Page 1
Operating Instructions Pharmaceutical Refrigerator MPR-1412R MPR-1412 Series MPR-1412 Please read the operating instructions carefully before using this product, and keep the operating instructions for future use. See page 41 for all model numbers.
CONTENTS INTRODUCTION ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ3 PRECAUTIONS FOR SAFE OPERATION ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ4 ENVIRONMENTAL CONDITIONS ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ9 SYMBOLS ON UNIT ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ9 REFRIGERATOR COMPONENTS Main body ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 10 Control panel ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 12 Inside the front over ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 13 INSTALLATION SITE ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 14 INSTALLATION ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 15 START-UP PROCEDURE ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 17 PLACEMENT OF STORED ITEMS ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ...
INTRODUCTION Read the operating instructions carefully before using the appliance and follow the instructions for safe operation. Handling of Operating Instructions PHC Corporation never guarantees any safety if the appliance is used for any objects other than intended use or used by any procedures other than those mentioned in the operating instructions. ...
PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with the operating instructions as they contain important safety advice. Items and procedures are described so that you can use this unit correctly and safely. Following these precautions will prevent possible injury to the user and any other person.
Page 5
PRECAUTIONS FOR SAFE OPERATION WARNING Never damage the power supply cord or power supply plug (by breaking, adapting, placing near a source of heat, bending with force, twisting, pulling, adding weight, or binding). A damaged power supply cord or power supply plug may cause electric shock, short circuit, or fire. Remove dust from the power supply plug periodically.
Page 6
PRECAUTIONS FOR SAFE OPERATION WARNING Never store volatile or flammable substances in this unit except in a sealed container. Such substances may cause explosion or fire if they leak. Never insert metal objects such as pins and wires into any vent, gap, or outlet on the unit. This may cause electric shock or injury by accidental contact with moving parts.
Page 7
PRECAUTIONS FOR SAFE OPERATION WARNING Ask a qualified contractor to carry out disassembly and disposal of the unit. Leaving the unit in a location that can be accessed by third parties may result in unexpected accidents (e.g. the unit may be used for unintended purposes).
Page 8
PRECAUTIONS FOR SAFE OPERATION CAUTION This unit must be plugged into a dedicated circuit protected by branch circuit breaker. Use a dedicated power source as indicated on the rating label attached to the unit. multiple-tap may cause fire resulting from abnormal heating. Do not climb on top of the unit or put any objects on the unit.
ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1): Indoor use; Altitude up to 2000 m; Temperature 5 C to 40 Maximum relative humidity 80 % for temperature up to 31 C decreasing linearly to 50 % relative humidity at 40 ...
REFRIGERATOR COMPONENTS Main body MPR-1412 Back side...
Page 11
REFRIGERATOR COMPONENTS 1. Front cover: Open this cover when using the remote alarm terminal or replacing the glow starter [page 13]. To open the cover, pull the lower portion forward. 2. Control panel: The chamber temperature and other alarms/functions can be set through the keys on the control panel.
REFRIGERATOR COMPONENTS Control panel 1. Door check indicator (DOOR): The red LED 5. Scroll key ( lamp is lit when the door is opened. At “temperature display mode”; pressing this key 2 minutes after the door check indicator is lit, for more than 5 seconds leads “setting of lock the buzzer is activated to notice the door opening function”...
REFRIGERATOR COMPONENTS Inside the front cover 1. Condenser filter This is a filter for preventing for clogging of the condenser. The clogged filter may cause failure of refrigerating circuit. Clean the filter every 3 months [page 32]. 2. Remote alarm terminal The alarm is relayed to a remote location when a remote alarm device (commercially available) is connected to the remote alarm terminal.
INSTALLATION SITE This unit must be installed in a location which meets all the conditions described below. If the unit is installed in a location which does not meet the conditions, its specified performance may not be achieved or malfunctions and accidents may occur. ...
INSTALLATION 1. Preparations after unpacking Remove all the tape used to secure the doors and interior parts, and leave the doors open for a short while for ventilation. If any surfaces of the outer cabinet are dirty, wipe the surface using a cloth moistened with a diluted neutral dish-washing detergent.
