User Manual
G3 B30VT
Luna G3 BPAP 30VT

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 3B G3 B30VT and is the answer not in the manual?

Questions and answers

Summary of Contents for React Health 3B G3 B30VT

  • Page 1 User Manual G3 B30VT Luna G3 BPAP 30VT...
  • Page 2: Table Of Contents

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Table of Contents 1. Symbols·············································································· 1 1.1 Control Buttons ····························································· 1 1.2 Device Symbols ····························································· 1 2. Warning, Caution and Important Tip ··········································· 3 3. Intended Use ······································································· 3 4. Contraindications ··································································· 4 5.
  • Page 3 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 15.2 Connecting to WiFi Network ··········································· 27 16. Navigating the Patient Menu··················································· 31 16.1 Steps to Navigating the Patient Menu ································ 31 16.1.1 Accessing the Main Interface ································· 31 16.1.2 Bringing up the Initial Setup Interface ······················ 32 16.1.3 Selecting Options ···············································...
  • Page 4: Symbols

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 1. Symbols 1.1 Control Buttons Home Button Start / Stop Button Knob 1.2 Device Symbols Follow Instructions for Use Operating Instructions Type BF Applied Part (mask) Class II (Double Insulated) For indoor use only AC Power DC Power IP22...
  • Page 5 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Do not use the product if the package is damaged Disassembly is prohibited Maximum water level European CE Declaration of Conformity Product is intended for use by a single patient only Lot number Non-Ionizing Radiation SD Card...
  • Page 6: Warning, Caution And Important Tip

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 2. Warning, Caution and Important Tip WARNING! Indicate the possibility of injury to the user or operator. CAUTION! Indicate the possibility of damage to the device. IMPORTANT TIP! Place emphasis on an operating characteristic. Warnings, Cautions, and Important Tips appear throughout this manual as they apply.
  • Page 7: Contraindications

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 • Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security systems (metal detectors) as it may cause unacceptable risk to the patient or damage to the device. Some electromagnetic sources may not be apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and discontinue use.
  • Page 8: Specifications

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 IMPORTANT! • An irregular sleep schedule, alcohol consumption, obesity, sleeping pills, or sedatives may aggravate your symptoms. • Please use a mask which meets ISO 17510: 2015. 5. Specifications Device Size Dimensions (L x W x H): 265 mm ×...
  • Page 9 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Main device input 24 V, 3.33 A Device offer to Heated Tubing Communications Port 24 V 18 W Type of Protection against Electric Shock Class II Equipment Degree of Protection against Electric Shock Type BF Applied Part Degree of Protection against Ingress of Water IP22...
  • Page 10 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Maximum Flow Test Pressures (cmH20) Measured Pressure at the Patient Connection Port (cmH20) Average Flow at the Patient Connection Port (L/min) Maximal Optical Output Power Less than 1.5 mw maximum average. Air Tubing Air tubing Length...
  • Page 11: Available Therapies

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 6. Available Therapies The device delivers the following therapies: CPAP – Delivers Continuous Positive Airway Pressure; CPAP maintains a constant level of pressure throughout the breathing cycle. If your health care professional has prescribed ramp for you, you can turn the Knob to reduce the pressure and then gradually increase the pressure to the therapeutic pressure setting so that you can fall asleep more...
  • Page 12: Glossary

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 7. Glossary Apnea A condition marked by the cessation of spontaneous breathing. Auto Off When this feature is enabled, the device automatically discontinues therapy whenever the mask is removed. Auto On With this feature, the device automatically initiates therapy when you breathe into the mask.
  • Page 13 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Liters Per Minute. Obstructive Sleep Apnea. Patient Menu The display mode in which you can change patient-adjustable device settings, such as the starting pressure for the Ramp feature. Ramp A feature that may increase patient comfort when therapy is started. It can reduce pressure and then gradually increase the pressure to the prescription setting so the patient can fall asleep more comfortably.
  • Page 14: Model

