Meraw Aspen - Arm Blood Pressure Monitor Manual
![]()
Device Overview
Indications for Use
This Blood Pressure Monitor Meraw aspen is a digital monitor intended for use in measuring blood pressure and heartbeat rate with wrist circumference ranging from 51/3˝-8½˝(13.5 cm to 21.5 cm). It is intended for indoor, adult use only.
Contraindications
- The device is not suitable for use on the women who are or may be pregnant.
- The device is not suitable for use on patients with implanted electrical devices, such as cardiac pacemakers, defibrillators.
The signs below might be in the user manual, labeling or other component. They are the requirement of standard and using.
![]() | Symbol for "DIRECT CURRENT" | ![warning]() | Caution Indicates that caution is necessary when operating the device or control close to where the symbol is placed, or that the current situation needs operator awareness or operator action in order to avoid undesirable consequences. |
Measurement Principle
This product uses the Oscillometric Measuring Method to detect blood pressure. Before every measurement, the unit establishes a "zero point" equivalent to the atmospheric pressure. Then it starts inflating the cuff. Meanwhile, the unit detects pressure oscillation generated by beat-to-beat pulsatile, which is used to determine the systolic pressure and diastolic pressure as well as pulse rate.
Device Components
![Meraw - Aspen - Device Components - Part 1 Device Components - Part 1]()
![Meraw - Aspen - Device Components - Part 2 Device Components - Part 2]()
- LCD DISPLAY
- BLUETOOTH BUTTON
- START/STOP BUTTON
- CUFF
Component list of pressure measuring system:
- PCBA;
- Air Pipe;
- Pump;
- Valve;
- Cuff.
List
- Blood Pressure Monitor TMB-2085-K
- 2× AAA Batteries
- User manual
LCD Display
![Meraw - Aspen - LCD Display LCD Display]()
SYMBOL | DESCRIPTION | EXPLANATION |
SYS | Systolic blood pressure | The high pressure measured. |
DIA | Diastolic blood pressure | The low pressure measured. |
mmhg | mmHg | Measurement Unit of the blood pressure |
PULSE /min | Pulse display | Pulse in beats per minute. |
![]() | Average value | Displays average of last 3 readings. |
![]() | Heartbeat | Heartbeat dectetion during measurement. |
![]() | Irregular heartbeat | Irregular heartbeat detected during measurement. |
![]() | Wrist Positioning Guide | Guide for you to make the wrist in ideal position. |
![]() | Bluetooth transfer icon | The bluetooth transfer icon blinks when the bluetooth is working. |
![]() | Battery indicator | Indicate the current battery. |
![]() | Excessive Body Motion Detector | Appears when talking, moving, or hand shaking is detectedduring the measurement. NOTE: The measured blood pressure reading may not beaccurate if the icon is displayed. |
![]() | Memory display | Indicate it is in the memory mode and which group of memory it is. |
![]() | Current time | Time and date (year/month/day; hour: minute) |
![]() | Blood pressure level | Indicates the blood pressure level, |
![]() | Data pending to transmit | Appears in next measurement, when the data transmission fails. |
Device Set Up
Power Supply
Battery powered mode: 3V DC 2× AAA batteries
![]()
In order to get the best effect and protect your monitor, please use the right batteries which complies with local safety standard.
Installing Batteries
- Slide off the battery cover.
- Install or replace 2× AAA size batteries as indicated in the battery compartment. The display will light up and show the icon
. Press "START/ STOP" button to turn off, otherwise it will power off automatically after 10 seconds. Any time the battery is low, it will display the icon" bAt Lo" &
. It will power off automatically after 5 seconds. - Close the battery cover.
Replace the batteries if: - The low battery symbol appears on the display.
- When any button is pressed and nothing is displayed on the screen.
![]()
Do not use new and used batteries together.
Do not use different types of batteries together.
Do not dispose the batteries in fire. Batteries may explode or leak.
Remove batteries if the device is not likely to be used for some time.
Worn batteries are harmful to the environment. Do not dispose with daily garbage.
Remove the old batteries from the device following your local recycling guidelines.
Setting Date and Time
It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory.
Auto Setting
Once connected to smartphone App via Bluetooth, date and time will be syned to BPM automatically.
Manual Setting
- When the monitor is off, press and hold the "START/ STOP " button to display the date format. Press the "Bluetooth" button to switch the date format between [month/day] and [day/month].