Page 16
INSTALLATION 5. Setting up the shelves (MPR-1412) All shelves are packaged at the bottom of the chamber. Set the shelves firmly in place on the shelf stoppers at the standard locations (4 locations) below. 1) The sixth hole under the second screw from the top (top screw) [Fig. 3] 2) The first hole under the fifth screw from the top.
START-UP PROCEDURE Follow this procedure for the initial operation of the unit and for consequent operations (after temporary stoppage for cleaning, maintenance or moving). After a power failure, the unit will restart operation automatically with the same settings as before the power failure [page 26].
PLACEMENT OF STORED ITEMS Do not place stored items so as to disrupt the air circulation in the chamber. Disruption of air circulation can cause the freezing of stored items or can impair the temperature distribution in the chamber. Air intake vent Do not block the air intake vent with the stored items.
SETTING THE CHAMBER TEMPERATURE Set the chamber temperature according to the condition of use. This refrigerator can keep the stored items for long period under appropriate temperature. Setting range of chamber temperature: between 2 C and 23 Initial setting (factory setting): 5 <Important>...
LOCK OF CHAMBER TEMPERATURE SETTING The chamber temperature setting can be locked to avoid accidental changes. When the lock is ON, a change of chamber temperature setting is not accepted even if the key on the control panel is operated. ...
SETTING THE TEMPERATURE DISPLAY MODE The display mode of chamber temperature is selectable; integer display or decimal display. Select the display mode as required. Setting range of temperature display mode: integer mode (display: -50 to 50) or decimal display (display:-19.9 to 50.0) ...
DEFROSTING THE EVAPORATOR The following 2 types of defrosting methods is provided for the refrigerator. Both of them are controlled automatically. Cycle defrosting The refrigerator temperature is kept stable by ON/OFF operation of the compressor. When the compressor is OFF (stopped), the frost on the evaporator is melted by the heater. This defrosting never influences to the refrigerator temperature.
SETTING THE ALARM TEMPERATURE Setting the high-temperature alarm An abnormality (chamber temperature rise) is indicated by the blinking of red LED lamp and the chamber temperature display and the buzzer sounding (15 minutes after blinking) if the chamber temperature exceeds the set value of the high-temperature alarm. Always set the high-temperature alarm to protect stored items from damage resulting from an increase in temperature.
ALARM TEMPERATURE SETTING Setting the low-temperature alarm An abnormality (chamber temperature drop) is indicated by the blinking of red LED lamp and the chamber temperature display and the buzzer sounding (15 minutes after blinking) if the chamber temperature falls below the set value of low-temperature alarm. Always set the low-temperature alarm to protect stored items from damage resulting from a drop in temperature.
SETTING THE DELAY OF DOOR ALARM The door check indicator is lit when the door is opened, and the buzzer sounds with some delay to notice the door opening. The delay time (between lighting of the door check indicator and activation of the buzzer) can be changed. Set an appropriate delay time according to the condition of use to prevent the rise of chamber temperature occurring as a result of the door being left open for too long.
SETTING THE BUZZER SUSPENDED PERIOD An abnormality is indicated again by the buzzer sounding after a certain period has elapsed (the buzzer recovery time) if the alarm mode continues even after the buzzer has already been silenced by pressing the buzzer stop key (BUZZER). Always set the “buzzer suspended period” to avoid incorrect recognition of alarm mode.
REMOTE ALARM TERMINAL The alarm status is noticed to a remote location when a remote alarm equipment (commercial item) is connected to the remote alarm terminal. It is recommended to install a remote alarm equipment (commercial item) when the refrigerator is installed in a desolate location so that an alarm status is noticed to an operator.
ALARM FUNCTIONS This unit has the alarms functions listed below. Contact our sales representative or agent if the alarm is not resolved by eliminating the following causes. Before calling for service, take some precautions to prevent the stored items (for example by transferring the items to another freezer).
SELF-DIAGNOSTIC FUNCTIONS Alarms involving an error code indicated on the temperature display are generated by the unit’s self-diagnostic function. If any of the error codes are displayed, contact our sales representative or agent without delay, report the error code, and request repairs. Self-diagnostic Cause Indication...
ROUTINE MAINTENANCE Cleaning of exterior, interior, and accessories Use a dry cloth to wipe down the outside and inside of the unit and all accessories. If the outside panels are dirty, clean them with a diluted neutral dish-washing detergent. Wipe off condensation from the exterior of the cabinet or glass with a dry, soft cloth. Using an undiluted solution of detergent may cause the unit’s plastic areas to crack.
