.
Adjust Test Oscillator outp u t frequency for 400
c.
Hz, output voltage for
cated on Model 3400A meter.
d.
Switch the Test Oscillator monitor switch to
expand mode and set a convenient reference
level.
e.
Change Test Oscillator frequency to 100 kHz;
adjust Test Oscillator output voltage to main
tain reference level set in step d.
Remove Model 3400A bottom cover.
l.
CI02 (I V. 100 KC ADJ) for 90% full scale as
indicated on Model 3400A meter.
5 - 6
001 V position.
O.
A4
board
meter reading varies after
If
the
calibrated
be
full scale as indi
Adjust
g.
Vary oscillator between 100 kHz and 10 MHZj
maintain test oscillator output voltage to
reference level set in step d.
meter reading varies more than ±1% to 1 MHz.
:f:: 2 % from 1 MHz to 2 MHz. ±3% from 2 MHz
to 3 MHz. or ±5% from 3 MHz to 10 MHz. re
adjust C102 until optimum response is obtain
ed.
5-28.
SECOND ATTENUATOR ADJUSTMENT.
The Test Oscillator used in
this procedure should be
calibrated at the end of its'
output cable.
a.
Connect test setup tllustrated in Figure 5-2.
b.
Set Model 3400A RANGE switch and Test
Oscillator output attenuator to a 0 . 3 volt
position.
c.
Adjust Test Oscillator output frequency for
400 Hz; output voltage for 90% full scale as
indicated on Model 340QA meter.
d.
Switch the Test Oscillator monitor switch to
expand mode and set a convenient reference
level.
e.
Change Test Oscillator output frequency to
3 MHZj adjust output voltage to maintain re
ference level set in step d.
!.
Adjust C303 (10 MC . 3V ADJ) for 90% full
scale as indicated on Model 3400A meter.
g.
Vary Test Oscillator between 3 MHz and 10
MHz; maintain test oscillator output voltage
to reference level
3400A meter reading varies below 85% or
above 95% of full scale, repeat steps e and f
until optimum response is obtained between
3 and 10 MHz.
S-29. TROUBLESH O OTING PROCEDURES.
5-30.
This section contains procedures designed to
assist in the isolation of maHunctions. These opera
tions should
undertaken only after it has been es
be
tablished that the difficulty cannot be eliminated
the Adjustment and Calibration Procedures,
graph 5-15.
An investigation should also
ensure that the trouble is not a result of conditions
external to the Model 3400A.
5-31.
Conduct a visual check of the Model 3400A for
possible burned or loose components, loose connec
tions. or any other condition which might suggest a
source of trouble.
5-32.
Table
contains a summary of known pro
5-5
blems by front panel symptoms.
5-33.
Table 5-6 contains procedures which may
used as a guide in isolating malfunctions.
outlined in Table 5-6 are not designed to measure all
Model 3400A
If
Model 3400A
set in step d.
If Model
by
Para
made to
be
be
The checks