![]()
- Press the "START/ STOP " button to confirm the date format, then the year will flash. Press the "Bluetooth" button to change the year.
![]()
- When you get the right year, press the "START / STOP" button to confirm the year. The screen will then show a blinking number representing the [MONTH].
- Repeat steps 2 and 3 to set the [MONTH] and [DAY].
![]()
- Repeat steps 2 and 3 to set the time format.
- Repeat steps 2 and 3 to set the [HOUR] and [MINUTE].
- Repeat steps 2 and 3 to confirm [Wrist Positioning Guide] on or off.
![]()
Note: When the Wrist Positioning Guide is OFF, the instruction will not appear at the start of the measurement.
- After confirming the [Wrist Positioning Guide],the LCD will display"donE"and then the monitor will turn off.
WRIST POSITIONING GUIDE
When using wrist blood pressure monitor, it is important to take the measurement while relaxing your wrist at heart level. This will help ensure an accurate reading will be taken. Because it is difficult to find the ideal location for a wrist measurement, we have included an optional Wrist Positioning Sensor that will assist in directing your wrist to the ideal location for a blood pressure measurement. Each time you take a measurement, the display will illuminate with different icons that are designed to help you move your wrist. Once the ideal location is found, the wrist symbol
will flash and after several seconds, measurement will begin. Keep your arm still until measurement is completed.
![]()
Move your wrist up.
![]()
Move your wrist down.
![]()
Wrist in ideal position, the symbol
flash.
![]()
Do not move, Measurement start.
Note: Due to differences in individual sizes and physique, this feature may not be helpful in all cases and you may wish to turn this feature OFF. If you feel the suggested wrist position does not match your heart level, please turn this feature OFF and follow your judgment.
Measurement
Put on device
- Remove all accessories (watch, bracelet, etc) from your wrist. If your physician has diagnosed you with poor circulation in your wrist, use the other one.
- Roll or push up your sleeve to expose the skin.
- Apply the cuff to your wrist with your palm facing up.
- Position the edge of the cuff about 1cm~2cm from wrist joints.
- Fasten the wrist cuff around your wrist, leaving no extra room between the cuff and your skin. If the cuff is too loose, the measurement will not be accurate.
- Sit comfortably with your tested wrist resting on a flat surface. Place your elbow on a table so that the cuff is at the same level as your heart. Turn your palm upwards. Sit upright in a chair, and take 5-6 deep breaths.
Sit Correctly
To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement.
- Keep palm relaxed during measurement
- Sit in a chair with your legs uncrossed and feet flat on the floor.
- Sit with your back and arm supported.
Start The Measurement
Notes:
- To stop the measurement, press the START/STOP button once to defalte the arm cuff.
- Remain still and quiet while taking a measurement.
Press the START/STOP button.
- The arm cuff will start to inflate automatically.
- The whole measurement takes around 35 seconds.
![Meraw - Aspen - Starting The Measurement Starting The Measurement]()
Measurement Tips
Patients with Hypertension:
The middle of the cuff should be at the level of the right atrium of the heart; Before starting measurement, please sit comfortably with legs uncrossed, feet flat on the floor, back and wrist supported. Rest for 5 minutes before measuring. Wait at least 3 minutes between measurements.
This allows your blood circulation to recover. The patient must relax as much as possible and do not move and talk during the measurement procedure. For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, on the same wrist, or as directed by a physician. Take the measurement in a silent room. The cuff should maintain at the same level as the right atrium of the heart. Do not cross your legs and keep your feet on the ground.
Keep your back against the backrest of the chair.
- Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for at least 30 minutes before taking a measurement.
- Rest for at least 5 minutes before taking a measurement.
Bluetooth & App
Download App
Meraw Health app is available both on Google play and App Store, search and download.
Make sure your phone has enough storage and meet lowest system version requirement before downloading.
![]()
Scan above QR code to download App and get App instruction.
![play.google.com]()
Android system:
Android 9.0 or above
![www.apple.com]()
iOS system:
iOS 10.0 or above
Add My Device
- Add device
After opening Meraw App for the first time, it will guide you to add new device.
