Page 1
Philips M1032A VueLink External Device Service Booklet Part Number M1032-9000Y 4512 610 18581 Published in October 2006 Edition 22 Y.00.00...
Page 2
Notice Philips makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Philips shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material.
Page 3
The Philips M1032A VueLink module is powered via the patient monitor. Purpose The Philips M1032A VueLink module transmits information from a connected external device to a Philips IntelliVue, V24 or V26 Patient Monitor or a Philips CMS Patient Monitoring System. The Philips M1032A VueLink module is not a therapeutic device.
Page 4
Philips IntelliVue, V24 or V26 Patient Monitor or a Philips CMS Patient Monitoring System. Part of Body The Philips M1032A VueLink module does not contact the body or tissue of the or Type of patient. Signals are obtained from external devices.
Page 5
Warnings, Cautions and Notes In this guide: • alerts you to a potential serious outcome, adverse event or Warning warning safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient. A caution alerts you where special care is necessary for the safe and effective Caution use of the product.
Page 6
Responsibility of the Manufacturer Philips Medical Systems only considers itself responsible for any effects on safety, reliability and performance of the equipment if: • assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons authorized by Philips, and •...
Page 7
VueLink module and all related Philips equipment. The configuration, setup and repair of external devices furnished by manufacturers other than Philips Medical Systems must be carried out by local site personnel, or by representatives of the external device manufacturer.
Page 8
Who Should This guide is for biomedical engineers or technicians responsible for connecting Use This external devices to Philips patient monitoring systems using the VueLink module. Guide How to Use This booklet provides information on the worldwide external devices that can be This Booklet connected to the VueLink module.
Connecting your Device to VueLink Connecting your Device to VueLink There are two different ways to connect your Device to VueLink: • Connection via a Supported Device driver. This uses one of the following device outputs: – Analog – RS-232 Digital –...
Philips personnel are only responsible for configuring Philips manufactured external devices. The signal labels used on the Philips patient monitor may be different from those Note given on the external device. The labels used on the Philips patient monitor are listed in the Philips M1032A VueLink External Device Instructions for Use.
Driver Part Numbers for Supported Devices Driver Part Numbers for Supported Devices Each driver is allocated a part number, using the driver key detailed below: Supported Device Information...
VueLink Cable Overview VueLink Cable Overview 3 Separate part numbers are assigned to each VueLink Cable: M1032A KXX option To order a cable with a new module. M1182A KXX option To order a cable for an existing module or for upgrades (Sales).
Auxiliary Devices Abbott Oximetrix 3 - SO Monitor Device Driver Name: Abbott Oximetrix 3 Device Driver P/N: M1032-A07rl where: r = revision l = language Supported Devices: Oximetrix (R)3 SO Monitor (version 104 and 105) Connection: RS-232 Digital (Fixed configuration) Baud Rate: 1200 Word Length:...
Page 24
Abbott Oximetrix 3 - SO2 Monitor Abbott Oximetrix 3 - SO Monitor Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Page 26
Critikon Dinamap 1846 / 1846 SX - NBP Monitor Critikon Dinamap 1846 / 1846 SX - NBP Monitor Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Danmeter AAI/AEP Monitor Danmeter AAI/AEP Monitor Device Driver Name: AAI/AEP Monitor Device Driver P/N: M1032-A08rl where: r = revision. l = language. Supported Devices: Danmeter AAI/AEP Monitor (SBC(software) Version 1.5) Connection: RS-232 Digital (at the rear of the device) Baud Rate: 9600 (Low Word Length: 8 bits...
Page 28
Danmeter AAI/AEP Monitor Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network. Auxiliary Devices...
Nellcor® N-200 - SpO2 Monitor Nellcor® N-200 - SpO Monitor Device Driver Name: Nellcor® N-200 Device Driver P/N: M1032-A05rl where: r = revision l = language Supported Devices: Nellcor® N-200 Connection: Analog / RS-232 Digital combination Baud Rate: 9600 Word Length: 8 bits Stop Bits: Parity:...
Page 30
Nellcor® N-200 - SpO2 Monitor Nellcor® N-200 - SpO Monitor Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Nellcor® N-100C - SpO2 Monitor Nellcor® N-100C - SpO Monitor Device Driver Name: Nellcor® N-100 Device Driver P/N: M1032-A04rl where: r = revision l = language Supported Devices: Nellcor® N-100 Connection: Analog: 1 wave, 2 numerics, no alarms Switch Settings: Set range switch to 0-10 V.
Page 32
The VueLink cable can get stuck in the outer cabinet without providing proper Note contact to the connectors. Apply moderate force to overcome this problem. Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Page 35
Nellcor Puritan-Bennett 740 / 760 / 840 Ventilators Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Puritan-Bennett 7200a / 7200ae Ventilator Puritan-Bennett 7200a / 7200ae Ventilator Device Driver Name: Bennett 7200a/ae Device Driver P/N: M1032-B02rl where: r = revision l = language Supported Devices: Puritan-Bennett 7200a (sw rev. V English only). Puritan-Bennett 7200ae (sw rev. D for Eng, Sp, Fr, Ger and Ital).
Page 37
Puritan-Bennett 7200a / 7200ae Ventilator Puritan-Bennett 7200a / 7200ae Ventilator Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Babylog 8000 Ventilator Dräger Babylog 8000 Ventilator Device Driver Name: Dräger Babylog 8000 Ventilator Device Driver P/N: M1032-B24rl where: r = revision l = language Supported Devices: Dräger Babylog 8000 with: • SW Revision: 5.01 • Digital Interface (optional): Babylink on MEDIBUS protocol 3.00.