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 8. Model Product Contents Maximum Work Model Work Mode Pressure Main Device Optional Accessory (cmH20) Mask, WiFi kit, Cellular Module, Main device CPAP, S, T, Heated Tubing, G3 B30VT (3.5-inch TFT) PM2.5 Filter 11 / 58...
  • Page 15: Package Contents

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 9. Package Contents After unpacking the system, make sure you have everything shown here: Articles Qty. Notes Device Air Filter Power Adapter Power Cord Mask Optional PM2.5 Filter Optional WiFi kit Optional Cellular Module Optional...
  • Page 16: System Features

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 IMPORTANT! • If any of the above parts are missing, contact your home care provider. • Contact your home care provider for additional information on the available accessories of this device. When using optional accessories, always follow the instructions enclosed with the accessories.
  • Page 17 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Fig. 10-2 Name Function Air Outlet Deliver pressurized air; connects to the tubing Heated Tubing Port Connected to the plug of the heated tubing DC Inlet An inlet for the DC power supply Place the cap on the air filter, which is used to filter dust and pollen Filter Cap (Air Inlet) in the air entering the device...
  • Page 18: First Time Setup

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 11. First Time Setup 11.1 Placing the Device Place the device on a firm, flat surface. WARNINGS! • If the device has been dropped or mishandled, if the enclosure is broken, or if water has entered the enclosure, disconnect the power cord and discontinue use.
  • Page 19: Connecting To Power

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Fig. 11-2 (3) Change the air filter and filter cap to the PM2.5 filter, as shown in Fig. 11-3. PM2.5 Filter Fig. 11-3 CAUTIONS! • The air filter or the PM2.5 filter must be in place when the device is operating. •...
  • Page 20: Connecting To Power Cord Locker

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 WARNINGS! • The device is powered on for use when the power cord and power adapter is connected. The Knob turns the blower On / Off. • Use of the device at an AC voltage beyond the stated range (see Section 5 “AC Power Consumption”) may damage the device or cause device failure.
  • Page 21: Assembling The Tubing / Heated Tubing And Mask

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 11.5 Assembling the Tubing / Heated Tubing and Mask (1) Connect one end of the tubing to the air outlet of the device, as shown in Fig. 11-7. Air Outlet Tubing Fig.
  • Page 22: Using Oxygen With The Device

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 There are five heat levels available, and the number of heat level will appear in the Main Interface on the screen of the device. The number 3 next to the icon indicating the heat is adjusted to Level 3, as shown in Fig.
  • Page 23: Inserting The Sd Card

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 • Turn on the device before turning on the oxygen. Turn off the oxygen before turning off the device. Explanation of Warning: When the device is turned off, but the oxygen flow still exists, oxygen may accumulate within the device's enclosure and pose a fire hazard.
  • Page 24: Starting Treatment

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 11.8 Starting Treatment Connect the device to a power outlet, press the Start / Stop Button and the device will start delivering air. WARNINGS! • Be sure to follow your physician’s instructions on adjusting the settings! To order any accessories not included with this device, contact your equipment supplier.
  • Page 25: Using The Ramp Feature

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 12.5 Using the Ramp Feature Every time the feature is enabled, the pressure will drop to the initial pressure, and then gradually rise to the prescribed treatment pressure according to the preset ramp time, so as to make the patient fall asleep easily.
  • Page 26: Heated Humidifier

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 13. Heated Humidifier The humidifier is available from your home care provider. The humidifier may reduce nasal dryness and irritation by adding moisture (and heat if applicable) to the airflow. 13.1 Filling the Water Chamber 13.1.1 Removing the Water Chamber Press down the water chamber, and then remove it, as shown in Fig.
  • Page 27: Returning The Water Chamber

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 (2) Open the cap, and fill the water chamber with approximately 360 ml of water, as shown in Fig. 13-4. Make sure that the water does not exceed the maximum water level line. Fig.
  • Page 28: Emptying The Water Chamber

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 WARNING! • For safety purposes, the device must be placed on a flat surface at a level lower than the patient's head on a bed, so that the condensation flows back to the water chamber rather than remain in the tubing causing rainout.
  • Page 29: Using The Cellular Module And The Wifi Kit

