Active Mode; Report Mode - Percussionaire Travel Air TRUE-IPV Therapy Device User Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Active Mode

Pulse Amplitude
Usage Timer
At 16 seconds the PDM enters Active mode. The timer bar will change to a numeric
display, showing the current usage Session Timer. Above the timer is the pulse
amplitude display. This is calculated from the pressure measurements at the moment of
instantaneous peak and trough amplitude averaged over 5 seconds. The Active mode
display also shows the currently measured percussion rate/pulse frequency.
The PDM displays the usage Session Timer in minutes and seconds. The Session Timer is
the total time of the current usage. The Session Timer can display a maximum of
59 minutes and 59 seconds. If usage has been stopped for more than 5 minutes, the
Session Timer will reset and start over.
NOTE: To display the most recent usage duration time, see Report mode.

Report Mode

A.
B.
C.
Home IPV: Impulsator® Display
The Session Timer and the Total Usage
Timer (A) are displayed for 2 seconds,
followed by the System Information
page (B) for 2 seconds, alternating.
The alternating page display continues
for 5 minutes, or until usage resumes and
the PDM enters Active mode.
During the 5-minute period, a horizontal
bar graph indicates the time by moving
from left to right at a fixed rate. After
5 minutes of no usage, the system
information page is no longer displayed,
and the time display flashes (2 seconds
on, 2 seconds off) (C) for an additional
25 minutes.
When the Travel Air® is switched off,
the measurements drop to zero after a
few seconds. The PDM enters Sleep
mode after 25 minutes.
Pulse Frequency Rate
EN-7
EN