Page 39
RS-232 Connector Dräger Babylog 8000 Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Evita Ventilator Dräger Evita Ventilator Device Driver Name: Dräger Evita Ventilator Device Driver P/N: M1032-B11rl where: r = revision l = language Supported Devices: Dräger Evita Typ 8410614 with • DW-Bus analog interface option number 8303940. • digital interface option number 8305327 •...
Page 41
Dräger Evita display after switching on the device are the same as the first two digits of the version number on the Evita’s interface cards. Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Evita 2 Ventilator Dräger Evita 2 Ventilator Device Driver Name: Dräger Evita 2 Ventilator Device Driver P/N: M1032-B23rl where: r = revision l = language Supported Devices: Dräger Evita 2 with: • SW Revision 2.00. • Digital interface (optional): EvitaLink 2.00 on MEDIBUS protocol 3.00.
Page 43
Dräger Evita 2 Ventilator Dräger Evita 2 Ventilator Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Page 45
1 Dräger Evita 2 Dura, Evita 4 and Evita XL Ventilator Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Graphic Screen for Babylog 8000 Dräger Graphic Screen for Babylog 8000 Device Driver Name: Dräger Graphic Screen for Babylog 8000 Device Driver P/N: M1032-B64rl where: r = revision l = language Supported Devices: Dräger Babylog 8000 Software 5.01 MEDIBUS Version 3.00. Dräger Graphic Screen version 2.n MEDIBUS Version 4.00.
Page 47
Dräger Graphic Screen for Babylog 8000 Dräger Graphic Screen Rear Panel Please refer to the Philips M1032A VueLink External Device User's Information for Note Dräger for Graphic Screen for Babylog 8000 for exact details of numerics, INOPs and alarms available from the external device via Philips patient monitoring network.
Dräger Graphic Screen for Savina Dräger Graphic Screen for Savina Device Driver Name: Dräger Graphic Screen for Savina Device Driver P/N: M1032-B63rl where: r = revision l = language Supported Devices: Dräger Savina Software 3.01 MEDIBUS Version 4.00 Dräger Graphic Screen version 2.n MEDIBUS Version 4.00.
Page 49
Dräger Graphic Screen for Savina Dräger Graphic Screen Rear Panel Please refer to the Philips M1032A VueLink External Device User's Information for Note Dräger for Graphic Screen for Savina for exact details of numerics, INOPs and alarms available from the external device via Philips patient monitoring network.
Dräger Savina Dräger Savina Device Driver Name: Dräger Savina Device Driver P/N: M1032-B66rl where: r = revision l = language Supported Devices: Dräger Savina Software 3.01 MEDIBUS Version 4.00 Connection: RS-232 port on rear side Baud Rate: 19200 Data Bits: 8 Start Bits: 1 Stop Bits: 1 Parity: None...
Page 51
Dräger Savina RS 232 Dräger Savina Rear Panel Please refer to the Philips M1032A VueLink External Device User's Information for Note Dräger for Savina for exact details of waves, numerics, INOPs and alarms available from the external device via Philips patient monitoring network.
GE Engström Carestation Ventilator GE Engström Carestation Ventilator Device Driver Name: GE Engström Device Driver P/N: M1032-B67rl where: r = revision l = language Supported Devices: GE Engström Software 3.x Ohmeda Com 1.3 Serial Protocol Connection: RS-232 port on rear side (Port 4) Baud Rate: 19200 Data Bits: 7 Start Bits: 1...
Page 53
GE Engström Carestation Ventilator RS232 GE Engström Carestation Ventilator Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs, and alarms available from the external device, and via Philips patient monitoring network.
Hamilton Veolar, Veolar FT, and Amadeus Ventilators Hamilton Veolar, Veolar FT, and Amadeus Ventilators Device Driver Name: Veolar Amadeus Device Driver P/N: M1032-B19rl where: r = revision l = language Supported Devices: • Veolar and Veolar FT Ventilators Control Processor and Front Panel sw rev: V30 and V33 Interface sw rev: NIK01 •...
Page 55
Hamilton Veolar, Veolar FT, and Amadeus Ventilators Hamilton Veolar Ventilator Rear Panel Hamilton Amadeus Ventilator Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs, and alarms available from the external device, and via Philips patient monitoring network.
BEAR 1000 Ventilator BEAR 1000 Ventilator Device Driver Name: BEAR 1000 Device Driver P/N: M1032-B28rl where: r = revision l = language Supported Devices: Bear 1000 (Basic and Comprehensive) SW Release 9.7 Bear GD 1000 Graphics Display (optional) Software P/N 51001-02007 Connection: Analog (2 waves) / RS-232 Digital combination.
Page 57
In the latter case, the RS-232 gender adapter must also be used. BEAR 1000 Ventilator Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Infrasonics® Infant Star / ISV Ventilator Infrasonics® Infant Star / ISV Ventilator Device Driver Name: Infrasonics® Infant Star / ISV Device Driver P/N: M1032-B30rl where: r = revision l = language Supported Devices: • Original Infant Star Ventilator (yellow casing): Standard Model TÜV Approved •...
Page 59
Infrasonics® Infant Star / ISV Ventilator Connection: The device driver supports the following configuration for the serial communications port: Baud Rate: 1200 Word Length: 8 bits Stop Bits: Parity: None Analog: 1 wave These settings are fixed and cannot be changed. Switch Settings: None Original Infrasonics Infant Star Ventilator Rear Panel...