![]()
- Choose device
Choose Meraw Aspen and continue
![]()
- Start Connection
Follow the instruction and start connection
![]()
- Connection successfuly
The connection process will complete automatically
View Data
![]()
All synced data will be listed by month, and show brief information including:
- Blood pressure level
- SYS
- DIA
- measured time
Some factors may influence the accuracy of measurement, click a data and add remark on it.
![]()
Track Trends
![]()
View your blood pressure in trend, clear at a glance
Data Management
Recall the records
- When the monitor is off, press "Bluetooth" button, the display will show the average value of the latest three records first.
- When the memory record is less than three groups, the display will show the latest record(memory record 01)
- Press "Bluetooth" button to get the record you want.
Each press "Bluetooth" butooth will increase the memory record by one in a cycling manner (AVG-01-02...).
Sync the records
- Turn on Bluetooth and open App on your smartphone before you sync the records.
![Meraw - Aspen - Data Management - Syncing the records Data Management - Syncing the records]()
- Each time you complete a measurement, data will be synced to App automatically,
will blink during uploading and turn solid on after uploading successful.
- If you data is not synced to App,
will show in data page, indicating the record is not uploaded to App.
- You could press 'Bluetooth' button when the blood pressure monitor is off to sync data that are not uploaded.
![]()
Delete the records
If you did not get the correct measurement, you can delete results by following steps below.
To delete a single measurement
- Enter the memory recall mode as described in section [Recall the Records]. Press "Bluetooth" button to get the measurement you would like to erase.
- Press and hold "Bluetooth" button for 3 seconds, and the display will show a blinking "dEL yES" along with the memory number of the reading.
- Use the "Start/Stop" or "Bluetooth" button to toggle between "dEL yES" and "dEL no", Press "START/STOP" to confirm the selection.
If "dEL yES" is selected, the unit will delete the record and display "dEL do nE" along with the memory number of the reading.
After then it will return to last memory record and All the records are pushed forward one digit (e.g., 03 becomes 02, and so on)
If "dEL no" is selected, it will stop the deletion.
![]()
To delete all measurements
- Enter the memory recall mode as described in section [Recall the Records].
- Press and hold "BLUETOOTH" and "RECORD" button for 5 seconds, and the display will show a blinking "dEL ALL" along with the user ID.
- Use the "BLUETOOTH" or "RECORD" button to toggle between "dEL ALL" and "dEL no", Press "START/STOP" to confirm the selection.
If "dEL ALL" is selected, the unit will display "dEL ALL do nE" + User ID and delete all the record of the current user. Several seconds later, it will display "---".
If "dEL no" is selected, it will stop the deletion.
![Meraw - Aspen - Data Management - Deleting all measurements Data Management - Deleting all measurements]()
ABOUT BLOOD PRESSURE
Systolic pressure and diastolic pressure
When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.
![]()
Only a physician can tell your normal BP range. Please contact a physician if your measuring result falls out of the range. Please note that only a physician can tell whether your blood pressure value has reached a dangerous point.
Standard blood pressure classification
The blood pressure classification published by AAC/AHA in 2017 is as follows:
![Meraw - Aspen - Standard blood pressure classification Standard blood pressure classification]()
Irregular Heartbeat Detector
An irregular heartbeat is detected when a heartbeat rhythm varies while the unit is measuring the systolic and diastolic blood pressure. During each measurement, the monitor records all the pulse intervals and calculate the average; if there are two or more pulse intervals, the difference between each interval and the average is more than the average value of ±25%, or there are four or more pulse intervals, the difference between each interval and the average is more than the average value of ±15%, the irregular heartbeat symbol appears on the display when the measurement results are appeared.
![]()
The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heart-beat was detected during measurement. Usually this is NOT a cause for concern. However, if the symbol appears often, we recommend you seek medical advice. Please note that the device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage
Blood pressure fluctuate throughout the day
- Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions.
- If the person takes medicine, the pressure will vary more.
- Wait at least 3 minutes for another measurement.
Different blood pressure at home compared to the hospital
The blood pressure is different even throughout the day due to weather, emotion, exercise etc.
Also, there is the "white coat" effect, which means blood pressure usually increases in clinical settings.
What you need to pay attention to when you measure your blood pressure at home:
If the cuff is tied properly.
If the cuff is too tight or too loose.
If the cuff is tied on the upper arm.
If you feel anxious.
Taking 2-3 deep breaths before beginning will be better for measuring.
