Advertisement

Quick Links

FUSION 4000-X OEM
USER MANUAL
INTELLECTUAL PROPERTY
All Intellectual Properties, as defined below, owned by or which is otherwise the property of Chemyx Inc. or its suppliers relating to the Chemyx syringe pumps, including but not limited to, accessories,
parts or software relating thereto (Chemyx Syringe Pumps), are proprietary to federal and state laws, and international treaty provisions. Intellectual Property includes but is not limited to, inventions
(patentable or unpatentable), patents, trade secrets, copyrights, software, firmware, computer programs, and related documentation and other works of authorship. Moreover, you agree that you will
not, and will not attempt to, modify, prepare derivative works of, reverse engineer, disassemble the Chemyx syringe pumps, decompile or otherwise attempt to create source code from the related
software/firmware. No title to or ownership in Intellectual Property is transferred to you. All applicable rights of the Intellectual Property shall remain with Chemyx and its suppliers.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FUSION 4000-X OEM and is the answer not in the manual?

Questions and answers

Summary of Contents for Chemyx FUSION 4000-X OEM

  • Page 1 INTELLECTUAL PROPERTY All Intellectual Properties, as defined below, owned by or which is otherwise the property of Chemyx Inc. or its suppliers relating to the Chemyx syringe pumps, including but not limited to, accessories, parts or software relating thereto (Chemyx Syringe Pumps), are proprietary to federal and state laws, and international treaty provisions. Intellectual Property includes but is not limited to, inventions (patentable or unpatentable), patents, trade secrets, copyrights, software, firmware, computer programs, and related documentation and other works of authorship.
  • Page 2: Table Of Contents

    Table of Contents Safety Information Product Overview Data Sheet Loading a Syringe Pump Control by Computer Controller Programs Commands Pump Maintenance Syringe Pump Accessories Warranty and Repair Information Appendix A: Syringe Volume/Diameter Reference Table...
  • Page 3: Safety Information

    CHEMYX PRODUCTS ARE NOT APPROVED FOR CLINICAL USE ON HUMANS. Use Proper Power Supply Chemyx is not responsible for the use of power supplies outside the stated electrical specifications or failure to switch the power converter from 220V to 110 V while in the 220V environment or vice versa.
  • Page 4: Product Overview

    Product Overview Fusion OEM 4000-X The Fusion OEM 4000-X syringe pump comes with a dual drive system that allows two independent pumping channels to be controlled for simultaneous infusion and withdrawal at same/different flows. Fusion OEM 4000-X 7.5 in 7.5 in 12 in 4 in...
  • Page 5: Data Sheet

    Fusion 4000-X OEM Module Data Sheet The Fusion 4000-x OEM syringe pump comes with a dual drive system that allows two independent pumping channels to be controlled for simultaneous infusion and withdrawal at same/different flows. TECHNICAL SPECIFICATIONS Infuse & Withdraw Mode ±...
  • Page 6: Loading A Syringe

    Motor Power The Fusion 4000-X OEM pump motor can generate up to a maximum of 80lbs linear force. Some syringes made of more fragile materials may not be able to withstand the full linear force of the pump.
  • Page 7: Pump Control By Computer

    The OEM syringe pumps can be controlled with an external computer using an RS232 serial connection. Cable Requirements Two types of connections are available for external computer control: DB9 and USB, cables for each are available for purchase through Chemyx (call 1-281-277-5499 or visit www.chemyx.com/contact-us). USB-A to USB-B cable DB9 Serial Cable (Male-to-Female) TTL Connection For the DB9 serial connection, a Male-to-Female DB9 cable must be used.
  • Page 8 Fusion OEM 4000-X Cable Connectors Power RS232 Foot switch Hit foot switch once, both pumps will start; hit twice, both pumps will stop Fusion OEM 4000-X Computer Connections...
  • Page 9: Controller Programs

    Chemyx Pump Controller Programs In addition to the manual control of the pumps through a terminal, Chemyx provides pump controller programs as an unsupported open-source applications written in Python, a LabVIEW package, and a MATLAB package. More information on these tools can be found on the website at www.chemyx.com/support.
  • Page 10: Commands

    Quick Start Guide - Commands IMPORTANT RS232 Protocol/Commands * Explanation of command function is located after command name* ALWAYS connect RS232 and RS232 Comport Settings: All system from 11/1/2010 Power supply BEFORE powering Bits per seconds - 38400 Data bits - 8 on the OEM Module! Parity - None Stop bit - 1...
  • Page 11: Pump Maintenance

    Pump lubrication should be performed every three months. Approved Lubricants • Engine/motor oil • Silicone oil • Machine oil • Tooling oil or “WD-40” • Lithium-based grease Please contact Chemyx call +1 281-277-5499 or email us at info@chemyx.com to confirm the safe use of a different lubricant.
  • Page 12: Syringe Pump Accessories

    Syringe Pump Accesories Streamline laboratory operations with this electric foot switch control that lets you start, stop, or pause your Chemyx syringe pump systems—even when your hands are full. The electric foot pedal switch connects seamlessly to computerized pumping programs, and its 10-...
  • Page 13: Warranty And Repair Information

    Warranty Information Chemyx provides a two-year limited warranty (1) first year for parts and labor (2) second-year for parts only from the shipment date for its pumps against defects in materials and workmanship. Chemyx will repair product that proves defective during its stated warranty period.
  • Page 14: Appendix A: Syringe Volume/Diameter Reference Table

    Appendices Appendix A: Syringe Volume/Inner Diameter Reference Chemyx Stainless Steel Hamilton Glass Size Diameter Inner | Outer Size Diameter 6.00 mL 9.5 mm inner | 13mm outer 0.50 μL 0.10 mm 20.00 mL 19.1 mm inner | 23 mm outer 1.00 μL...
  • Page 15 JMC Air-Tite Plastic Norm-Ject Plastic Size Diameter Size Diameter 1.00 mL 4.66 mm 1.00 mL 4.70 mm 2.00 mL 6.90 mm 2.50 mL 9.70 mm 2.50 mL 9.10 mm 5.00 mL 12.48 mm 5.00 mL 12.62 mm 10.00 mL 15.89 mm 10.00 mL 14.34 mm 20.00 mL...
  • Page 16 SGE Glass Terumo Japan Size Diameter Size Diameter 25.00 μL 0.73 mm 1.00 mL 4.73 mm 50.00 μL 1.03 mm 1.00 mL 6.50 mm 100.00 μL 1.46 mm 3.00 mL 8.65 mm 250.00 μL 2.30 mm 5.00 mL 13.00 mm 500.00 μL 3.26 mm 10.00 mL...
  • Page 17 Unimetrics Size Diameter 10.00 μL 0.46 mm 25.00 μL 0.73 mm 50.00 μL 1.03 mm 100.00 μL 1.46 mm 250.00 μL 2.30 mm 500.00 μL 3.26 mm 1000.00 μL 4.61 mm Kendall Monoject Plastic Size Diameter 1.00 mL 4.65 mm 3.00 mL 8.94 mm 6.00 mL...