Page 60
Infrasonics® Infant Star / ISV Ventilator New Infrasonics Infant Star Ventilator Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Siemens SCM 990 Ventilator Siemens SCM 990 Ventilator Device Driver Name: Siemens SCM 990 AD (Adult) Siemens SCM 990 NEO (Neonatal) Device Driver P/N: M1032-B08rl (Adult) M1032-B14rl (Neonatal) where: r = revision l = language Supported Devices: Siemens 900 C/D/E with Servo Computer Module 990 (serial number 713 or greater, sw rev 2.0).
Page 62
Siemens SCM 990 Ventilator Transmission Switch Settings Switch Name Position Number ADDRESS Code Switches 0-9 (ANY) 0-9 (ANY) PARITY Switches (for RS-232 configuration) BAUD RATE switch The restart switch is to the right of the BAUD RATE switch. If any of the above Note switches are changed during power-on;...
Page 63
Siemens SCM 990 Ventilator Siemens SCM990 Ventilator Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Page 65
Siemens 900 C/D/E Ventilator Siemens 900 C/D/E Ventilator Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Maquet Servo 300/300A Ventilator Maquet Servo 300/300A Ventilator Device Driver Name: Maquet Servo 300/300A Device Driver P/N: M1032-B20rl where: r = revision l = language Supported Devices: Maquet Servo 300/300A Software revision 2.01 Connection: Port N82 on rear side Baud Rate: 9600 Word length: 8 bits...
Page 67
All waves and numerics are now transferred to the VueLink module using the digital cable connected to port N82. Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Maquet SERVO-i Ventilator Maquet SERVO-i Ventilator Device Driver Name: Maquet SERVO-i Device Driver P/N: M1032-B55rl where: r = revision l = language Supported Devices: SERVO-i System version 2.0, 3.0 Connection: Computer Interface Emulator Baud Rate: 9600 Word length: 8 bits Start Bits: Stop Bits: Parity:...
Page 69
Maquet SERVO-i Ventilator Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network. Ventilators...
Maquet SERVO-s Ventilator Maquet SERVO-s Ventilator Device Driver Name: Maquet SERVO-s Device Driver P/N: M1032-B65rl where: r = revision l = language Supported Devices: SERVO-s System version 3.0 Connection: Computer Interface Emulator Baud Rate: 9600 Word length: 8 bits Start Bits: Stop Bits: Parity: Even...
Page 71
Maquet SERVO-s Ventilator RS 232 Maquet SERVO-s Ventilator Back Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Device Driver Name: Philips M1025A/B Device Driver P/N: M1032-B06rl where: r = revision l = language Supported Devices: Philips M1025A (Protocol Rev. A, serial Number 31xx and lower) and Philips M1025B. Connection: RS-232 Digital Baud Rate: 9600 Parity: Even Stop Bit:...
Page 74
HP M1025A/B Gas Analyzer Philips M1025A/B Gas Analyser Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Datex Capnomac (II and Ultima) Gas Analyzer Datex Capnomac (II and Ultima) Gas Analyzer Start Up Procedure The driver automatically identifies the connected device at start-up. For this reason, the following start-up procedure must be followed whenever the user changes the Datex Device, even if it is the same model: 1.
Page 76
Datex Capnomac (II and Ultima) Gas Analyzer l = language Supported Devices: • Datex Capnomac II • Datex Capnomac II four channel analog output option • Datex Ultima All units shipped after Feb 21st 1994 Connection: Analog / RS-232 Digital combination Baud Rate: 1200 Word Length:...
Page 77
Datex Capnomac (II and Ultima) Gas Analyzer Datex Capnomac (II and Ultima) Gas Analyzer rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Vamos Variable Anaesthetic Gas Monitor Dräger Vamos Variable Anaesthetic Gas Monitor Device Driver Name: Dräger Vamos Device Driver P/N: M1032-B54rl where: r = revision l = language Supported Devices: Dräger Vamos Software 2.0 MEDIBUS Version 4.03 Connection: MEDIBUS RS-232 port on rear side Baud Rate: 9600 Data Bits:...
Page 79
Dräger Vamos Variable Anaesthetic Gas Monitor Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network. Gas Analyzers...
Ohmeda RGM 5250 Respiratory Gas Monitor Ohmeda RGM 5250 Respiratory Gas Monitor Device Driver Name: Ohmeda RGM 5250 Respiratory Gas Monitor Device Driver P/N: M1032-B22Bl where: r = revision l = language Supported Devices: Ohmeda RGM 5250 with SW Revision 6.0 Supported Device Options: •...
Page 81
Ohmeda RGM 5250 Respiratory Gas Monitor Ohmeda RGM 5250 Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Ohmeda Rascal II Anesthetic Gas Monitor Ohmeda Rascal II Anesthetic Gas Monitor Device Driver Name: Ohmeda Rascal II Device Driver P/N: M1032-B44rl where: r = revision l = language Supported Devices: Ohmeda Rascal II Anesthetic Gas Monitor with SW Revision 1.12 Connection: RS-232 Digital and Analog Baud Rate:...
Page 83
Ohmeda Rascal II Anesthetic Gas Monitor Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network. Gas Analyzers...
Page 84
Ohmeda Rascal II Anesthetic Gas Monitor Gas Analyzers...
Anesthesia Machines Dräger Cato Anesthesia Device Device Driver Name: Dräger Cato Anesthesia Device Device Driver P/N: M1032-B25rl where: r = revision l = language Supported Devices: Dräger Cato - Screen Type PM 8050 CD with • SW Revision 2.01, 2.02 •...