ROUTINE MAINTENANCE Replacement of fluorescent lamp Follow the procedure below when replacing a fluorescent lamp. The fluorescent lamp is located vertically at the back of the center mullion. It is recommended to replace a glow starter when replacing a fluorescent lamp 1.
ROUTINE MAINTENANCE Replacement of glow starter A glow starter is located inside the front cover [page 13]. Glow starter (with water-proof cover) 1. Turn off the light switch and disconnect the power supply plug of the refrigerator. 2. Remove the water-proof cover [Fig. 1] and replace the glow starter.
TROUBLESHOOTING If the unit malfunctions, check out the following before calling for service. <Attention> If the malfunction is not resolved after checking the following items or if the malfunction is not shown in the table below, contact our sales representative or agent. Malfunction Check/Remedy Nothing operates even...
DISPOSAL OF UNIT WARNING If the unit is to be stored unused in an unsupervised area for an extended period ensure that children do not have access and doors cannot be closed completely. The disposal of the unit should be accomplished by appropriate personnel. Always remove doors to prevent accidents such as suffocation.
Page 35
DISPOSAL OF UNIT Information on Disposal for Users of Waste Electrical & Electronic Указания за изхвърляне на електрически и електронни отпадъци (за Equipment (private households) битови потребители) This symbol on the products and/or accompanying documents Този символ върху изделия и-или в придружаваща means that used electrical and electronic products should not документация...
Page 36
DISPOSAL OF UNIT Oplysninger til brugerne om afhændelse af elektriske apparater og Informaţii pentru utilizatori, privind eliminarea echipamentelor electrice şi elektronisk udstyr (private husholdninger) electronice uzate (proprietate particulară) Når produkter og/eller medfølgende dokumenter indeholder Acest simbol, de pe produse şi/sau documentele însoţitoare, dette symbol, betyder det, at elektriske apparater og elektronisk are semnificaţia că...
Page 37
DISPOSAL OF UNIT Informace pro uživatele k likvidaci elektrických a elektronických zařízení Informacija naudotojams dėl nenaudojamos elektros ir elektroninės (domácnosti) įrangos utilizacijos (naudojantiems ją buityje) Tento symbol na produktech anebo v průvodních dokumentech Šis simbolis, pažymėtas ant gaminių ir / arba lydimųjų znamená, že použité...
Page 38
DISPOSAL OF UNIT Informacije za uporabnike o odstranitvi odpadne električne in elektronske Tietoja sähkö- ja elektroniikkalaitteiden hävittämisestä (kotitaloudet) opreme (zasebna gospodinjstva) Tämä symboli tuotteissa ja/tai niiden käyttöohjeissa osoittaa, Ta simbol, ki se nahaja na izdelkih in/ali priloženih dokumentih, että käytettyjä sähkö- ja elektroniikkalaitteita ei saa hävittää pomeni, da se rabljeni električni in elektronski izdelki ne smejo tavallisen kotitalousjätteen mukana.
TEMPERATURE RECORDER (OPTION) The chamber temperature can be recorded and checked by installing an optional temperature recorder (MTR-G04 or MTR-0621LH). Contact our sales representative or agent for the purchase of temperature recorder. Main specifications of temperature recorder MTR-G04 MTR-0621LH Application Refrigerator &...
SPECIFICATIONS Product name Pharmaceutical Refrigerator MPR-1412 MPR-1412R External dimensions W1440 mm x D830 mm + (90) mm x H1950 mm Internal dimensions W1320 mm x D710 mm x H1500 mm Effective capacity 1364 L 1359 L Exterior Painted steel Interior Painted steel Door Painted steel, Provided with glass window...
PERFORMANCE Product name Pharmaceutical Refrigerator MPR-1412 Model number MPR-1412-PA MPR-1412-PR MPR-1412-PK MPR-1412-PE Temperature control range C to 14 C (ambient temperature; -5 C to 0 C, no load) C to 23 C (ambient temperature; 0 C to 35 C, no load) Noise level 48 dB (A scale) Maximum pressure...
CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety check sheet 1. Unit contents: □ □ Risk of infection: □ □ Risk of toxicity: □ □...