9 Supplement
9
Supplement
9.1
Technical data
Note for approved instruments
The technical data in the respective safety instructions are valid for approved instruments (e.g. with
Ex approval). In some cases, these data can differ from the data listed herein.
General data
Series
Weight
Housing material
Connection terminals
Ʋ Type of terminal
Ʋ Wire cross-section
Voltage supply
Operating voltage
Ʋ Nominal voltage AC
Ʋ Nominal voltage DC
Max. power consumption
Sensor input
Quantity
Input type
Measured value transmission
Switching threshold
Ʋ On
Ʋ Off
Ʋ Tolerance
Current limitation
Terminal voltage
Internal resistance
Detection line break
Detection shortcircuit
Relay output
Quantity
Contact
Contact material
Switching voltage
Switching current
22
Module unit for mounting on carrier rails 35 x 7.5 acc. to
EN 50022/60715
125 g (4.02 oz)
Polycarbonate PC-FR
Screw terminal
0.25 mm² (AWG 23) ... 2.5 mm² (AWG 12)
24 ... 230 V (-15 %, +10 %) 50/60 Hz
24 ... 65 V DC (-15 %, +10 %)
3 W (8 VA)
2 x analogue
Active (sensor power supply by VEGATOR 122)
Analogue 8/16 mA
12.1 mA
11.9 mA
± 500 µA
23 mA (permanently short-circuit proof)
18.2 V DC, ± 5 %
200 Ω, ± 1 %
≤ 3.6 mA
≥ 21 mA
2 x operating relay
Floating spdt
AgSnO2, hard gold-plated
min. 10 mV DC, max. 253 V AC/50 V DC
min. 10 µA DC, max. 3 A AC, 1 A DC
VEGATOR 122 •