Page 86
Dräger Cato Anesthesia Device Dräger Cato Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Cicero Anesthesia Machine Dräger Cicero Anesthesia Machine Device Driver Name: Dräger Cicero Anesthesia Machine Device Driver P/N: M1032-B12rl where: r = revision l = language Supported Devices: PM8020 datamanager Software revision 4.6 only Connection: RS-232 Digital Baud Rate: 9600 (fixed) Word Length: 8 bits fixed) Stop Bits:...
Page 88
Philips cannot accept any responsibility for information displayed on the Philips patient monitor via VueLink. Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Cicero EM Anesthesia Machine Dräger Cicero EM Anesthesia Machine Device Driver Name: Dräger Cicero EM Anesthesia Machine Device Driver P/N: M1032-B36rl where: r = revision l = language Supported Devices: PM 8060 datamanager Software revision 2.01 only. Connection: RS-232 Digital Baud Rate: 9600 Word Length:...
Page 90
Note VueLink supports Isoflurane, Enflurane, Halothane, Sevoflurane, and Desflurane Note gases only. If the Cicero EM is configured with any other unsupported gas, Philips cannot accept any responsibility for information displayed on the Philips patient monitor via VueLink. Please refer to the Philips M1032A VueLink External Device Instructions for Use...
Dräger Fabius GS/Tiro Anesthesia Machine Dräger Fabius GS/Tiro Anesthesia Machine Device Driver Name: Dräger Fabius GS/Tiro Device Driver P/N: M1032-B53rl where: r = revision l = language Supported Devices: Dräger Fabius GS/Tiro Software 3.01 Connection: RS-232 port COM1 Baud Rate: 9600 Word Length: Start Bits:...
Page 92
Dräger Fabius GS/Tiro Anesthesia Machine Dräger Fabius GS/Tiro Rear Panel showing the location of COM1 Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Julian Anesthesia Workstation Dräger Julian Anesthesia Workstation Device Driver Name: Dräger Julian Device Driver P/N: M1032-B42rl where: r = revision l = language Supported Devices: PM 8055 datamanager • Software Revision 3.02. • Digital Interface: MEDIBUS Protocol 3.00. Connection: RS-232 Digital Baud Rate: 9600...
Page 94
Dräger Julian Rear Panel VueLink supports Isoflurane, Enflurane, Halothane, Sevoflurane and Desflurane Note gases only. If the Dräger Julian is configured with any other unsupported gas, Philips cannot accept any responsibility for information displayed on the Philips patient monitor via VueLink.
COM 1 (Medibus) Dräger Pallas Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger PM 8050 Anesthesia Machine Dräger PM 8050 Anesthesia Machine Device Driver Name: Dräger PM 8050 Device Driver P/N: M1032-B32rl where: r = revision l = language Supported Devices: Dräger PM 8050 with: • Software Revision 2.02, 2.03. • Digital Interface: MEDIBUS Protocol 3.00. Connection: RS-232 Digital Baud Rate:...
Page 97
Philips cannot accept any responsibility for information displayed on the Philips patient monitor via VueLink. Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Primus/Apollo Anesthesia Workstation Dräger Primus/Apollo Anesthesia Workstation Device Driver Name: Dräger Primus/Apollo Device Driver P/N: M1032-B52rl where: r = revision l = language Supported Devices: Dräger Primus Software 2.02 MEDIBUS Version 4.03 Dräger Apollo Software 3.0 MEDIBUS Version 4.03 Connection: RS-232 port COM1 on rear side Baud Rate:...
Page 99
COM 1 (Medibus) Dräger Primus/Apollo Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Dräger Zeus Anesthesia Device Dräger Zeus Anesthesia Device Device Driver Name: Dräger Zeus Device Driver P/N: M1032-B58rl where: r = revision l = language Supported Devices: Dräger Zeus Software 3.n • MEDIBUS Version 4.03 Connection: RS-232 port COM1 or COM2 on rear side Baud Rate: 9600 Word Length:...
Page 101
COM2 COM1 Dräger Zeus Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for Dräger Zeus for exact details of waves, numerics, INOPs and alarms available from the external device via Philips patient monitoring network.
GE Aestiva and Avance GE Aestiva and Avance Device Driver Name: GE Avance and Aestiva Device Driver P/N: M1032-B39rl where: r = revision l = language Supported Devices: GE Avance Software 3.0 Ohmeda Com 1.2 Serial Protocol GE Aestiva Software 4.5 Ohmeda Com 1.0 Serial Protocol Connection: RS-232 Digital...
Page 103
GE Aestiva and Avance • RS-232 Rear of GE Aestiva Anesthesia Machines...
Page 104
• RS-232 Rear of GE Avance Please refer to the Philips M1032A VueLink External Device Instructions for Use Note Information for GE Avance and Aestiva for exact details of waves, numerics, INOPs and alarms available from the external device via Philips patient monitoring network.
GE Aisys GE Aisys Device Driver Name: GE Aisys Device Driver P/N: M1032-B60rl where: r = revision l = language Supported Devices: GE Aisys Software 1.x Ohmeda Com 1.2 Serial Protocol Connection: RS-232 port on rear Baud Rate: 19200 Word Length: 7 bits Start Bits: Stop Bits:...
Page 106
GE Aisys Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device via Philips patient monitoring network. Anesthesia Machines...
North American Dräger Narkomed 2B and Narkomed 2C Anesthesia Systems North American Dräger Narkomed 2B and Narkomed 2C Anesthesia Systems Device Driver Name: NARKOMED 2B/2C/3200 Device Driver P/N: M1032-B31rl where: r = revision l = language Supported Devices: Narkomed 2B Anesthesia System Software Revision 2.05 Narkomed 2C Anesthesia System Software Revision 1.11...
