Factory Calibration Reset - Tintometer Lovibond MD110 Instruction Manual

Photometer-system
Hide thumbs Also See for Lovibond MD110:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 16
Calibration Mode
GB
Store
Date
Cal
RESULT
RESULT
Time
Cal
Zero
Mode
Test
On
Off
SEL
CAL
SEL
cAL
Mode
SEL
CAL
On
Off
MD110_3b 12/2020

Factory calibration reset

Resetting the user calibration to the original factory calibration will
reset all methods and ranges.
A user calibrated method is indicated by an arrow while the test result
is displayed.
To reset the calibration press both the [MODE] and [ZERO/TEST] key
and hold.
Switch the unit on using the [ON/OFF] key.
Release the [MODE] and [ZERO/TEST] keys after approx. 1 second.
The following messages will appear in turn on the display:
The factory setting is active.
(SEL stands for Select)
or:
Calibration has been set by the user.
(If the user calibration is to be retained, switch the unit off using the
[ON/OFF] key).
Calibration is reset to the factory setting by pressing the [MODE] key.
The following messages will appear in turn on the display:
Switch the unit off using the [ON/OFF] key.
25

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Lovibond MD110 and is the answer not in the manual?

Questions and answers

Table of Contents