Advice: Relax yourself for 4-5 minutes until you calm down.
Measuring on the right arm
It is ok for both arms, but there will be some different results for different people. We suggest you measure the same arm every time.
Trouble Shooting
If any abnormality arises during use, please check the following points:
PROBLEM | SYMPTOM | CHECK THIS | REMEDY |
No power | | Batteries are exhausted. | Replace with new batteries. |
Batteries are inserted incorrectly. | Insert the batteries correctly. |
Low batteries | bAt Lo & shows | Batteries are low. | Please replace batteries. |
Warning message | "out" shows | Out of measurement range | Relax for a moment. Refastenthe cuff and vthen measure again. If the problem persists, contact your physician. |
Error message | E 01 shows | The cuff is not secure | Refasten the cuff and then measure again. |
E 02 shows | The monitor detected motion, talking, or the pulse is too poor while measuring. | Relax for 5 minutes. and then keep still, measure again. |
E 03 shows | Pulse is not detected during measuring. | Loosen the clothing on the wrist and then measure again. |
E 04 shows | The measurement failed. | Relax for 5 minutes and measure again. |
EEx shows | Hardware error (X can be some digital symbol, such as 1, 2, etc.) | Turn off monitor and measure again. If EEx still appears on the display, please contact the retailer or our customer service |
NOTE: If the product still does not work, contact Meraw Customer Service. Under no circumstance should you disassemble or attempt to repair the unit by yourself.
Specifications
Power supply | Battery powered mode: 2x AAA batteries 3V ![]() |
Display mode | Digital LCD V.A.43 mm x 46 mm |
Measurement mode | Oscillographic testing mode |
Measurement range | Rated cuff pressure: 0 mmHg ~ 299 mmHg Measurement pressure: SYS: 60 mmHg ~ 230 mmHg DIA: 40 mmHg ~ 130 mmHg Pulse value: (40-199) beat/minute |
Accuracy | Pressure: 5°C-40°C within ±3 mmHg Pulse value: ±5% |
Working condition | A temperature range of: +5°C to +40°C A relative humidity range of 15% to 90%, non-condensing, but not requiring a water vapour partial pressure greater than 50 hPa An atmospheric pressure range of: 700 hPa to 1060 hPa |
Storage & transportation condition | Temperature: -20°C to +60°C A relative humidity range of ≤ 93%, non-condensing, at a water vapour pressure up to 50hPa |
Measurement perimeter of the wrist | About 13.5-21.5 cm |
Weight | Approx.103 g (Excluding the batteries) |
External dimensions | Approx. 85.7 mm × 60.8 mm × 24.7 mm (Excluding the cuff) |
Attachment Mode of operation | 2x AAA batteries, user manual Continuous operation |
Degree of protection | Type BF applied part |
Device Classification | Internally Powered ME Equipment |
IP Classification | IP22: The first number 2: Protected against solid foreign objects of 12,5mm Ф and greater. The second number: Protected against vertically falling water drops when enclosure titled up to 15º. Vertically falling drops shall have no harmful effects when the enclosure is titled at any angle up to 15º on either side of the vertical. |
Software Version | A01 |
Product Maintainance
![]()
- This device is intended for adult use in homes only.
- The device is not suitable for use on neonatal patients, pregnant women, patients with implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mastectomy. Please consult your doctor prior to using the unit if you suffer from illnesses.
- The device is not suitable for measuring the blood pressure of children. Ask your doctor before using it on older children.
- The device is not intended for patient transport outside a healthcare facility.
- The device is not intended for public use.
- This device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than the wrist or for functions other than obtaining a blood pressure measurement.
- Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood pressure. Do not begin or end medical treatment without asking a physician for treatment advice.
- If you are taking medication, consult your physician to determine the most appropriate time to measure your blood pressure. Never change a prescribed medication without consulting your physician.
- Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a medicine prescribed by a doctor. Consult your doctor if you have any question about your blood pressure.
- When the device is used to measure patients who have common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation, the best result may occur with deviation. Please consult your physician about the result.
![]()
Do not apply the cuff over a wound; otherwise it can cause further injury. - Do not inflate the cuff on the same limb which other monitoring ME equipment is applied around simultaneously, because this could cause temporary loss of function of those simultaneously-used monitoring ME equipment.