Page 108
• The above VueLink module is measuring one or more numerics which are also output by the Narkomed VueLink interface. (For example, the Philips M1025A/ B Anesthetic Gas Analyzer numerics include !imCO2, !etCO2, !inN2O, etN2O, !FIO2, !inAGT, !etAGT, !PULSE, !SpO2 and !AWRR. These correspond to the Narkomed VueLink interface numerics: IMCO2, ETCO2, inN2O, etN2O, FIO2, inAGT, etAGT, PULSE, SpO2 and AWRR.)
Page 109
North American Dräger Narkomed 2B and Narkomed 2C Anesthesia Systems Narkomed 2B: Rear View Narkomed 2C: Rear View Anesthesia Machines...
Page 110
North American Dräger Narkomed 2B and Narkomed 2C Anesthesia Systems Vitalert 3200: Rear View Communications Hub: Rear View Anesthesia Machines...
Page 111
North American Dräger Narkomed 2B and Narkomed 2C Anesthesia Systems Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
• The above VueLink module is measuring one or more numerics which are also output by the Narkomed VueLink interface. (For example, the Philips M1025A/ B Anesthetic Gas Analyzer numerics include !imCO2, !etCO2, !inN2O, etN2O, !FIO2, !inAGT, !etAGT, !PULSE, !SpO2 and !AWRR. These correspond to the Narkomed VueLink interface numerics: IMCO2, ETCO2, inN2O, etN2O, FIO2, inAGT, etAGT, PULSE, SpO2 and AWRR.)
Page 113
Narkomed VueLink interface. Narkomed 4: Rear View Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
North American Dräger Narkomed GS Anesthesia System North American Dräger Narkomed GS Anesthesia System Device Driver Name: NAD NARKOMED GS Device Driver P/N: M1032-B45rl where: r = revision l = language Supported Devices: Narkomed GS Anesthesia System Software Revision 1.10 Connection: RS-232 Digital Protocol:...
Page 115
North American Dräger Narkomed GS Anesthesia System Narkomed GS: Rear View Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
North American Dräger Narkomed 6000 Anesthesia System North American Dräger Narkomed 6000 Anesthesia System Device Driver Name: NAD NARKOMED 6000 Device Driver P/N: M1032-B49rl where: r = revision l = language Supported Devices: North American Dräger Narkomed 6000 Anesthesia System Software Revision 1.06 Connection: RS-232 Digital (use one of the ports: com5 -...
Page 117
North American Dräger Narkomed 6000 Anesthesia System com5 to com8 NAD Narkomed 6000: Rear View Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Ohmeda 7800/7810 Ohmeda 7800/7810 Device Driver Name: OHMEDA 7800/7810 Device Driver P/N: M1032-B37rl where: r = revision l = language Supported Devices: Ohmeda 7800 in Excel Anesthesia Machine Software Revision 4.22 Ohmeda 7810 in Modulus II Anesthesia Machine Software Revision 4.22 Connection: Baud Rate: 1200...
Page 119
Ohmeda 7800/7810 Rear Panel of Ohmeda 7810 Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Ohmeda 7900 Ohmeda 7900 Device Driver Name: OHMEDA 7900 Device Driver P/N: M1032-B39rl where: r = revision l = language Supported Devices: Software Revision 3.5 Connection: RS-232 Digital Baud Rate: 19200 Word Length: 7 bits Stop Bits: Parity: RS-232 Underside of Ohmeda 7900 Anesthesia Machines...
Page 121
Ohmeda 7900 Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network. Anesthesia Machines...
Ohmeda Modulus CD Anesthesia Machine Ohmeda Modulus CD Anesthesia Machine Device Driver Name: Ohmeda Modulus CD Device Driver P/N: M1032-B27rl where: r = revision l = language Supported Devices: Ohmeda Modulus CD (Base Type) Software versions 3.50 and 4.00 are supported. Connection: Connect the interface cable to the A1 serial port of the Ohmeda Modulus CD, which is located on...
Page 123
Ohmeda Modulus CD Anesthesia Machine Set up the Ohmeda Modulus CD as follows: • For software version 3.50: 1. Press MANUAL key twice on display pod. 2. Select EXTERNAL PORT DEVICE and change it to ARKIVE. 3. Change the Printer Logging Interval to 20 seconds. •...
Page 124
Ohmeda Modulus CD Anesthesia Machine Rear Panel of the Ohmeda Modulus CD Anesthesia Machine Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Taema Alys Anesthesia Machine Taema Alys Anesthesia Machine Device Driver Name: Taema Alys Device Driver P/N: M1032-B29rl where: r = revision l = language Supported Devices: Taema Alys, all versions with software driver version V02.40 and upwards. Connection: Connect the device driver to the serial port of the Taema Alys.
Page 126
Taema Alys Anesthesia Machine Rear Panel of the Taema Alys Ventilator Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Auxiliary Plus Devices Aspect Medical Systems A-2000™ BIS™ Monitor Device Driver Name: Aspect A-2000 BIS Device Driver P/N: M1032-B47rl where: r = revision l = language Supported Devices: Aspect Medical Systems A-2000 BIS Serial Protocol Revision 1.07 Connection: RS-232 Digital Protocol: ASCII Word Length:...