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 WARNING! • Do not touch the heater plate of the device when it is working, otherwise you may get burned. Turn off the heat when the heated humidifier is not in use. CAUTIONS! •...
  • Page 30: Connecting To Wifi Network

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 (1) When the signal is weak, data transmission may become slow and even stop. (2) The Cellular Module will keep searching for signals until one is found. If the signal is strong, the signal icon appears in the Main Screen, as shown in Fig. 15-2 (the signal icons of different strength appear in a similar way).
  • Page 31 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Fig. 15-4 (2) Turn the Knob until the cursor stays on the “WiFi” option, as shown in Fig. 15-5. Press the Knob , and the interface shown in Fig. 15-6 appears. Wait for 0-5 seconds to automatically access the “WiFi”...
  • Page 32 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Fig. 15-7 Fig. 15-8 Note: are page turning symbols. If no WiFi networks are found, the “WiFi”setup interface displays “No Available WiFi”, as shown in Fig. 15-9. Fig. 15-9 (4) After the desired WiFi network is found, press the Knob , turn the Knob select this WiFi network.
  • Page 33 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Fig. 15-10 Fig. 15-11 If the WiFi network is connected successfully, the screen will return to the “WiFi” setup interface, displaying “connected” to the right of the WiFi network name, and the WiFi symbol will become blue, as shown in Fig.
  • Page 34: Navigating The Patient Menu

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 If the desired WiFi network is a public network that does not require a password, turn the Knob directly after accessing the password input interface until the cursor stays on the Confirmation Key .
  • Page 35: Bringing Up The Initial Setup Interface

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Fig. 16-4 The first icon on the upper part of the screen indicates the Preheat Function Icon, the second icon indicates the Accessorie, and the third icon indicates Mask Setup Icon, the fourth icon indicates the Report Interface Icon, the fifth icon...
  • Page 36: Selecting Options

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 16.1.3 Selecting Options As you turn the Knob clockwise, the cursor moves downwards from one option to another. When the cursor is on a certain option, press the Knob , and the color of the option is changed, meaning that the option can now be adjusted, as shown by the Humidifier option in Fig.
  • Page 37: Turning Pages

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 16.1.6 Turning Pages When the cursor is on Work screen saver, the last option shown in Fig. 16-9, the remaining options will appear on a new page if you continue to turn the Knob clockwise, as shown in Fig.
  • Page 38: Options Of The Patient Menu And Corresponding Descriptions

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 16.2 Options of the Patient Menu and Corresponding Descriptions Option Range Description There are six humidity levels available. As the Off, Auto, Humidifier numbering increases, the humidity rises accordingly. 1 ~ 5 “Off”...
  • Page 39 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 12-hour / Turn the Knob to choose between the two Time Format 24-hour time formats. Brightness High / Low Setting screen brightness by adjusting this option. The backlight of the LCD screen can be set to “Auto” or “On.”...
  • Page 40: Alarm

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 17. Alarm This chapter describes device alarms and the responses operators make to different alarms. After running, disconnect the device from the power supply by unplugging the power cord, an audible alarm sounds like “beep beep beep, beep-beep, beep beep beep, beep-beep”, which means the alarming system of the device works normally.
  • Page 41: Auditory Alarming

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 17.3 Auditory Alarming In the case of alarming, the alarming sounds at different grades will occur and are described as follows: Grade Auditory Description beep beep beep beep-beep High beep beep beep beep beep Intermediate beep beep beep...
  • Page 42: Alarming Information And Description

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 17.5 Alarming Information and Description Alarm Alarm Alarm Message Description Priority Type An audible alarm will sound in 6s if the device is accidentally disconnected from power when it is delivering air. Alarming duration time is no less than 30s.
  • Page 43 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 When the airflow is on, an audible alarm will sound if the respiratory rate is below the limen; the screen will display “Low RR!!!”. Low RR!!! Function High Setting range: Off, 4 ~ 40 BPM, in 1 Priority Alarm BPM increments, the default setting is “6...
  • Page 44 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Priority Alarm replacement time reaches but without replacing the air filter; the screen will display “Please Change Filter!”. The default setting is “Off”. When the tubing Alarm feature is enabled, an audible alarm will sound if Technology Please Change the preset replacement time reaches but...
  • Page 45: Reposition Of Alarming

