Page 3
This product is not designed for, nor intended for use in hazardous duty areas as defined by ATEX or the NEC (National Electrical Code); including, but not limited to use with flammable liquids. Masterflex L/S Precision Variable-Speed Modular Drives Operating Manual...
Page 4
CAUTION: Replace the fuse only with one of the same type and rating. The fuse rating and type are stated on the rear panel. 7. The L/S Precision Variable-Speed Modular Drives should not be used in outdoor or hazardous locations. L/S Precision Variable-Speed Modular Drives Operating Manual Masterflex...
General Description The L/S Precision Variable-Speed Modular Drives control the speed of MASTERFLEX® Pump Heads to provide flow rates from 0.06 to 3400 mL/min. The 600 rpm L/S Precision Variable-Speed Modular Drives can mount up to two (2) MASTERFLEX Pump Heads or other pumps adapted to MASTERFLEX drives.
DB9 connector (Round 18 pin for NEMA 4X) connector. G. IEC Power Entry Module/Line Cord: Utilized to connect line cord to drive. See page 4-1 for alternative cords. H. Motor Connector: Utilized to connect motor to controller. L/S Precision Variable-Speed Modular Drives Operating Manual Masterflex...
Mounting the Mount Pump Head and load tubing (See Pump Head datasheets within this flash drive or on the web). Check to ensure that rollers are clean and Pump Head free of defects. Masterflex L/S Precision Variable-Speed Modular Drives Operating Manual...
20 minutes for more stable flow conditions. NOTE: When in INTernal mode, pump will start automatically after brownout, powerout or if drive was turned off using the power switch while pump was running. Masterflex L/S Precision Variable-Speed Modular Drives Operating Manual...
4 (green) CW/CCW 6 (blue) Chassis (Earth) Ground 9 (violet) NOTE: Colors for NEMA 4X are those of the Remote Cable, Cat. number MN-77300-32. Figure 3-1. DB9 Pin Configuration with Wiring Scheme L/S Precision Variable-Speed Modular Drives Operating Manual Masterflex...
AC line to the drive to start and stop the motor can cause drive hardware damage. The drive is designed to use control input signals that will start and stop the motor. Masterflex L/S Precision Variable-Speed Modular Drives Operating Manual...
Transmitter Type ISA Type 2 (Two-wire) Transmitter Receiver Power Supply ISA Type 3 (Three-wire) Transmitter Receiver Power Supply ISA Type 4 (Four-wire) Transmitter Receiver Power Supply Figure 3-2. Typical 4-20 mA wiring L/S Precision Variable-Speed Modular Drives Operating Manual Masterflex...
Line Cord (115V), United States 50001-68 Line Cord, Israel 50001-69 Line Cord, Europe 50001-70 Line Cord, England 50001-72 Line Cord, Switzerland 50001-74 Line Cord, Italy 50001-76 Line Cord (230V), United States 50001-78 Masterflex L/S Precision Variable-Speed Modular Drives Operating Manual...
2. Disconnect the AC power input line cord from the receptacle. 3. Remove and check the fuse and replace if defective. Figure 4-1. Fuse Replacement Item Description IEC Power Entry Module / Line Cord 3.15A (5×20 mm) Fuse – Do Not Substitute L/S Precision Variable-Speed Modular Drives Operating Manual Masterflex...
Figure 4-2. Motor Gear and Brush Replacement Cleaning Remove power from the pump before any cleaning WARNING: operation is started. Keep the drive enclosure clean with mild detergents. Do not immerse or use excessive fluid when cleaning. Masterflex L/S Precision Variable-Speed Modular Drives Operating Manual...
Unit will turn on but display would Verify that the incoming voltage meets the required dim and pump will not spin minimum of 90Vrms. Unit vibrates excessively when Check that the tubing was loaded properly. pump is running Masterflex L/S Precision Variable-Speed Modular Drives Operating Manual...
Less than 6562 ft (2000 m) Pollution Degree: 07557-00, 07557-10 Pollution degree 2 (indoor use–lab, office) 07557-60, 07557-70 Pollution degree 3 (Indoor use-sheltered locations) Noise Level: <70 dBA @ 1 meter with Pump Head Masterflex L/S Precision Variable-Speed Modular Drives Operating Manual...
Page 26
CAN/CSA-C22.2 No. 61010-1, second edition, including Amendment 1, or a later version of the same standard incorporating the same level of testing requirements. (For CE Mark): EN61010-1: (EU Low Voltage Directive) and EN61326: (EU EMC Directive) L/S Precision Variable-Speed Modular Drives Operating Manual Masterflex...
Warranty, Product Return and Section 8 Technical Assistance Warranty Use only MASTERFLEX precision tubing with MASTERFLEX pumps to ensure optimum performance. Use of other tubing may void applicable warranties. This product is warranted against defects in material or workmanship, and...
Any damages resulting from improper packaging are your responsibility. Technical Assistance If you have any questions about the use of this product, contact the Manufacturer or authorized seller. L/S Precision Variable-Speed Modular Drives Operating Manual Masterflex...
Page 30
® ® US & Canada only *EN809 manufactured by: Masterflex LLC. Toll Free 1-800-MASTERFLEX | 1-800-637-3739 28092 W Commercial Avenue, Barrington, IL 60010 Outside US & Canada masterflex.tech@avantorsciences.com 1-847-381-7050 www.avantorsciences.com\masterflex...
Need help?
Do you have a question about the L/S MFLX07557-00 and is the answer not in the manual?
Questions and answers