Page 128
Aspect Medical Systems A-2000™ BIS™ Monitor A-2000 BIS Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Baxter Explorer Cardiopulmonary Monitor Baxter Explorer Cardiopulmonary Monitor Device Driver Name: Baxter Explorer Device Driver P/N: M1032-B15rl where: r = revision l = language Supported Devices: Baxter Explorer (sw rev. 06.37) Connection: RS-232 Digital (Fixed configuration) Baud Rate: 9600 Word Length: 8 bits Stop Bits: Parity:...
Page 130
Baxter Explorer Cardiopulmonary Monitor Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network. Auxiliary Plus Devices...
B|Braun SpaceCom B|Braun SpaceCom Device Driver Name: B|Braun SpaceCom Device Driver P/N: M1032-B61rl where: r = revision l = language Supported Devices: B|Braun SpaceCom with Protocol Version V3.25 Infusomat Space Version 686 Infusomat Space P Version 687 Perfusor Space Version 688 Connection: RS232 port on rear side Baud Rate:...
Page 132
B|Braun SpaceCom and the VueLink module is lost. You have to restart the BBraun SpaceCom to reestablish the communication. Please upgrade the IntelliVue Software to Rev. C.00.90 (patch) or higher or Caution IntelliVue Rev. D.0 to avoid any loss of VueLink data from the resting display.
Page 133
B|Braun SpaceCom Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device via Philips patient monitoring network. Auxiliary Plus Devices...
Braun FM Device Information Braun FM Device Information Device Driver Name: Braun fm-system Device Driver P/N: M1032-B33rl where: r = revision l = language Supported Devices: Braun fm system with fm computer Software Revision 3.10 Connection: RS-232 Baud Rate: 9600 Word Length: 8 bits Stop Bits:...
Page 135
Braun FM Device Information RS-232 Connector Braun FM Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device via Philips patient monitoring network.
Diametrics Medical IRMA SL Series 2000 Diametrics Medical IRMA SL Series 2000 Device Driver Name: IRMA SL Series 2000 Device Driver P/N: M1032-B50rl where: r = revision l = language Supported Devices: Diametrics Medical IRMA SL Series 2000 Software version 5.1.x Connection: RS-232 Digital Baud Rate:...
Page 137
Diametrics Medical IRMA SL Series 2000 Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device via Philips patient monitoring network. Auxiliary Plus Devices...
Diametrics Medical Trendcare Monitor Diametrics Medical Trendcare Monitor Device Driver Name: Trendcare Monitor Device Driver P/N: M1032-B51rl where: r = revision l = language Supported Devices: Diametrics Medical Trendcare Monitor Software version 2.70.01 Connection: RS-232 Digital (Fixed configuration) Baud Rate: 9600, 19200 Data Bits: Start Bits:...
Page 139
COM2 COM1 Connector Senior Monitor Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device via Philips patient monitoring network. Auxiliary Plus Devices...
Edwards CardiacOutput-Oximetry Monitor Family Edwards CardiacOutput-Oximetry Monitor Family Device Driver Name: Edwards CardiacOutput-Oximetry Monitor Family Device Driver P/N: M1032-B16rl where: r = revision l = language Supported Devices: Edwards Vigilance Edwards Vigilance II Edwards Vigileo Connection: RS-232 Digital Baud Rate: 19200 Parity: None...
Page 141
Edwards CardiacOutput-Oximetry Monitor Family COM1 Fig. 1:COM1 position on Vigilance (backside) Edwards Vigilance Rear Panel COM1 Edwards Vigilance II Rear Panel Auxiliary Plus Devices...
Page 142
Connecting to devices with male connectors Some releases of the Edwards CardiacOutput-Oximetry Monitor Family devices have a male COM-Port socket. The Philips cable provides a “gender-changer” to connect these devices to the VueLink module. The following instructions describe, how to assemble the “gender-changer”...
Page 143
Connect the assembled VueLink-cable and gender-changer to COM1 of the Edwards CardiacOutput-Oximetry Monitor Family device. VueLink-cable Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device via Philips patient monitoring network.
Fresenius Vial Base A/DPS/MVP Fresenius Vial Base A/DPS/MVP The Fresenius Vial Base A can be configured to support either six or eight Warning DPS or MVP modules. If the base is configured to support six modules, and more than six modules are connected, VueLink provides only numerics, alarms, and inops from the first six modules.
Page 145
Fresenius Vial Base A/DPS/MVP Fresenius Vial Base A Module Rear Panel Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for Fresenius Vial Base A/DPS/MVP for exact details of numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
GE DINAMAP ProCare 100 Series GE DINAMAP ProCare 100 Series Device Driver Name: GE DINAMAP ProCare 100 Series Device Driver P/N: M1032-B72rl where: r = revision l = language Supported Devices: GE DINAMAP ProCare 100 Series Connection: RS-232 port on rear side Baud Rate: 9600 Data Bits: 8 Start Bits: 1...
Page 147
GE DINAMAP ProCare 100 Series Note Please refer to the Philips M1032A VueLink External Device Instructions for Use for exact details of numerics and INOPs available from the external device. Auxiliary Plus Devices...
Page 149
1 Analyzer Note Please refer to the Philips M1032A VueLink External Device Instructions for Use for exact details of numerics and INOPs available from the external device. Auxiliary Plus Devices...
Mortara ELI 100/STM 12-Lead Monitor Mortara ELI 100/STM 12-Lead Monitor Device Driver Name: Mortara ELI 100/STM Device Driver P/N: M1032-B26rl where: r = revision l = language Supported Devices: Mortara ELI 100/STM with: • Software part no: 11001-001-20 • Software revision: 1.95, 11.95 Connection: RS-232 Digital (fixed configuration) Baud Rate:...