- On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff immediately. Prolonged high pressure (cuff pressure 300mmHg or constant pressure 15mmHg for more than 3 minutes) applied to the wrist may lead to an ecchymosis.
- Please check that operation of the device does not result in prolonged impairment of patient blood circulation.
- When measurement, please avoid compression or restriction of the connection tubing.
- The device cannot be used with HF surgical equipment at the same time.
- The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically investigated according to the requirements of ISO 81060-2:2018.
- To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the manufacturer.
- This device is contraindicated for any female who may be suspected of, or is pregnant. Besides providing inaccurate readings, the effects of this device on the fetus are unknown.
- Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries.
- This unit is not suitable for continuous monitoring during medical emergencies or operations. Otherwise, the patient's wrist and fingers will become anaesthetic, swollen and even purple due to a lack of blood.
- When not in use, store the device in a dry room and protect it against extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.
- This device may be used only for the purpose described in this booklet. The manufacturer cannot be held liable for damage caused by incorrect application.
- This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in this booklet.
- The maximum temperature that the applied part can be achieved is 42.5 while the environmental temperature is 40.
- The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air of with oxygen or nitrous oxide.
![]()
No servicing/maintenance while the ME equipment is in use. - The patient is an intended operator.
- The patient can measure data and change batteries under normal circumstances and maintain the device and its accessories according to the user manual.
- To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated interference signal or electrical fast transient/burst signal.
- The blood pressure monitor and the cuff are suitable for use within the patient environment. If you are allergic to polyester, nylon or plastic, please don't use this device.
- During use, the patient will be in contact with the cuff. The materials of the cuff have been tested and found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any potential sensization or irritation reaction.
- If you experience discomfort during a measurement, such as pain in the wrist or other complaints, press the START/STOP button to release the air immediately from the cuff. Loosen the cuff and remove it from your wrist.
- If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. Should the cuff not deflate when pressure reaches 40 kPa (300 mmHg), detach the cuff from the wrist and press the START/STOP button to stop inflation.
- Before use, make sure the device functions safely and is in proper working condition. Check the device, do not use the device if it is damaged in any way. The continuous use of a damaged unit may cause injury, improper results, or serious danger. Do not wash the cuff in a washing machine or dishwasher!
- The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. Thetypical service life is 10000 times.
- It is recommended that the performance should be checked every 2 years and after maintenance and repair, by retesting at least the requirements in limits of the error of the cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).
- Please dipose of any accessories, detachable parts, and the ME equipment according to the local guidelines
- Manufacturer will make available on request circuit diagrams, component part lists, descriptions, calibration instructions, etc., to assist to service personnel in parts repair.
- The operator shall not touch output of batteries and the patient simultaneously.
- Cleaning: Dust environment may affect the performance of the unit. Please use the soft cloth to clean the whole unit before and after use. Don't use any abrasive or volatile cleaners.
- The device doesn't need to be calibrated within two years of reliable service.
- If you have any problems with this device, such as setting up, maintaining or using, please contact the SERVICE PERSONNEL of Transtek. Don't open or repair the device by yourself in the event of malfunctions. The device must only be serviced, repaired and opened by individuals at authorized sales/service centers.
- Please report to Transtek if any unexpected operation or events occur.
- Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing of small parts. It is dangerous or even fatal.
- Be careful to strangulation due to cables and hoses, particularly due to excessive length.
- At least 30 min required for ME equipment to warm from the minimum storage temperature between uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use.
- This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS;
- Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a distance d away from the equipment. The distance d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz.
- Please use ACCESSORIES and detachable parts specified/ authorised by MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the user/patients.
- There is no luer lock connectors are used in the construction of tubing, there is a possibility that they might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood vessel.
- Please use the device under the environment which was provided in the user manual. Otherwise, the performance and lifetime of the device will be impacted and reduced.
Guangdong Transtek Medical Electronics Co., Ltd.
Zone A, No.105, Dongli Road, Torch Development District,
528437 Zhongshan, Guangdong, China
Distributor & Address
Meraw Labs Inc.
4500 Great America Parkway, Suite 1023, Santa Clara, CA 95054
Global Customer Service
support@merawlabs.com
www.merawlabs.com
![]()
Documents / Resources
References
Download manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Meraw Aspen - Arm Blood Pressure Monitor Manual
Need help?
Do you have a question about the Aspen and is the answer not in the manual?
Questions and answers