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 17.6 Reposition of Alarming After the elimination of the alarming faults, the residual alarming information still exists (alarming information is shown on the top of the screen without any visual and auditory alarming), and turn the button leftwards or rightwards to reduce the residual alarming information.
  • Page 46: Cleaning And Disinfection

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 (2) Press the home button and it will become silent for 100 to 120 seconds. If the alarm state has not been eliminated, and then the alarm sounds again immediately after the end of the silence.
  • Page 47: Cleaning The Mask And Headgear

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 • Do not use solutions containing chlorinated lime, chlorine, or aromatic to clean the device and its accessories. Liquid soap containing moisturizing agents or antimicrobials should not be used either. These solutions may harden cleaned materials or reduce their lifespan. •...
  • Page 48: Cleaning The Transfer Box

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 18.3 Cleaning the Transfer Box (1) Removing the Transfer Box: First remove the water chamber from the device, and then remove the transfer box, as shown in Fig. 18-2. Transfer Box Fig.
  • Page 49: Replacing The Air Filter / Pm2.5 Filter

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 CAUTION! • It is recommended to clean the tubing once a week. 18.6 Replacing the Air Filter / PM2.5 Filter Attach the air filter to the filter cap, as shown in Fig. 18-4. Air Filter Filter Cap Fig.
  • Page 50: Disinfection

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 18.7 Disinfection If you have followed the above cleaning instructions, you do not have to disinfect the device and / or water chamber. CAUTIONS! • Disinfectants tend to damage materials and reduce the life of components. Try to select the appropriate disinfectant, and follow the disinfectant manufacturer's instructions and recommendations.
  • Page 51: Traveling With The Device

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 19. Traveling with the Device CAUTIONS! • Empty the water chamber before packing the device for your trip; in order to prevent any remaining water from entering the device. • Using the device at an incorrect elevation setting could result in airflow pressures higher than the prescribed setting.
  • Page 52: Transferring The Device To Another Patient

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 20. Transferring the Device to Another Patient If the device is transferred to another patient, components in close contact with the previous owner, including the mask, headgear, tubing, and air filter, should be replaced to prevent cross-infection.
  • Page 53: Troubleshooting

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 24. Troubleshooting The table below lists common problems you may have with the device and possible solutions to those problems. If none of the corrective actions solve the problem, contact your home care provider.
  • Page 54 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 part of the tubing should be lower than the patient's head Nasal, sinus, or Sinus middle Contact your physician ear pain inflammation immediately the treatment pressure is Discomfort due to determined according to the inability to adapt patient's conditions,...
  • Page 55: Common Problems In The Device And Corresponding Solutions

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 24.2 Common Problems in the Device and Corresponding Solutions Problem Possible Cause Solution(s) Take a few deep breaths with the The Auto On / Off feature is mask on, and the device will start enabled automatically Ensure that the power cord, power...
  • Page 56: Emc Requirements

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 25. EMC Requirements Guidance and manufacturer's declaration - electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The user of the device should ensure that it is used in such an environment. Electromagnetic Environment –...
  • Page 57 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment. 60601 Compliance Electromagnetic...
  • Page 58 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment. Immunity IEC 60601 Compliance...
  • Page 59 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Recommended separation distances between portable and mobile RF communications equipment and the device The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
  • Page 60 Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 Recommended separation distances between wireless communications equipment The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between RF wireless communications equipment and the device as recommended below, according to the maximum output power of the communications equipment.
  • Page 61: Limited Warranty

    Luna G3 BPAP 30VT (G3 B30VT) User Manual MID-99/v1.0 26. Limited Warranty 3B Medical, Inc. warrants that the device shall be free from defects of workmanship and materials and will perform in accordance with the product specifications for a period of two (2) years for main unit and three (3) months for all accessories from the date of sale by 3B Medical, Inc.