Page 151
Mortara ELI 100/STM 12-Lead Monitor Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network. Auxiliary Plus Devices...
Organon Teknika TOF-Watch SX Organon Teknika TOF-Watch SX This information only applies to the special Philips version of the TOF-Watch SX. Note Device Driver Name: TOF-Watch SX Device Driver P/N: M1032-B46rl where: r = revision l = language Supported Devices: Organon Teknika TOF-Watch SX (Philips version) Software Revision 1.2...
Page 153
TOF Watch SX Interface TOF-Watch SX: Connecting Please refer to the Philips M1032A VueLink External Device Instructions for Use Note for exact details of waves, numerics, INOPs and alarms available from the external device and via Philips patient monitoring network.
Page 154
Organon Teknika TOF-Watch SX Auxiliary Plus Devices...
Free Analog Device Information Free Analog Device Information This appendix contains information about Free Analog Devices that can be connected to the VueLink module. The following information is provided: • General configuration and cable information about Free Analog Devices. • Step by step instructions on how to configure the following Free Analog Device into the VueLink module.
If an external device name does not appear in the Preselection Task Window on the Philips patient monitor, but has an analog interface, it can be connected to the VueLink Module using the Free Analog Device driver. The device information must however be manually configured into the module using the Task Windows in the System’s Configuration Mode.
Page 157
Configuring the Free Analog Device The Free Analog cable options are detailed in the matrix below: Patient New Module SMO/SLI Monitor Cable Option Option Option Devicecol (M1032A-) (M1032-) (M1182A-) Free Analog 61611 61612 Devices Free Analog Device Information...
Dräger Evita Ventilator Dräger Evita Ventilator This note provides information on configuring the Evita Free Analog Device into the VueLink module. This procedure should take approximately 2 hours and consists of 6 steps:- General information on configuring the free analog devices has not been included in this technical note.
Page 159
Dräger Evita Ventilator Step 1: Ensure that all the necessary hardware is available (10 Minutes) You will need:- • An M1032A #A02 VueLink Module • 4m Free Analog (open end) cable M1032A #K6A or M1182A #K6A 2m Free Analog cable M1032A #K60 or M1182A #K60. •...
Page 160
Dräger Evita Ventilator Step 2: Prepare the free analog cable (45 Minutes) 1. Connect the black and white wires that are twisted together to the Evita Analog ground (Evita pin 1). Do NOT separate these wires. 2. Connect the thick black wire, or the copper shield, to the case resp. ground on the device (Evita pin 32).
Page 161
Dräger Evita Ventilator Free Analog Cable EVITA Designation Channel Wire Colours Black and White Analog ground (twisted) Black (thick) Shield Wave 1 Brown Airway Pressure Wave 2 Expiratory Flow Numeric 1 Orange Minute Volume Numeric 2 Yellow Airway Resp. Rate Numeric 3 Green Tidal Volume...
Page 162
Dräger Evita Ventilator Pins used for the Evita free analog interface Step 3: Enter a device name, device-id and channel labels (10 Minutes) Only the Expiratory Flow Label (EXPFL) needs to be entered with the handheld keypad. All the other Evita waves and numerics are offered in the configuration Task Window of the Ventilator Free Analog Device Driver.
Page 163
Dräger Evita Ventilator Name Recommended Label Expired Tidal Volume V Inspired O Concentration !FIO2 1. Referred to as Airway Respiration Rate in patient monitor Alarms No alarms are activated for the Evita Free Analog Device Driver. Free Analog Device Information...
Page 164
Dräger Evita Ventilator Step 4: Configure each wave and numeric channels (30 Minutes) The configurations for each wave and numeric channel are provided in the following table. Value Range = Decimal Channel Label Unit Voltage Range Represent. Wave 1 mbar -100 to 0 mbar-10 to 0 V 0 to 100 mbar0 to 10 V Wave 2...
Page 165
Dräger Evita Ventilator Step 5: Connect the free analog cable to the Evita analog output (10 Minutes) Step 6: Verify configuration with the user (15 Minutes) Ensure that the correct waves and numerics are being interfaced from the Evita. Free Analog Device Information...
Engström ELSA/EAS Anesthesia System Engström ELSA/EAS Anesthesia System (This device can be connected in Germany - SUV approval is available). This note provides information on configuring the ELSA/EAS Free Analog device into the VueLink module. This procedure should take approximately 130 minutes and consists of 7 steps:- General information on configuring the free analog devices has not been included in this technical note.
Page 167
Engström ELSA/EAS Anesthesia System Step 1: Ensure that all the necessary hardware is available (10 Minutes) You will need:- • An M1032A #A04 VueLink Module • 4m Free Analog (open end) cable M1032A #K6A or M1182A #K6A 2m Free Analog cable M1032A #K60 or M1182A #K60. •...
Page 168
Engström ELSA/EAS Anesthesia System Step 2: Select two waveforms (10 Minutes) The following parameters are provided as a waveform at the Analog Output port of ELSA/EAS: • Expiration Flow (ELSA/EAS pin 2) • Airway Pressure (ELSA/EAS pin 15) • Concentration (ELSA/EAS pin 5) •...
Page 169
Engström ELSA/EAS Anesthesia System Step 3: Prepare the free analog cable (45 Minutes) 1. Connect the black and white wires that are twisted together to the ELSA/ EAS Analog ground (pin 18). Do NOT separate these wires. 2. Connect the thick black wire, or the copper shield, to the case resp. ground on the device (ELSA/EAS pin 1).
Page 170
Engström ELSA/EAS Anesthesia System Pins used for the ELSA/EAS free analog interface Free Analog Device Information...
Page 171
Engström ELSA/EAS Anesthesia System Step 4: Enter a device name, device-id and channel labels (10 Minutes) Only the AWP and the MIVOL labels are offered in the configuration window of the Ventilator Free Analog Driver. All the other ELSA/EAS parameter labels need to be entered with the handheld keypad.
Page 172
Engström ELSA/EAS Anesthesia System The configurations for each wave and numeric channel are provided in the following table. Value Range = Decimal Channel Label Unit Voltage Range Represent. Wave 1 CO2CC mmHg 0 to 76 mmHg = 0 to 1 V Wave 2 0 to 100 cmH O = 0 to 1 V...
Page 173
Engström ELSA/EAS Anesthesia System Step 6: Connect the free analog cable to the ELSA/EAS analog output (10 Minutes) Step 7: Verify configuration with the user (15 Minutes) Ensure that the correct waves and numerics are being interfaced from the ELSA/ EAS.
Germany. • Differences in the displayed numerics can occur between the Nellcor and the Philips patient monitor, if exceptionally unstable monitoring occurs (such as crying babies). This is due to the different numeric update rates between both devices.
Page 175
Nellcor N1000 and N2500 Multi-Function Monitor General information on configuring the free analog devices has not been included in this technical note. Please refer to Chapter 3 of the VueLink Handbook, if you are not familiar with this procedure. Step 1: Ensure that all the necessary hardware is available (10 Minutes) You will need:- •...
Page 176
Nellcor N1000 and N2500 Multi-Function Monitor 3. Put Cable 1 through the sleeve in A and crimp the sleeve on the inside of the connector hood, so that the cable cannot be torn out. Repeat this procedure for all the other cables. (As there are only 5 holes, two cables must be fitted through single holes in two cases).
Page 177
Nellcor N1000 and N2500 Multi-Function Monitor VueLink Free Analog Cable Designation Cable Channel Wire Colours Numeric 2 Yellow EtCO Numeric 3 Green Resp Numeric 4 Blue Numeric 5 Purple Pulse Free Analog Device Information...
Page 178
Nellcor N1000 and N2500 Multi-Function Monitor Step 3: Enter a device name, device-id and channel labels (10 Minutes) Only the Pleth, Resp and Pulse labels need to be entered with the handheld keypad. All the other Nellcor waves and numerics are offered in the configuration window of the Nellcor Free Analog Device Driver.
Page 179
Nellcor N1000 and N2500 Multi-Function Monitor The configurations for each numeric are provided in the following table. Value Range = Decimal Channel Label Unit Voltage Range Represent. Wave 1 Pleth 0 to 100 = 0 to 10 V Wave 2 mmHg 0 to 76 mmHg = 0 to 10 V Numeric 1...
Page 180
Nellcor N1000 and N2500 Multi-Function Monitor Step 5: Configure each Nellcor analog output (30 Minutes) The following settings should be stored in slot 10 (defaults) of the Nellcor device: Analog Signal Full scale output voltage Output Port Pulse wave 10.00V wave 10.00V =76 mmHg 10.00V = 100%...
Page 181
Nellcor N1000 and N2500 Multi-Function Monitor Step 6: Connect the free analog cable to the Nellcor analog output (10 Minutes) Step 7: Verify configuration with the user (15 Minutes) Ensure that the correct waves and numerics are being interfaced from the Nellcor monitor.
Radiometer Transcutaneous pO2/pCO2 Monitor Radiometer Transcutaneous pO /pCO Monitor This note provides information on configuring the TCM3 Free Analog device into the VueLink module. This procedure should take approximately 140 minutes and consists of 8 steps:- General information on configuring the free analog devices has not been included in this technical note.
Page 183
Radiometer Transcutaneous pO2/pCO2 Monitor Step 1: Ensure that all the necessary hardware is available (10 Minutes) You will need:- • An M1032A #A01 VueLink Module • 4m Free Analog (open end) cable M1032A #K6A or M1182A #K6A 2m Free Analog cable M1032A #K60 or M1182A #K60. •...
Page 184
Radiometer Transcutaneous pO2/pCO2 Monitor Step 2: Select two Numerics (10 Minutes) The following parameters are provided at the Analog Output port of TCM3: • Alarm (TCM3 pin 5) *Transducer Temperature (TCM3 pin 8) • Transcutaneous partial oxygen pressure (TCM3 pin 9) •...
Page 185
Radiometer Transcutaneous pO2/pCO2 Monitor VueLink Free Analog Cable Designation TCM3 PIN Channel Wire Colours Numeric 1 TCPO Numeric 2 Orange CPCO Alarm Yellow Alarm Pins used for the TCM3 free analog interface Free Analog Device Information...
Page 186
Radiometer Transcutaneous pO2/pCO2 Monitor Step 4: Enter a device name, device-id and channel labels (10 Minutes) All the TCM3 parameter labels need to be entered with the handheld keypad. The following parameter abbreviations are recommended as labels for these, but the customer can decide on his/her own abbreviations.
Page 187
Radiometer Transcutaneous pO2/pCO2 Monitor Value Range = Decimal Channel Label Unit Voltage Range Represent. Numeric 1 TCPO mmHg 0 to 200 mmHg = 0 to 1 V Numeric 2 CPCO mmHg 0 to 200 mmHg = 0 to 2 V 1.
Page 188
Radiometer Transcutaneous pO2/pCO2 Monitor VueLink only provides a visual alarm, not an audible alarm. Note Free Analog Device Information...